Product Overview
Ensure your sugar analysis refractometer is properly calibrated with our ready-to-use 50% Brix solution. This high-quality standard is manufactured in our state-of-the-art facilities and comes with a Certificate of Analysis. The Certificate of Analysis provides the lot number, reference value and expiry date for traceability.
Key Features
- Supplied with a Certificate of Analysis
- 50% Brix standard value
- Reusable
Instructions for Use
Preparing a standard Brix (sucrose) solution: to make a Brix solution, follow the procedure below:
- Place a container (such as a glass bottle with a cap or a dropper bottle) on an analytical balance.
- Tare the balance.
- Weigh X grams of high-purity sucrose (CAS#: 57-50-1) directly into a beaker.
- Add distilled or deionized water to the beaker until the total weight of the solution is 100 g.
Note: solutions above 65% may require vigorous stirring or shaking and heating in a water bath to approximately 40°C (104°F). Once the sucrose has dissolved, remove the solution, allow it to cool and then use it.
Example of preparing a 25% Brix solution:
| % Brix | Sucrose (g) | Water (g) | Total (g) |
| 25 | 25.000 | 75.000 | 100.000 |