Instrumentation · Engineering · Package · PLC
0755-82221901 sales@ainstru.com

HI764113/10

HI764113/10
Optical DO Probe
HANNA Instruments HI764113/10
The probe measures dissolved oxygen based on the principle of fluorescence quenching, featuring an immobilized Pt based luminophore excited by a blue LED and emitting a red light.

Overview

The probe measures dissolved oxygen based on the principle of fluorescence quenching, featuring an immobilized Pt based luminophore excited by a blue LED and emitting a red light. The Smart Cap™ is locked in place on the optical probe and includes the immobilized O2 sensitive luminophore with a rugged insoluble black oxygen permeable protective layer. The probe comes with a built-in temperature sensor, protective shield and 10 m (32.8') cable.

Specifications

ParameterValue
SKUHI764113/10
Range0.00 to 50 mg/L (ppm); 0.0 to 500.0% saturation
Resolution0.01 mg/L (ppm); 0.1% saturation
Dissolved Oxygen Accuracy1.5% of reading ± 0.01mg/L for 0.00-20.00mg/L; 5% of reading for 20.00-50.00mg/L; 1.5% of reading ±0.1% for 00-200.0%; 5% of reading for 200.0-500.0%
Calibrationone or two points automatic calibration at 100% (8.26 mg/L) and 0% (0 mg/L); Single point manual using a value entered by the user in % saturation or mg/L
Temperature Range-5.0 to 50.0°C (23 to 122°F)
Temperature Resolution0.1°C (0.1°F)
Temperature Accuracy±0.3°C (±0.4°F)
Temperature Compensationautomatic from -5.0 to 50.0°C (23.0 to 122.0°F)
Pressure Compensationautomatic from 420 to 850 mmHg
Salinity Compensationautomatic from 0 to 70 PSU (manually set)
ProbeHI764113 optical DO probe with stainless steel, weighted protective sleeve, internal temperature sensor, 7-pin DIN connector and 4m (13') cable (included)