Instrumentation · Engineering · Package · PLC
0755-82221901 sales@ainstru.com

HI7698594/30

HI7698594/30
Multiparameter pH/EC/Turbidity/opdo Probe
HANNA Instruments HI7698594/30
This model is a multiparameter pH/EC/turbidity/opdo probe for HI98594, supplied with a 30 m (98') cable.

Overview

This model is a multiparameter pH/EC/turbidity/opdo probe for HI98594, supplied with a 30 m (98') cable. The page lists measurement ranges, resolution, accuracy, and calibration for TDS, salinity, seawater Sigma, turbidity, atmospheric pressure, and temperature. The instrument has IP67 protection rating and uses 4 x 1.5 V AA alkaline batteries plus an internal Li-ion rechargeable battery.

Specifications

ParameterValue
SKUHI7698594/30
TDS Range0 to 400000 ppm (mg/L) (the maximum value depends on the TDS factor)
TDS ResolutionManual: 1 ppm (mg/L); 0.001 ppt (g/L); 0.01 ppt (g/L); 0.1 ppt (g/L); 1 ppt (g/L); Automatic: 1 ppm (mg/L) from 0 to 9999 ppm (mg/L); 0.01 ppt (g/L) from 10.00 to 99.99 ppt (g/L); 0.1 ppt (g/L) from 100.0 to 400.0 ppt (g/L)
TDS Accuracy±1 % of reading or ±1 ppm (mg/L), whichever is greater
TDS CalibrationBased on conductivity or salinity calibration
Salinity Range0.00 to 70.00 PSU
Salinity Resolution0.01 PSU
Salinity Accuracy±2% of reading or ±0.01 PSU, whichever is greater
Salinity CalibrationOne point, using a custom solution
Seawater Sigma Range0.0 to 50.0 ıt, ı0, ı15
Seawater Sigma Resolution0.1 ıt, ı0, ı15
Seawater Sigma Accuracy±1.0 ıt, ı0, ı15
Seawater Sigma CalibrationBased on conductivity or salinity calibration
Turbidity Range0.0 to 99.9 FNU; 100 to 1000 FNU
Turbidity Resolution0.1 FNU from 0.0 to 99.9 FNU; 1 FNU from 100 to 1000 FNU
Turbidity Accuracy±0.3 FNU or ±2 % of reading, whichever is greater
Turbidity CalibrationUp to three points using 0 FNU, 20 FNU, 200 FNU, and a custom