HI-97757-11 CAL Check Standards for Ozone Validation of the HI-97757 involves verifying the concentration of the certified CAL Check standards. The CAL Check screen guides the user step-by-step through the validation process and user calibration (if necessary). WARNING: Do not use any solutions or standards other than the Hanna Instruments® CAL Check Standards.
For accurate validation and calibration results, please perform these at room temperature, 18 to 25 °C. Note: CAL Check Standards will not read the specified value in measurement mode. Protect the CAL Check cuvettes from direct sunlight by keeping them in the original packing.
Store between 5 and 30 °C, do not freeze. How to perform a CAL Check: Press CAL Check from measurement mode. The “Not Available“ message or the date, time and status of the last CAL Check will be displayed on the screen.
Press Check to start a new CAL Check. Press the key at any time to abort the validation process. Use the functional keys to enter the certificate value of the calibration standard found on the CAL Check™ Standard Certificate.
Press Next to continue. Note: This value will be saved in the instrument for future validation. If a new set of calibration standards is obtained, please update the certificate value.
Insert the A ZERO CAL Check Cuvette A then press Next to continue. The “Please wait...” message will be displayed during the measurement. Insert the HI-97757B CAL Check Cuvette B then press Next to continue.
The “Please wait…“ message will be displayed during the measurement. When the CAL Check is complete, the display will show one of the following messages and the value obtained during the measurement: “PASSED“: The measured value is within the accuracy specification, no user calibration is required. “OUT OF SPECIFICATION“ and Calibrate is available: The measured value is near the expected value.
To update the user calibration press Calibrate. Press Accept to confirm or Cancel to return to the previous screen. “OUT OF SPECIFICATION“: A user calibration is not allowed, the measured value is outside of the tolerance window.
Check the certified value, expiration date and clean the outside of the cuvette. Repeat the CAL Check procedure. If this error continues contact Hanna Instruments UK here.
Related Products
HANNA Instruments1382mg/L (ppm) TDS Calibration Solution, 230mL bottle
HANNA Instruments111 800uS/cm Conductivity Solution, 230 mL bottle
HANNA Instruments5000µS/cm Conductivity Solution, 230mL bottle
HANNA InstrumentsPotassium Iodide Powder Packets, 100 Packets
HANNA InstrumentsCombination Buffer Kit 10.01 & 7.01 pH 5 x 5 Sachets & Certificate of Analysis
HANNA InstrumentsCombination Buffer Kit 10.01 & 7.01 pH 5 x 5 20ml Sachets
HANNA InstrumentsCombination pH/TDS Buffer Kit 1500 ppm & 7.01 pH
HANNA Instruments