TriOxmatic® 702 IQ
Digital electro-chemical IQ sensors for dissolved oxygen, connectable to the IQ SENSOR NET
TriOxmatic® IQ: Reliable and proven digital electro-chemical oxygen sensors with 3 electrode system (ECDO) for precise and accurate measurements:
- Low investment costs
- No running in, no long-term drift - stable from the beginning to the end
- Self-diagnosis system SensReg/SensLeck by means of 3 electrode system
The electro-chemical sensors are known for their high accuracy - from the beginning. Their teflon membrane makes them extremely insensitive to biological deposits. Therefore, in most cases, you will not need an additional cleaning unit.
Also, the membrane - just like the electrolyte consumption - is monitored continuously. Our self-diagnosis system SensLeck and SensReg show occurring errors automatically.
The sea water model is optimized for the use in special media: All wetted parts consist of titanium and plastic and are therefore extremely corrosion-resistant.
Furthermore, if you want, you will find a sensor with a faster response time or for the measurement of trace oxygen.
|
Model |
TriOxmatic® 700 IQ |
TriOxmatic® 700 IQ SW* |
TriOxmatic® 701 IQ |
TriOxmatic® 702 IQ |
|
Order No. |
201640 |
201641 |
201644 |
201646 |
| * SW: Sensor as sea water model (with plastic arming (POM)) | ||||
|
Measuring method |
Electrochemical |
|||
|
Measuring range (25 °C) |
||||
|
O2 concentration |
0.0 ... 60.0 mg/l |
0.00 ... 20.00 mg/l |
0 ... 2000 µg/l |
|
|
O2 saturation |
0 ... 600% |
0.0 … 200.0% |
0 … 110% |
|
|
Resolution |
||||
|
O2 concentration |
0.1 mg/l |
0.01 mg/l |
0.001 mg/l |
|
|
O2 saturation |
1% |
0.1% |
0.1% |
|
|
Response time at 25 °C |
t90: 180 s |
t90: 30 s |
t90: 30 s |
|
|
Minimum flow rate |
0.05 m/s |
0.23 m/s |
0.3 m/s |
|
|
SensCheck |
SensLeck |
|
SensLeck |
— |
|
Temp. measurement |
Integrated NTC, 23 °F ... 140 °F (-5 °C ... +60 °C) ± 0.5 °C |
|||
|
Temp. compensation |
32 °F ... 140 °F (0 °C ... +60 °C) |
|||
|
Pressure Resistance |
10 bar (incl. sensor connection cable) |
|||
|
Ambient Conditions |
Operating temperature: 32 °F ... 140 °F (0 °C ... +60 °C); Storage temperature: 23 °F ... 149 °F (-5 °C ... +65 °C) |
|||
|
Electrical connections |
2-wire shield cable with quick fastener to sensor |
|||
|
Electromagnetic Compatibility |
EN 61326, Class B, FCC Class A; |
|||
|
Certifications |
CE, cETL, ETL |
|||
|
Mechanical |
Membrane head assembly, locking cap: POM |
|||
|
Weight (without cable) |
Approx. 1.46 lb (660 g) |
Approx. 2.58 lb (1,170 g) |
Approx. 1.46 lb (660 g) |
|
|
Warranty |
2 years for defects in quality |
|||
Case Studies
Municipal Wastewater Treatment · WTW — Stock Configuration Case
Customer Profile
The typical setting in Municipal Wastewater Treatment is the inlet, biological and outlet stages of a municipal plant. The measuring point is WTW; the focus is covering online and laboratory water analysis with WTW instruments.
Key Difficulties
Conditions are high solids and corrosive fractions, with large diurnal flow swings and, on top of that, effluent parameters drift and probes foul with sludge and biofilm, which a standard configuration cannot handle for long. The design therefore starts with selected by parameter set, range and digital sensor protocol.
Implementation
The scope ran through the stock channel: configuring from stock and delivered quickly. Technically it hinges on selected by parameter set, range and digital sensor protocol, and on these requirements: stable compliance with Grade 1A discharge limits and online data reporting.
Customer Benefit
In practice this means plug-and-play digital sensors that keep calibration when swapped, which cut the workload on the site maintenance team.
Food & Beverage · WTW — Stock Configuration Case
Project Background
The project covers dosing, pasteurising, filling and CIP stages. The work centres on WTW and the core task is covering online and laboratory water analysis with WTW instruments.
Operating Challenges
The medium and duty are characterised by syrups, dairy and beverages that scale easily, under strict hygienic requirements, and frequent cleaning demands instruments that withstand hot CIP/SIP caustic cycles. So WTW cannot be specified generically — it calls for selected by parameter set, range and digital sensor protocol.
Solution
The Ainstru engineering team worked through the stock channel: configuring from stock and delivered quickly. The configuration involves selected by parameter set, range and digital sensor protocol, with these requirements in view: food-safety and hygienic design per 3-A / EHEDG.
Results
Since start-up the system has run steadily; the practical result is plug-and-play digital sensors that keep calibration when swapped.
Chemical & Pharmaceutical · WTW — Stock Configuration Case
Project Overview
The site is reactors, metering vessels and finished-product tanks. The customer wanted WTW properly configured on the existing plant, with the aim of covering online and laboratory water analysis with WTW instruments.
Technical Challenges
What actually causes trouble is highly corrosive media with solvents and crystallisation, partly in explosive atmospheres, and corrosion plus Ex protection shortens instrument life and makes downtime costly. All of this means the WTW configuration has to involve selected by parameter set, range and digital sensor protocol.
Configuration
Supply and installation were organised through the stock channel: configuring from stock and delivered quickly. On site this meant selected by parameter set, range and digital sensor protocol, checked line by line against: Ex certification and GMP traceability of records.
Performance
After commissioning the outcome is plug-and-play digital sensors that keep calibration when swapped, and the readings are now used for process tuning.
Power & Energy · WTW — Stock Configuration Case
Site Conditions
This is a Power & Energy customer whose plant is at boiler feedwater, turbine condensate and FGD/DeNOx systems. The main task was covering online and laboratory water analysis with WTW instruments.
The Problem
The difficulty lies in high temperature and pressure with both steam and saline water; at the same time drift under high temperature and pressure, with very costly unplanned outages. Specification therefore means selected by parameter set, range and digital sensor protocol.
What We Did
We took a stock channel approach — configuring from stock and delivered quickly. Delivery was built around selected by parameter set, range and digital sensor protocol, with point-by-point checks on: compliance with power-industry codes and unit availability targets.
Operating Outcome
In service the system performs as intended — plug-and-play digital sensors that keep calibration when swapped, with routine maintenance falling noticeably.