Instrumentation · Engineering · Package · PLC
0755-82221901 sales@ainstru.com

302222 LRD 01-7

302222 LRD 01-7

LRD 01-7

WTW 302222
2 electrode screw-in conductivity cell, 1/2 inch NPT, for use with analogue monitors. With integrated temperature sensor, 7 m (23 ft) cable and watertight plug (IP68)

Product Overview

  • Conductivity measuring cells to be connected to analog transmitters

  • All analog conductivity measuring cells are equipped with an integrated temperature compensation and cover all applications:

  • In wastewater treatment plants, as flow measurement or in the pipeline: The right solution for every application

  • Highest linearity by 4 electrode measuring cell

  • High operational safety by robust design

  • You have the application - we have the application - we have the measuring cell! The unique 4 electrode measuring cell compensates for possible contamination and polarization effects. All measuring cells can be connected to the LF monitor LF 298 NTC (exception: TetraCon® 700).

  • TetraCon® 700 – Designed as an immersion encoder specifically for use in wastewater treatment plants

  • Conductivity measuring cell with 4 electrode system, integrated temperature sensor, anti-slip electrodes and breakage-resistance shaft; measuring range 10 µS/cm to 1000 mS/cm. Only as replacement armature for analog monitors of the Ecoline® or QuadroLine® series. Still available as digital version (IQ SENSOR NET System 181).

  • LRD 01 – for the installation into pipelines

  • Conductivity measuring cell as V4A model. With integrated temperature sensor (up to 130 °C), measuring range 0.001 to 200 µS/cm, pressure-resistant to 14 bar, 1/2 inch NPT screw thread.

  • LRD 325 – for the installation into pipelines

  • Conductivity measuring cell with integrated temperature sensor (up to 100 °C). Measuring range 1 µS/cm to 2 S/cm, pressure-resistant to 10 bar, 1/2 inch NPT screw thread.

Model

LRD 01

LRD 325

LR 325/01

LR 325/001

TetraCon® 325

TetraCon® DU/T

Order No.

302220, 302222

302225, 302229

301961

301962

301960, 301970, 301971

301252

Measuring principle

2 electrodes measuring cell

4 electrodes measuring cell

2 electrodes measuring cell

4 electrodes measuring cell

Measuring Range

0.001 µS/cm ... 200 µS/cm

1 µS/cm ...
2 S/cm

0.001 µS/cm ... 200 µS/cm

0.0001 µS/cm ... 30 µS/cm

1 µS/cm ...
2 S/cm

Cell constant

0.1 cm-1, ±2%

0.475 cm-1, ±1.5%

K = 0.1 cm-1

K = 0.01 cm-1

K = 0.475 cm-1

K = 0.778 cm-1

Resolution

Depends on measuring range

Temperature sensor

Integrated NTC

Temperature Measurement

0 °C ... +130 °C

0 °C ... +100 °C

-5 °C ... 80 °C

0 °C ... 60 °C

Maximum pressure

14 bar (at 20 °C)

10 bar (at 20 °C)

2 bar

Electrical Connection

Integrated PU connection cable with 7-pole screw plug (IP 65)

Integrated cable mit 8-pole plug

8-pole socket for cable KKDU 325

Mechanical

Shaft

Stainless steel 1.4571

Epoxy

POM

Kable gland

Brass, nickel-plated

—

—

—

—

Connection head

—

—

POM

—

Electrodes

Stainless steel 1.4571

Graphite

Stainless steel 1.4571

Graphite

Protection Rating

IP65

IP68
(Sensor and cable)

IP65
(when connected)

(when connected)

(Measuring cell until immersion length)

Weight (without cable)

Approx. 350 g

Approx. 300 g

Approx. 135 g

Approx. 280 g

Approx. 135 g

Approx. 170 g

Warranty

2 years on defects in quality according to § 10 terms of conditions

Municipal Wastewater Treatment

Municipal Wastewater Treatment · WTW — Stock Configuration Case

Customer Profile

The typical setting in Municipal Wastewater Treatment is the inlet, biological and outlet stages of a municipal plant. The measuring point is WTW; the focus is covering online and laboratory water analysis with WTW instruments.

Key Difficulties

Conditions are high solids and corrosive fractions, with large diurnal flow swings and, on top of that, effluent parameters drift and probes foul with sludge and biofilm, which a standard configuration cannot handle for long. The design therefore starts with selected by parameter set, range and digital sensor protocol.

Implementation

The scope ran through the stock channel: configuring from stock and delivered quickly. Technically it hinges on selected by parameter set, range and digital sensor protocol, and on these requirements: stable compliance with Grade 1A discharge limits and online data reporting.

Customer Benefit

In practice this means plug-and-play digital sensors that keep calibration when swapped, which cut the workload on the site maintenance team.

Food & Beverage

Food & Beverage · WTW — Stock Configuration Case

Project Background

The project covers dosing, pasteurising, filling and CIP stages. The work centres on WTW and the core task is covering online and laboratory water analysis with WTW instruments.

Operating Challenges

The medium and duty are characterised by syrups, dairy and beverages that scale easily, under strict hygienic requirements, and frequent cleaning demands instruments that withstand hot CIP/SIP caustic cycles. So WTW cannot be specified generically — it calls for selected by parameter set, range and digital sensor protocol.

Solution

The Ainstru engineering team worked through the stock channel: configuring from stock and delivered quickly. The configuration involves selected by parameter set, range and digital sensor protocol, with these requirements in view: food-safety and hygienic design per 3-A / EHEDG.

Results

Since start-up the system has run steadily; the practical result is plug-and-play digital sensors that keep calibration when swapped.

Chemical & Pharmaceutical

Chemical & Pharmaceutical · WTW — Stock Configuration Case

Project Overview

The site is reactors, metering vessels and finished-product tanks. The customer wanted WTW properly configured on the existing plant, with the aim of covering online and laboratory water analysis with WTW instruments.

Technical Challenges

What actually causes trouble is highly corrosive media with solvents and crystallisation, partly in explosive atmospheres, and corrosion plus Ex protection shortens instrument life and makes downtime costly. All of this means the WTW configuration has to involve selected by parameter set, range and digital sensor protocol.

Configuration

Supply and installation were organised through the stock channel: configuring from stock and delivered quickly. On site this meant selected by parameter set, range and digital sensor protocol, checked line by line against: Ex certification and GMP traceability of records.

Performance

After commissioning the outcome is plug-and-play digital sensors that keep calibration when swapped, and the readings are now used for process tuning.

Power & Energy

Power & Energy · WTW — Stock Configuration Case

Site Conditions

This is a Power & Energy customer whose plant is at boiler feedwater, turbine condensate and FGD/DeNOx systems. The main task was covering online and laboratory water analysis with WTW instruments.

The Problem

The difficulty lies in high temperature and pressure with both steam and saline water; at the same time drift under high temperature and pressure, with very costly unplanned outages. Specification therefore means selected by parameter set, range and digital sensor protocol.

What We Did

We took a stock channel approach — configuring from stock and delivered quickly. Delivery was built around selected by parameter set, range and digital sensor protocol, with point-by-point checks on: compliance with power-industry codes and unit availability targets.

Operating Outcome

In service the system performs as intended — plug-and-play digital sensors that keep calibration when swapped, with routine maintenance falling noticeably.