WTW 600320
根据DIN EN ISO 7027-1,用于浊度测量(90°)的便携式浊度计,包括校准标准试剂盒(0.02-10-1000 NTU/FNU)、5个空试管、组织、棉絮。(4 AA Mignon),适合饮用水。可选:配备软件包LSdata Flex/430的LabStation LS Flex/430,用于符合GLP标准的数据导出和rech.batt。RB Flex/430(订单号251 301)或可充电电池RB Flex/440(订单号251300,PC软件LSdata Flex/430,独立902762)。

Turbidity measurement acc. To DIN ISO 27027 for any demand up tp 1100 NTU with lab option
Turb® 430 IR meet the requirements of nephelometric measurement at 90° stray light for drinking water with highest precision particularly suitable in the range < 1 NTU and is therefore suitable for water analysis,control, such, fuel, cell cultures, food and beverage industries and (health) administration surveillance.

• Measuring range comprises 0.02 -1100 NTU/FNU with automatic measuring range switching
• Adjustable calibration intervals and GLP-compliant documentation
• Stray light behavior according to pharmacopoeia 5.0
• Higly precise AMCO Clear® Standards
The measurement and automatic 3-point calibration runs menu-guided. The calibration takes place via an AMCO Clear® standard set (0.02-10-1000 NTU). Up to 1000 data sets including sample ID numbers can be stored and later put out GLP-compliant via the PC software LSdata . The quality assurance of the measured results is supported by calibration interval setting with documentation. This makes Turb® 430 not only a portable meter – also available in a practical field set - but also a precise and reliable benchtop meter for applications of up to 1100 NTU/FNU with optimized data administration!

AMCO Clear® standards

Polymer calibration standards resist against Formazine by significantly higher accuracy and stability without drifting, which increases the measuring accuracy significantly:
•Manufacturing tolerance of only ±1% for highest result accuracy
•Long-term stable particle distribution and size
•No health hazard
•Easy to dispose of
•N.I.S.T. traceability

More informationen in: "Turbidity An interesting parameter - Brief Overview, Measurement Tips & Tricks"

Method Nephelometrisch (90° Straylight)
Light Source IR LED 830-890 mm
acc. to Standard Method DIN EN ISO 7027-1
Measuring Range 0-1100 FNU/NTU
Resolution 0,01 for 0,02 -9,99
  0,1 for 10 – 99,90
  1 for 100 - 1100
Accuracy 0.01 NTU or +/- 2% of meas.value
Reproducability < 0.5%
Calibration Preset standard 3 point Cal, selectable 2-5 point (FreeCal), 1P QuickCal
Calibration Protocol yes
Calibration Interval yes
Data Storage 2000 values
Response Time Ca. 3 s
Cuvette 28 x 60 mm, 20 ml Sample Volume
Ident Number yes
Firmware-Update yes
Interface RS 232, USB via Adapter (optional)
Operation Temp. 0... +50°C
Power Supply 4 Mignon (AA) > 3000 Measurements
Rech.Batt. optional
Data Management GLP compliant optional PC software LSdata
LabStation+ PC-Software LSdata optional as small benchtop solution