Instrumentation · Engineering · Package · PLC
0755-82221901 sales@ainstru.com

600325 Turb 430 T

600325 Turb 430 T

Turb 430 T

WTW 600325
Portable turbidimeter compliant with US EPA 180.1 (90°, tungsten), including calibration standard kit (0.02-10-1000 NTU) and accessories: 5 empty test tubes (28 mm), cleaning tissues, cotton wool. Optional: LabStation LS Flex/430 with software package LS Flex/440 for GLP-compliant data export (order no. 251301) or rechargeable battery RB Flex/430 (order no. 25.1300), PC software LS Flex/490 (separate order no. 902762)

Product Overview

  • Turbidity measurement acc. to US EPA 180.1 for any demand up tp 1100 NTU with lab option

  • Turb® 430 T meet the requirements of nephelometric measurement at 90° stray light for drinking water with highest precision particularly suitable in the range < 1 NTU and is therefore suitable for water analysis,control, such, fuel, cell cultures, food and beverage industries and (health) administration surveillance.

  • Measuring range comprises 0.02 -1100 NTU/FNU with automatic measuring range switching

  • AQA: Calibration interval setting and GLP-compliant documentation

  • Stray light behavior according to pharmacopoeia 5.0

  • Higly precise AMCO Clear® Standards

  • The measurement and automatic 3-point calibration runs menu-guided. The calibration takes place via an AMCO Clear® standard set (0.02-10-1000 NTU). Up to 1000 data sets including sample ID numbers can be stored and later put out GLP-compliant via the PC software LSdata . The quality assurance of the measured results is supported by calibration interval setting with documentation. This makes Turb® 430 not only a portable meter – also available in a practical field set - but also a precise and reliable benchtop meter for applications of up to 1100 NTU/FNU with optimized data administration!

  • AMCO Clear® standards

  • Polymer calibration standards resist against Formazine by significantly higher accuracy and stability without drifting, which increases the measuring accuracy significantly:

  • Manufacturing tolerance of only ±1% for highest result accuracy

  • Long-term stable particle distribution and size

  • No health hazard

  • Easy to dispose of

  • N.I.S.T. traceability

  • More informationen in: "Turbidity An interesting parameter - Brief Overview, Measurement Tips & Tricks".

Method Nephelometrisch (90° Straylight)
Light Source Tungsten
acc. to Standard Method US-EPA 180.1
Measuring Range 0-1100 NTU
Resolution 0,01 for 0,02 -9,99
  0,1 for 10 – 99,90
  1 for 100 - 1100
Accuracy 0.01 NTU or +/- 2% of meas.value
Reproducability < 1,0%
Calibration Preset standard 3 point Cal, selectable 2-5 point (FreeCal), 1P QuickCal
Calibration Protocol yes
Calibration Interval yes
Data Storage 1000 values
Response Time Ca. 7s
Cuvette 28 x 60 mm, 20 ml Sample Volume
Ident Number yes
Firmware-Update yes
Interface RS 232, USB via Adapter (optional)
Operation Temp. 0... +50°C
Power Supply 4 Mignon (AA) > 3000 Measurements
Rech.Batt. optional
Data Management GLP compliant optional PC software LSdata
LabStation+ PC-Software LSdata optional as small benchtop solution
Municipal Wastewater Treatment

Municipal Wastewater Treatment · WTW — Stock Configuration Case

Customer Profile

The typical setting in Municipal Wastewater Treatment is the inlet, biological and outlet stages of a municipal plant. The measuring point is WTW; the focus is covering online and laboratory water analysis with WTW instruments.

Key Difficulties

Conditions are high solids and corrosive fractions, with large diurnal flow swings and, on top of that, effluent parameters drift and probes foul with sludge and biofilm, which a standard configuration cannot handle for long. The design therefore starts with selected by parameter set, range and digital sensor protocol.

Implementation

The scope ran through the stock channel: configuring from stock and delivered quickly. Technically it hinges on selected by parameter set, range and digital sensor protocol, and on these requirements: stable compliance with Grade 1A discharge limits and online data reporting.

Customer Benefit

In practice this means plug-and-play digital sensors that keep calibration when swapped, which cut the workload on the site maintenance team.

Food & Beverage

Food & Beverage · WTW — Stock Configuration Case

Project Background

The project covers dosing, pasteurising, filling and CIP stages. The work centres on WTW and the core task is covering online and laboratory water analysis with WTW instruments.

Operating Challenges

The medium and duty are characterised by syrups, dairy and beverages that scale easily, under strict hygienic requirements, and frequent cleaning demands instruments that withstand hot CIP/SIP caustic cycles. So WTW cannot be specified generically — it calls for selected by parameter set, range and digital sensor protocol.

Solution

The Ainstru engineering team worked through the stock channel: configuring from stock and delivered quickly. The configuration involves selected by parameter set, range and digital sensor protocol, with these requirements in view: food-safety and hygienic design per 3-A / EHEDG.

Results

Since start-up the system has run steadily; the practical result is plug-and-play digital sensors that keep calibration when swapped.

Chemical & Pharmaceutical

Chemical & Pharmaceutical · WTW — Stock Configuration Case

Project Overview

The site is reactors, metering vessels and finished-product tanks. The customer wanted WTW properly configured on the existing plant, with the aim of covering online and laboratory water analysis with WTW instruments.

Technical Challenges

What actually causes trouble is highly corrosive media with solvents and crystallisation, partly in explosive atmospheres, and corrosion plus Ex protection shortens instrument life and makes downtime costly. All of this means the WTW configuration has to involve selected by parameter set, range and digital sensor protocol.

Configuration

Supply and installation were organised through the stock channel: configuring from stock and delivered quickly. On site this meant selected by parameter set, range and digital sensor protocol, checked line by line against: Ex certification and GMP traceability of records.

Performance

After commissioning the outcome is plug-and-play digital sensors that keep calibration when swapped, and the readings are now used for process tuning.

Power & Energy

Power & Energy · WTW — Stock Configuration Case

Site Conditions

This is a Power & Energy customer whose plant is at boiler feedwater, turbine condensate and FGD/DeNOx systems. The main task was covering online and laboratory water analysis with WTW instruments.

The Problem

The difficulty lies in high temperature and pressure with both steam and saline water; at the same time drift under high temperature and pressure, with very costly unplanned outages. Specification therefore means selected by parameter set, range and digital sensor protocol.

What We Did

We took a stock channel approach — configuring from stock and delivered quickly. Delivery was built around selected by parameter set, range and digital sensor protocol, with point-by-point checks on: compliance with power-industry codes and unit availability targets.

Operating Outcome

In service the system performs as intended — plug-and-play digital sensors that keep calibration when swapped, with routine maintenance falling noticeably.