Instrumentation · Engineering · Package · PLC
0755-82221901 sales@ainstru.com

Proline Promag 10D 诊断代码

共 16 条诊断代码 · 6 条常见故障 · 来源手册 BA00082DEN_1917.pdf

全部 16 未标注 16
代码 信号 简述 行为 页码
001 ? CRITICAL FAIL. Alarm P78
011 ? AMP HW EEPROM Alarm P78
012 ? AMP SW EEPROM Alarm P78
101 ? GAIN ERROR AMP Alarm P78
111 ? CHECKSUM TOTAL. Alarm P78
321 ? TOL. COIL CURR. Alarm P78
351 ? CURRENT RANGE Warning P78
359 ? RANGE PULSE Warning P79
401 ? EMPTY PIPE Warning P79
461 ? ADJ. NOT OK Warning P79
463 ? FULL = EMPTY Alarm P79
501 ? SW.-UPDATE ACT. Warning P79
502 ? UP-/DOWNL. ACT. Warning P79
601 ? POS. ZERO-RET. Warning P79
691 ? SIM. FAILSAFE Alarm P79
692 ? SIM. VOL. FLOW Warning P79

常见故障(无代码号)

现象可能原因处理方法
Flow values are negative, even though the fluid is flowing forwards through the pipe. 1. Remote version: – Switch off the power supply and check the wiring → 46 – If necessary, reverse the connections at terminals 41 and 42 2. Change the setting in the "INSTALLATION DIRECTION SENSOR" function accordingly
Measured-value reading fluctuates even though flow is steady. 1. Check grounding and potential equalization → 54 2. Check the fluid for presence of gas bubbles. 3. In the "SYSTEM DAMPING" function → increase the value
Measured-value reading shown on display, even though the fluid is at a standstill and the measuring tube is full. 1. Check grounding and potential equalization → 54 2. Check the fluid for presence of gas bubbles. 3. Activate the "LOW FLOW CUTOFF" function, i.e. enter or increase the value for the switching point.
Measured-value reading on display, even though measuring tube is empty. 1. Perform empty-pipe/full-pipe adjustment and then switch on Empty Pipe detection → 72 2. Remote version: Check the terminals of the EPD cable → 46 3. Fill the measuring tube.
The current output signal is always 4 mA, irrespective of the flow signal at any given time. 1. Select the "BUS ADDRESS" function and change the setting to "0". 2. Value for creepage too high. Reduce the value in the "LOW FLOW CUTOFF" function.
The fault cannot be rectified or some other fault not described above has arisen. In these instances, please contact your Endress+Hauser service organization. The following options are available for tackling problems of this nature: Request the services of an Endress+Hauser service technician If you contact our service organization to have a service technician sent out, please be ready to quote the following information: – Brief description of the fault – Nameplate specifications (→ 6): order code, serial number Returning devices to Endress+Hauser The necessary procedures on → 5 must be carried out before you return a flowmeter requiring repair or calibration to Endress+Hauser. Always enclose a duly completed "Declaration of Conformity" form with the flowmeter. You will find a master copy of this form at the back of this manual. Replace transmitter electronics Components in the measuring electronics defective →order spare parts → 82
Flow values are negative, even though the fluid is flowing forwards through the pipe. 1. Remote version: – Switch off the power supply and check the wiring → 46 – If necessary, reverse the connections at terminals 41 and 42 2. Change the setting in the "INSTALLATION DIRECTION SENSOR" function accordingly
Measured-value reading fluctuates even though flow is steady. 1. Check grounding and potential equalization → 54 2. Check the fluid for presence of gas bubbles. 3. In the "SYSTEM DAMPING" function → increase the value
Measured-value reading shown on display, even though the fluid is at a standstill and the measuring tube is full. 1. Check grounding and potential equalization → 54 2. Check the fluid for presence of gas bubbles. 3. Activate the "LOW FLOW CUTOFF" function, i.e. enter or increase the value for the switching point.
Measured-value reading on display, even though measuring tube is empty. 1. Perform empty-pipe/full-pipe adjustment and then switch on Empty Pipe detection → 72 2. Remote version: Check the terminals of the EPD cable → 46 3. Fill the measuring tube.
The current output signal is always 4 mA, irrespective of the flow signal at any given time. 1. Select the "BUS ADDRESS" function and change the setting to "0". 2. Value for creepage too high. Reduce the value in the "LOW FLOW CUTOFF" function.
The fault cannot be rectified or some other fault not described above has arisen. In these instances, please contact your Endress+Hauser service organization. The following options are available for tackling problems of this nature: Request the services of an Endress+Hauser service technician If you contact our service organization to have a service technician sent out, please be ready to quote the following information: – Brief description of the fault – Nameplate specifications (→ 6): order code, serial number Returning devices to Endress+Hauser The necessary procedures on → 5 must be carried out before you return a flowmeter requiring repair or calibration to Endress+Hauser. Always enclose a duly completed "Declaration of Conformity" form with the flowmeter. You will find a master copy of this form at the back of this manual. Replace transmitter electronics Components in the measuring electronics defective →order spare parts → 82
需要 Proline Promag 10D 的备件或技术支持? 联系我们 ,或致电 0755-82221901。