The Hach model 2200 PCX is the first particle counter designed specifically to meet the unique requirements of water treatment. The model 2200 PCX can be integrated into your plant control system, used with our water quality software products, or both in parallel. Collected data can be utilized for optimizing chemical dosage, backwash cycles, and filter-to-waste times. Particle count reports can also be used to apply for removal credits and filter high rate studies.
主要特点
- Supports up to 32 size channels and 8 analog inputs (turbidity, pH, etc.):See a wide range of particle data at one time, ensuring you never miss an event in your system.
- Operates stand-alone or as part of a networked system:Use the 2200 PCX as a single data point or as an integrated part of your plant.
- Volumetric-all particles pass through the sensing area:See the whole picture with analysis of the entire sample.
- Superior Sizing Accuracy and Sensor-to-Sensor Matching:Industry-leading calibration yields excellent repeatability.
技术参数
| 参数名称 | 说明 |
|---|---|
| Alarm | Power, counting display, clean sensor and alarm |
| Certifications | Product safety: UL/CSA approved 100-115 V, 50/60 Hz external wall-style power supply. Immunity: EN 50081-2; 1992 (Generic Immunity Standard) per 89/366/EEC EMC: Supporting test records with Elliot Laboratories, Inc. Required Standards include: IEC 801-2: 1991 Electro-static Discharge: IEC 801-3 Radiated RF Electro-magnetic fields: IEC 801-4 Electrical Fast Transients/Burst Emissions: EN50081-2 (Generic Emissions Standard) per 89/336/EEC EMC: Supporting test records by Elliot Laboratories, Inc. (NVLAP # 200069-0). This Class A digital Apparatus Meets all requirements of the Canadian Interference - Causing Equipemtn Regulations. FCC Part 15, Class "A" Limits: Supporting test records with Elliot Laboratories, Inc. (NVLAP #200069-0) |
| Communication | RS485 and Analog I/O |
| Communication Capabilities | RS485 and Analog I/O |
| Data Communications Distance | 4000 foot maximum for entire RS485 signal path |
| Enclosure Rating | NEMA 4X |
| Flow Rate | 100 mL/min |
| Fluid Connections Inlet | Barbed fitting with self-sealing quick disconnect for 1/4 inch O.D. tubing |
| Fluid Connections Outlet | Barbed fitting with self-sealing quick disconnect for 1/4 inch O.D. tubing |
| Operating Temperature Range | 0 - 50 °C |
| Particle Size | 2 - 750 microns |
| Power Requirements (Hz) | 60 Hz |
| Power Requirements (Voltage) | 115 VAC |
| Pressure Range | 65 psig not more than 1 minute; 55 psig continuous |
| Warranty | 12 months |
| Weight | 4.89 kg |
| What's included? | Hach 2200PCX Particle Counter, Analog I/O Card, Water Weir, Power Supply with Cable, Cleaning Brush and Solution, PCX Manual, and Water Weir Manual. |
同系列其他配置
| 型号 | |
|---|---|
| 5704000 | 2200 PCX Particle Counter with RS485 I/O and Water Weir, 115 VAC |
资料下载
- Scaling Analog Inputs with the 2200 PCX Explorer
- Particle Counting in a Municipal Drinking Water Treatment Plant
- Introduction to Hach Services
- 2200 PCX Online Particle Counter Data Sheet
- 2200 PCX Particle Counter CSI Specifications
- 2200 PCX Dimensional Drawings
- Basic User Manual: 2200 PCX
- Full User Manual: 2200 PCX Particle Counter
Recomendados similares
HACHNP6000sc Phosphate Analyzers
HACHMódulo de relé SC1000, montaje en carril DIN, 4 relés, máx. 240 V
COVER-SERVICE REPLACEMENT
Kit de montaje con cadena de acero inoxidable
HACHNH6000sc Ammonia Analyzers
Accu4 Spare Chamber with Plumbing, High Pressure
HACHEZ Series Arsenic Analyzers
HACH