产品概述
Simple to Integrate. Simple to Operate.
Our most basic version of the degassed cation conductivity panel (DCCP) system removes CO₂ from the sample stream and measures degassed cation conductivity (DCC). This comprehensive solution saves you time on design, installation, training, maintenance, operation, and plant start-up, while helping protect your most critical assets and equipment.
- Designed with operator safety in mind - Compact, wall-mounted design - Operates in accordance with ASTM D4519
主要特点
- Reliable Measurements:A new way of monitoring based on 9523sc + DCCP calculated pH, delivering reliable results that save you critical time and effort.
- Save Time on Maintenance and Set-Up:The degassed conductivity system is easy to install and requires low annual maintenance as well as easy set-up with cation conductivity and specific conductivity instrumentation.
- Flexible and Compact Design:The 9523sc cation conductivity analyzer can be combined with the degassed conductivity system in a single panel or separated in order to meet the need of each customer.
- Reduce Training Complexity:A single platform minimizes time required to teach and learn product operations, getting new systems in use faster.
技术参数
| 参数名称 | 说明 |
|---|---|
| Accuracy | < 2% |
| Ambient Temperature | 2 - 50 °C (36 - 122 °F) |
| Cell Constant k | 0.01 cm-1 |
| Certifications | CE, UL, CSA |
| Cooling Water | 3.8 - 11.4 L/min (1 to 3 gpm) |
| Enclosure Rating | IP66 / NEMA 4X |
| Fitting Type | Sample inlet/outlet connections: ¼ inch tube fitting |
| Flow | 100 - 150 cc/min (mL/min); 6 - 9 L/h |
| Includes 9523sc | No |
| Measuring Range Conductivity | 0.01 - 200 µS/cm |
| Operating Temperature Range | Recommended operating temperature: 23 - 27 °C (75 - 80 °F) |
| Parameter | Degassed Cation Conductivity |
| Power Consumption | 1.6 k |
| Power Requirements (Hz) | 50/60 Hz |
| Power Requirements (Voltage) | 110 - 120 VAC |
| Regenerative Cooling | No |
| Storage Conditions | 0 - 50 °C (32 - 120 °F) 0 - 85 % relative humidity, non-condensing |
| Weight | 36 kg (79 lbs) |
| What's included? | Kit includes (1 ea.): 9525.99.0010 9525sc Degas System, 9523.99.09P2 Hach 9523sc Specific and Cationic Conductivity Analyzer, and pH Calculator, with 5x 4-20 mA Outputs, 100-240 V AC, 9525.99.7040 Degas System Extension Panel Kit |
同系列其他配置
| 型号 | |
|---|---|
| 9525.99.10CCDG | Hach 9525sc DCCP System with Specific Conductivity (SC), Cation Conduc |
| 9525.99.0011 | Hach 9525sc DCCP System, Degassed Conductivity only, with Regenerative |
| 9525.99.10CCR | Hach 9525sc DCCP System with Specific Conductivity (SC), Cation Conduc |
资料下载
- Ultra-Pure Water Measurements
- Introduction to Hach Services
- Care Advantage Flyer : Complete service solution for all your instruments
- Care Advantage Brochure : Complete service solution for all your instruments
- Complete Water Analysis Solutions for Power and Steam Generation
- Complete Water Analysis Solutions for Power and Steam Generation
- Complete Water Analysis Solutions for the Pulp and Paper Industry
- Contrôle de la qualité de l’eau dans les usines de valorisation énergétique des déchets
Recomendados similares
HACH1990 µS/cm Conductivity Standard; (995 mg/L TDS, 1000 mg/L NaCl), 1L
HACHRadiometer Analytical CDC565 Conductivity Cell (4-pole Pt sensor, k = 1.0 cm-1, epoxy body)
HACHConductivity Solution, 1000-2000 µS/cm, 1 L
HACHConductivity Solution, 2000-150,000 µS/cm, 1 L
HACHConductivity Solution, 200,000-300,000 µS/cm, 1 L
HACHConductivity Solution, 300,000-800,000 µS/cm, 1 L
HACH