产品概述
Simple to Integrate. Simple to Operate.
Hach's most featured version of our degassed cation conductivity panel (DCCP) system measures specific conductivity (SC), cation conductivity (CC), and calculated pH in addition to degassed cation conductivity (DCC) after CO₂ removal from the sample stream. Also included is a built-in regenerative cooler to ensure consistent and effective sampling temperature. This comprehensive solution saves you time on design, installation, training, maintenance, operation, and plant start-up, while helping protect your most critical assets and equipment.
- Designed with operator safety in mind - Compact, wall-mounted design - Operates in accordance with ASTM D4519
主要特点
- Reliable measurements:Hach's newly redesigned Degassed Cation Conductivity Analyzer efficiently removes CO₂ from the sample to deliver fast reliable results that save you critical time and effort.
- Space-Saving Design:Compact, wall mounted footprint to allow for easy integration into existing or new SWAS systems.
- Easy and Safe Handling:The new 9525sc requires no additional cooling water supply but instead allows for a regenerative cooling option and condenses the steam to allow for safe discharge via a vent drain.
技术参数
| 参数名称 | 说明 |
|---|---|
| Accuracy | < 2% |
| Ambient Temperature | 2 - 50 °C (36 - 122 °F) |
| Cell Constant k | 0.01 cm-1 |
| Enclosure Rating | IP66 / NEMA 4X |
| Flow | 100 - 150 cc/min (mL/min); 6 - 9 L/h |
| Includes 9523sc | Yes |
| Measuring Range Conductivity | 0.01 - 200 µS/cm |
| Operating Temperature Range | Recommended operating temperature: 23 - 27 °C (75 - 80 °F) |
| Parameter | Degassed Cation Conductivity, Specific Conductivity, Cation Conductivity, Calculated pH |
| Power Requirements (Hz) | 60 Hz |
| Power Requirements (Voltage) | 110 - 120 VAC |
| Regenerative Cooling | Yes |
| Sample Cooler Type | DTC-DCCP |
| Storage Conditions | 0 - 50 °C (32 - 120 °F) 0 - 85 % relative humidity, non-condensing |
| Warranty | 12 months |
| Weight | 50 kg (110 lbs) |
| What's included? | Kit includes (1 ea.): 9525.99.0010 9525sc Degas System, 9523.99.09P2 Hach 9523sc Specific and Cationic Conductivity Analyzer, and pH Calculator, with 5x 4-20 mA Outputs, 100-240 V AC, 9525.99.7040 Degas System Extension Panel Kit |
同系列其他配置
| 型号 | |
|---|---|
| 9525.99.10CCDG | Hach 9525sc DCCP System with Specific Conductivity (SC), Cation Conduc |
| 9525.99.0010DG | Hach 9525sc DCCP System, Degassed Conductivity only |
| 9525.99.0011 | Hach 9525sc DCCP System, Degassed Conductivity only, with Regenerative |
资料下载
- Ultra-Pure Water Measurements
- Introduction to Hach Services
- Care Advantage Flyer : Complete service solution for all your instruments
- Care Advantage Brochure : Complete service solution for all your instruments
- Complete Water Analysis Solutions for Power and Steam Generation
- Complete Water Analysis Solutions for Power and Steam Generation
- Contrôle de la qualité de l’eau dans les usines de valorisation énergétique des déchets
- Water quality at waste-to-energy plants
Recomendados similares
HACHConductivity Standard Solutions, 1000 µS/cm (491 mg/L NaCl)
HACHPolymetron Conductivity Module for SC200 Ultra Pure Controller
HACH1990 µS/cm Conductivity Standard; (995 mg/L TDS, 1000 mg/L NaCl), 1L
HACHRadiometer Analytical CDC565 Conductivity Cell (4-pole Pt sensor, k = 1.0 cm-1, epoxy body)
HACHCDW97-K10, Conductivity probe for salinity and brackish water
Solution, Conductivity Assembly
HACHRadiometer Conductivity Cells
HACH