Hach 9586 sc Oxygen Scavenger Analyzer, 100 - 240 V ACDescontinuado
HACH 9586.99.00P2
Online analyzer for monitoring hydrazine and carbohydrazide oxygen scavengers. An integral part of a complete water analytics system designed for the Power industry. A broad range of product options m

产品概述

Simple to Integrate. Simple to Operate.

Online analyzer for monitoring hydrazine and carbohydrazide oxygen scavengers.

An integral part of a complete water analytics system designed for the Power industry. A broad range of product options made to work together as flexible solutions meet your unique needs. The comprehensive approach saves you time on design, installation, training, maintenance, and operation.

主要特点

  • Save Time on Design:A single design source and one product platform means you spend less time searching for design files or configuring components. Create and reuse your optimal design templates.
  • Accelerate Your Installation:One source, interchangeable components, a common user interface, and one support team make installation faster and less complicated. Quickly and easily transfer user settings between oxygen scavenger analyzers.
  • Reduce Training Complexity:A single platform minimizes time required to teach and learn product operations, getting new systems in use faster.
  • Simplify Maintenance and Operation:Common menu guides reduce variability and provide step-by-step procedures for maintenance and calibration. Standard visual alerts across parameters notify operators when troubleshooting is required. The 9586 sc has a fast response time of less than 60 seconds. Unlike traditional amperometric techniques that use two electrodes, the Hach 9586 sc oxygen scavenger analyzer uses a three-electrode design; eliminating voltage drift due to the composition of the water. Self-cleaning electrodes reduce maintenance costs and analyser downtime via PTFE beads that prevent deposits on the electrode surfaces.

技术参数

参数名称说明
Operational ModePrimary or secondary measurement, calculated value (dual channel only) or timer/scheduler
Altitude< 2000 m
Analog Output Functional ModeLinear, Logarithmic, Bi-linear, PID
Analog OutputsTwo (Five with optional expansion module) 0/4 to 20 mA isolated current outputs, max. 550 Ω, Accuracy: ±0.1% of FS (20 mA) at 25 °C, ±0.5% of FS over -20 °C to 60 °C range
Calibration Methoda) Zero: electrically, with hydrazine-free water or with optional zero cartridge
b) Slope: using a laboratory reference value (e.g. LCW025)
CommunicationModbus RS232/RS485, Profibus DPV1, Hart optional
Communication CapabilitiesStandard (2x 4-20 mA outputs)
Conduit Openings1/2" NPT Conduit
Connection Drain Line6 x 8 mm (tubing must not exceed 4 feet and must drain straight down)
Connection Sample Line4 x 6 mm stainless steel tubing
DisplayGraphic dot matrix LCD with LED backlighting, transreflective
Display Resolution240 x 160 pixels
Display Size48 x 68 mm
Electrical CertificationsEMC CE compliant for conducted and radiated emissions: - CISPR 11 (Class A limits) - EMC Immunity EN 61326-1 (Industrial limits) Safety CAN/CSA C22.2 No. 61010-1 cETLus safety mark for: - General Locations per ANSI/UL 61010-1 & CAN/CSA C22.2. No. 61010-1
Enclosure RatingIP66 / NEMA 4X
Flow166 - 250 mL/min (10 - 15 L/h) recommended
Lower Limit of Detection (LOD)Drift is negligible; 1 ppb
Maintenance IntervalMonthly: Calibration and reagent refill
Manual LanguagesBulgarian, Chinese (PRC), Croatian, Czech, Danish, Dutch, English, Estonian, Finnish, French, German, Greek, Hungarian, Italian, Japanese, Korean, Lithuanian, Polish, Portuguese (Brazil), Portuguese (Portugal), Romanian, Russian, Slovak, Slovenian, Spanish, Swedish, Thai, Turkish
MaterialCounter-Electrode: Stainless steel; Working Electrode: Platinum
Material EnclosuresPolycarbonate Aluminium (powder coated) Stainless Steel
Number of Channels1
Operating Temperature Range5 - 45 °C at 0 - 95% RH (non-condensing)
Power Requirements (Voltage)100 - 240 VAC
Pressure Range0.5 - 6 bar (7.2 - 87 psi) or 12 L/h
Range0 - 100 ppb carbohydrazide; programmable
Range0 - 500 ppb hydrazine; programmable
Range 20 - 100 ppb carbohydrazide (also known as ELIMIN-OX); programmable ELIMIN-OX is a registered trademark of Nalco Chemical Co., Naperville, IL.
ReagentSample conditioning with diisopropylamine, ammonia, or diethylamine
Reference TypeAg/AgCl electrode with electrolyte KCl 0.1 M
Relative Humidity10 - 90 % relative humidity (non-condensing)
Relay FunctionsScheduler (Timer), Alarm, Feeder Control, Event Control, Pulse Width Modulation, Frequency Control, and Warning
RelaysFour electromechanical SPDT (Form C) contacts, 1200 W, 5 A
Repeatability± 2 % or 1 ppb (whichever is greater)
Response Time T90< 60 s
Sample RequirementsSample needs to be free of undissolved matter.
Sample Temperature5 - 45°C (41 - 113°F)
Security Levels2
Storage Conditions-20 to 70 °C
Temperature Compensation5 - 45 °C
Temperature Compensationto
Warranty12 months
Weight14.6 kg
What's included?Controller installed in Panel, Working Electrode, Reference Electrode, Counter Electrode, Diisopropylamine (99%), Basic User Manual