产品概述
Simple to Integrate. Simple to Operate.
Online analyzer for monitoring hydrazine and carbohydrazide oxygen scavengers.
An integral part of a complete water analytics system designed for the Power industry. A broad range of product options made to work together as flexible solutions meet your unique needs. The comprehensive approach saves you time on design, installation, training, maintenance, and operation.
主要特点
- Save Time on Design:A single design source and one product platform means you spend less time searching for design files or configuring components. Create and reuse your optimal design templates.
- Accelerate Your Installation:One source, interchangeable components, a common user interface, and one support team make installation faster and less complicated. Quickly and easily transfer user settings between oxygen scavenger analyzers.
- Reduce Training Complexity:A single platform minimizes time required to teach and learn product operations, getting new systems in use faster.
- Simplify Maintenance and Operation:Common menu guides reduce variability and provide step-by-step procedures for maintenance and calibration. Standard visual alerts across parameters notify operators when troubleshooting is required. The 9586 sc has a fast response time of less than 60 seconds. Unlike traditional amperometric techniques that use two electrodes, the Hach 9586 sc oxygen scavenger analyzer uses a three-electrode design; eliminating voltage drift due to the composition of the water. Self-cleaning electrodes reduce maintenance costs and analyser downtime via PTFE beads that prevent deposits on the electrode surfaces.
技术参数
| 参数名称 | 说明 |
|---|---|
| Operational Mode | Primary or secondary measurement, calculated value (dual channel only) or timer/scheduler |
| Altitude | < 2000 m |
| Analog Output Functional Mode | Linear, Logarithmic, Bi-linear, PID |
| Analog Outputs | Two (Five with optional expansion module) 0/4 to 20 mA isolated current outputs, max. 550 Ω, Accuracy: ±0.1% of FS (20 mA) at 25 °C, ±0.5% of FS over -20 °C to 60 °C range |
| Calibration Method | a) Zero: electrically, with hydrazine-free water or with optional zero cartridge b) Slope: using a laboratory reference value (e.g. LCW025) |
| Communication | Modbus RS232/RS485, Profibus DPV1, Hart optional |
| Communication Capabilities | Standard (2x 4-20 mA outputs) |
| Conduit Openings | 1/2" NPT Conduit |
| Connection Drain Line | 6 x 8 mm (tubing must not exceed 4 feet and must drain straight down) |
| Connection Sample Line | 4 x 6 mm stainless steel tubing |
| Display | Graphic dot matrix LCD with LED backlighting, transreflective |
| Display Resolution | 240 x 160 pixels |
| Display Size | 48 x 68 mm |
| Electrical Certifications | EMC CE compliant for conducted and radiated emissions: - CISPR 11 (Class A limits) - EMC Immunity EN 61326-1 (Industrial limits) Safety CAN/CSA C22.2 No. 61010-1 cETLus safety mark for: - General Locations per ANSI/UL 61010-1 & CAN/CSA C22.2. No. 61010-1 |
| Enclosure Rating | IP66 / NEMA 4X |
| Flow | 166 - 250 mL/min (10 - 15 L/h) recommended |
| Lower Limit of Detection (LOD) | Drift is negligible; 1 ppb |
| Maintenance Interval | Monthly: Calibration and reagent refill |
| Manual Languages | Bulgarian, Chinese (PRC), Croatian, Czech, Danish, Dutch, English, Estonian, Finnish, French, German, Greek, Hungarian, Italian, Japanese, Korean, Lithuanian, Polish, Portuguese (Brazil), Portuguese (Portugal), Romanian, Russian, Slovak, Slovenian, Spanish, Swedish, Thai, Turkish |
| Material | Counter-Electrode: Stainless steel; Working Electrode: Platinum |
| Material Enclosures | Polycarbonate Aluminium (powder coated) Stainless Steel |
| Number of Channels | 1 |
| Operating Temperature Range | 5 - 45 °C at 0 - 95% RH (non-condensing) |
| Power Requirements (Voltage) | 100 - 240 VAC |
| Pressure Range | 0.5 - 6 bar (7.2 - 87 psi) or 12 L/h |
| Range | 0 - 100 ppb carbohydrazide; programmable |
| Range | 0 - 500 ppb hydrazine; programmable |
| Range 2 | 0 - 100 ppb carbohydrazide (also known as ELIMIN-OX); programmable ELIMIN-OX is a registered trademark of Nalco Chemical Co., Naperville, IL. |
| Reagent | Sample conditioning with diisopropylamine, ammonia, or diethylamine |
| Reference Type | Ag/AgCl electrode with electrolyte KCl 0.1 M |
| Relative Humidity | 10 - 90 % relative humidity (non-condensing) |
| Relay Functions | Scheduler (Timer), Alarm, Feeder Control, Event Control, Pulse Width Modulation, Frequency Control, and Warning |
| Relays | Four electromechanical SPDT (Form C) contacts, 1200 W, 5 A |
| Repeatability | ± 2 % or 1 ppb (whichever is greater) |
| Response Time T90 | < 60 s |
| Sample Requirements | Sample needs to be free of undissolved matter. |
| Sample Temperature | 5 - 45°C (41 - 113°F) |
| Security Levels | 2 |
| Storage Conditions | -20 to 70 °C |
| Temperature Compensation | 5 - 45 °C |
| Temperature Compensation | to |
| Warranty | 12 months |
| Weight | 14.6 kg |
| What's included? | Controller installed in Panel, Working Electrode, Reference Electrode, Counter Electrode, Diisopropylamine (99%), Basic User Manual |
资料下载
- Hach 9586 sc Hydrazine Analyzer Calibration
- Introduction to Hach Services
- Complete Water Analysis Solutions for Power and Steam Generation
- Soluciones completas de análisis de agua para Generación de energía y vapor
- Solutions complètes d’analyse de l’eau pour La production de vapeur et d’électricité
- Complete Water Analysis Solutions for Power and Steam Generation
- Hach 9586sc CSI Specification PDF
- User Manual: Hach 9586sc Oxygen Scavenger Analyzer
Related Products
A
HACHReplacement Protector for Clark DO sensor
View →
HACHBench Service Partnership, LDO Model 2, 1 Service/Year
View →
HACHOrbisphere 510 Ozone Controllers AC Power
View →
HACHOrbisphere M1100 Oxygen (LDO) Sensors
View →
HACHOrbisphere 510 Oxygen Controllers Panel Mount
View →
HACHOrbisphere 51x Nitrogen Controllers
View →
HACHOrbisphere 510 Controller O₂ (EC), CO₂ (TC), Wall Mount, 100-240 VAC, 0/4-20mA, Ext. Press.
View →
HACHOrbisphere 3100 Portable Oxygen Analyzers
View →