Instrumentación · Ingeniería · Paquete · PLC
0755-82221901 sales@ainstru.com

EZ3008.9700140T

EZ3008.9700140T
Familia de producto: EZ sc Inorganics Analyzers
EZ3008sc Fluoride Ion-Selective Electrode (ISE) Analyzer, 1 stream, 4x mA
HACH EZ3008.9700140T
Fluoride Monitoring for Operational Stability

产品概述

Fluoride Monitoring for Operational Stability

"Stable fluoride control is essential in drinking water and surface-water treatment, but manual sampling can be labor intensive and introduce variability. The EZ3008sc automates fluoride measurement using discontinuous ISE analysis with temperature-controlled readings for dependable, repeatable data. Gel-filled ISE technology eliminates electrolyte refilling, reducing routine service and helping lower total cost of ownership.

With multi-stream capability, optional grab sampling for spot checks, and seamless integration into SCADA/PLC/DCS, the analyzer provides continuous results and diagnostics for confident compliance reporting and tighter process control.

Benefits: - Temperature-controlled, discontinuous ISE analysis for repeatable fluoride results - Gel-filled ISE electrodes eliminate electrolyte refilling - Multi-stream monitoring and optional grab sampling for spot checks - Reduced maintenance and service interventions - Integrates with SCADA/PLC/DCS for real-time data and diagnostics"

主要特点

  • Stable Performance in Changing Conditions:The EZ3008sc delivers temperature controlled, discontinuous fluoride measurements with multi stream capability. Operators get stable, reliable data during process fluctuations, reducing manual checks and adjustments while maintaining confident day to day water quality control.
  • Reduced Maintenance and Service Costs:Designed for long-term operation, the analyzer uses gel-filled ion-selective electrodes that eliminate the need for electrolyte refilling. This technology reduces reagent consumption, routine service interventions, maximizes analyzer uptime, and lowers the total cost of ownership over the instrument’s lifetime.
  • Flexible Connectivity for Legacy and Modern Platforms:Integrate seamlessly into existing PLC, SCADA, DCS and plant network architectures with a built-in industrial panel PC. Real-time data transmission provides immediate visibility into diagnostics and status, enabling faster responses and improved decision-making across your facility.
  • Flexible Analysis for Multiple Streams:Conduct quick spot checks and investigations without interrupting routine monitoring using built-in grab sampling. When combined with the optional 8-channel sample stream, one analyzer sequentially measures multiple process points: maximizing asset utilization and coverage while simplifying your system architecture.

技术参数

参数名称说明
Ambient Temperature50 - 86 °F (10 - 30 °C) ± 4 °C deviation at 5 - 95% relative humidity (non-condensing)
Analog OutputsActive 0 - 20 mA (or 4 - 20 mA) max. 500 Ohm load, standard 4, optional: 8
Automatic cleaningYes; Frequency freely programmable: 6 hours, 12 hours, daily, weekly
CalibrationAutomatic; 2-point; Frequency freely programmable: 6 hours, 12 hours, daily, weekly. Note: manufacturer recommends that a calibration is done when the reagents are replaced
CertificationsCE, ETL certified to UL and CSA safety standards, UKCA
Cycle TimeDefault: 5 minutes Continuous: 5 minutes
Demineralized WaterDoes not apply
Digital OutputsRelays: 5 contacts, not user configurable: Malfuntion, maintenance, analysis ready, sample ready, sample ready (EZ9150) Ethernet Connections: Optional: Claros Ethernet connection and Modbus TCP/IP Ethernet connector; LAN version; 10/100 Mbps or Profinet or Ethernet IP RS485 communication: Profibus DP or Modbus RTU
Dimensions (H x W x D)688 mm x 460 mm x 340 mm
DrainAtmospheric pressure, vented, min. Ø 32 mm
Earth ConnectionEarth connection Dry and clean earth pole with low impedance (< 1 Ohm) using an earth cable of > 2.5 mm²
Instrument AirDry and oil free according to ISA-S7.0.01-1996 quality standard for instrument air. Used to flush the instrument in corrosive environement. Min. 0.2 bar - Max. 0.5 bar
InterferencesMetal ions like aluminium [(Al)³+] > 72 mg/L, calcium [(Ca)²+] > 108 mg/L and iron [(Fe)²+/(Fe)³+] > 150 mg/L. Fats, oil, proteins, surfactants and tar.
Lower Limit of Detection (LOD)0.05 - 2.5 mg/L F-: 0.05 mg/L 0.1 - 5 mg/L F-: 0.1 mg/L 0.2 - 10 mg/L F-: 0.2 mg/L
MaterialHinged part: Thermoform ABS, Door: PMMA Wall section: Galvanized steel, powder coated
Measurement methodDiscontinuous, direct measurement by combined Ion-Selective Electrode
Measurement MethodIon-Selective Electrode
Number of sample streamsStandard: 1 Optional accessory for 2, 4, or 8 sample streams
ParameterFluoride
Power100 - 240 VAC, 50/60 Hz Max. power consumption: 120 VA
PrecisionBetter than 2% full scale range for standard test solutions
Protection ClassIP44
Range0.05 - 2.5 mg/L F- 0.1 - 5 mg/L F- 0.2 - 10 mg/L F-
Reagent RequirementsKeep between 50 - 86 °F (10 - 30 °C)
Sample Flow Rate100 - 300 mL/min
Sample PressureBy external overflow vessel (open to atmospheric pressure)
Sample QualityMaximum particle size 100 µm, < 0.1 g/L; Turbidity < 50 NTU
Sample Temperature50 - 86 °F (10 - 30 °C)
ValidationAutomatic; Frequency freely programmable: 6 hours, 12 hours, daily, weekly
Warranty12 months
WeightMax. 77 lb (35 kg)
What's included?EZ3008sc Fluoride Ion-Selective Electrode (ISE) Analyzer, Instruction Manual, 1 x Double Bit Door Key, 1 x Mounting Brackets, 1 x ISE Fluoride Electrode and 1 x empty 2.5L Reagent Container with Fittings (for Buffer Solution)

应用领域

Municipal Water

同系列其他配置

型号
EZ1012.9700140TEZ1012sc Cyanide Colorimetric Analyzer, 1 stream, 4x mA Output
EZ1035.9700140TEZ1035sc Silica Colorimetric Analyzer, 1 stream, 4x mA Output
EZ1036.9700140TEZ1036sc Sulfate Colorimetric Analyzer, 1 stream, 4x mA Output
EZ1037.9700140TEZ1037sc Sulfide Colorimetric Analyzer, 1 stream, 4x mA Output
EZ3005.9700140TEZ3005sc Chloride Ion-Selective Electrode (ISE) Analyzer, 1 stream, 4x
EZ3505.9700140TEZ3505sc Chloride Ion-Selective Electrode (ISE) Standard Addition Anal
EZ3508.9700140TEZ3508sc Fluoride Ion-Selective Electrode (ISE) Standard Addition Anal
EZ1016.9700140TEZ1016sc Hardness, total Colorimetric Analyzer, 1 stream, 4x mA Output
EZ1017.9700140TEZ1017sc Hardness, calcium Colorimetric Analyzer, 1 stream, 4x mA Outp

资料下载

10 modelos · clic para contraer