产品概述
By pairing Hach's Sumberged AV Sensor with our state-of-the-art Hach AV9000 Analyzer Module, you create a powerful Hach Submerged Area Velocity Meter (FL900AV) which provides "cleaner", more precise data than ever before.
Our Submerged AV Sensor (sold separately) is a robust, 1 MHz Acoustic Doppler velocity sensor designed to measure wastewater flow with improved accuracy and reliability. This tried-and-true sensor also uses a pressure transducer to measure flow level and incorporates advanced technologies to ensure precision, including automatically correcting for temperature and velocity effects on measurements. With advanced signal processing and filtering options, the AV9000 Analyzer Module expands the applicability of the sensor into more difficult applications. Plus, advanced analyzer diagnostics, including capture and display of the Doppler spectra, allow you to verify that the sensor is working properly even before you leave the site, giving you peace of mind.
主要特点
- Reliable, Budget-Saving Monitoring:The AV9000 Area Velocity Analyzer module is compensated for temperature, thus eliminating potential velocity errors of 2.7% over a 10ºC seasonal swing*. Its advanced multi-scale digital Doppler analysis provides the optimal combination of resolution and noise immunity. Mirror Image Processing™ eliminates sign errors and the advanced Target Set Processing™ reduces the impact of dominant targets (particles) in the stream to deliver a more representative velocity.
- Increase Monitoring Crew Safety:With the time saving features designed into the FL900 Flow Loggers, crews spend less time in the manhole and less time on site to decrease monitoring costs while increasing the safety of flow monitoring crews.
- Improved Accuracy:The AV9000 Area Velocity Analyzer module is compensated for temperature, thus eliminating potential velocity errors of 2.7% over a 10ºC seasonal swing*. Its advanced multi-scale digital Doppler analysis provides the optimal combination of resolution and noise immunity. Mirror Image Processing eliminates sign errors and the advanced Target Set Processing reduces the impact of dominant targets (particles) in the stream to deliver a more representative velocity.
- Easy Installation/Versatile Mounting Options:The logger can be quickly attached to a wall, pole or manhole ladder in minutes. Users can choose to hang logger from standard carabiner or optional 4-bolt wall mount for pole, horizontal or vertical wall mount or ladder rung mount.
技术参数
| 参数名称 | 说明 |
|---|---|
| Height | 400 mm |
| Accuracy | Timebase Accuracy: ± 0.002 % , synhchronized every 24 hours with server software and modem |
| Alarm | 16 maximum channels |
| Allowable Level | Standard: 10.5 m (34.5 ft) Extended: 31.5 m (103.5 ft) |
| Alternate KTO? | No |
| Battery Life | 296 days With Sub AV Sensor with AV9000 Analyzer - 4 Lantern Batteries - (no modem, room temperature) at 15 minute logging intervals |
| Certifications | Logger: CE, optional AC power supply: UL/CSA/CE |
| Communication | Local Communications: USB, RS232 (Baud rates: 9600, 19200, 38400, 57600, 115200) |
| Communication (optional) | Remote Communications (optional) Wireless modem: CDMA 1xRTT or GPRS |
| Data Storage | events maximum in non-volatile flash memory |
| Data Storage 2 | Sample History: 2000 sample events maximum in non-volatile flash memory |
| Data Storage 3 | Datalog: 325 k data points, 1128 days for 3 channels at 15-minute log intervals |
| Datalog Channels | maximum |
| Dimensions (H x W x D) | 400 mm x 254 mm x 220 mm |
| Environmental Rating | NEMA 6P |
| Level Accuracy | ±0.16% full scale ±1.5% of reading @ constant temp (±2.5ºC) ±0.20% full scale ±1.75% of reading from 0-30ºC (32-86ºF) ±0.25% full scale ±2.1% of reading from 0-70ºC (32-158ºF) |
| Level Measurement Method | Differential pressure transducer with stainless steel diaphragm and atmospheric pressure reference |
| Level Range | Standard: 0–3 m (0–10 ft) Extended: 0–9 m (0–30 ft) |
| Logging | 1, 2, 3, 4, 5, 6, 10, 12, 15, 20, 30 or 60 minutes. |
| Manual Languages | English |
| Material | PC ABS structural foam |
| Modem | Non-Wireless |
| Operating Temperature Range | -18 - 60 °C at 95% Relative Humidity |
| Power Requirements (Voltage) | 8 - 18 V DC |
| Pre-configuration? | AV9000 Module |
| Protocol | Local Modbus RTU |
| Recommended Range | -1.52 to 6.10 m/s (-5 to 20 ft/sec) |
| Sensor Connector | 1, with AV9000 |
| Sensor Connectors | 1, no aux |
| Sensor Ports | 1, 2 or 4 ports |
| Sensors Included | AV9000 Analyzer Module (Submerged AV Sensor sold separately) |
| Storage Conditions | -40 °C to 60 °C |
| Supported Sensors | Submerged Area Velocity Sensor |
| Typical Minimum Operating Depth | 2 cm (0.8cm) |
| Velocity Accuracy | ±2% of reading or 0.05 fps |
| Velocity Measurement Method | Doppler ultrasonic; twin 1 MHz piezoelectric crystals |
| Velocity-Induced Depth Error | Compensated based on flow velocity |
| Warranty | 12 months |
| Weight | Four batteries: 8.2 kg |
| Weight 2 | No batteries: 4.5kg |
同系列其他配置
| 型号 | |
|---|---|
| FL902AVUS.97 | Redundant-Level Flow Monitoring System |
资料下载
- Open Flow Measurement Trends
- Introduction to Hach Services
- Submerged Area Velocity Flow Meter
- FL900 Series Flow Logger
- Hach AV9000 Area Velocity Analyzer Module
- Full User Manual: FL900 Series Flow Logger
- Manuel d'utilisation: Enregistreur d'écoulement série FL900
- User Manual: FSDATA Desktop Instrument Manager
Recomendados similares
HACHSC1000 relay module for DIN-rail mounting, 4 Relays, max. 240V
HACHNP6000sc Phosphate Analyzers
Stainless Steel Chain Mount Kit
Mercury Thermometer, Celsius, -10 to 200°C
Mercury Thermometer, Fahrenheit, -30 to 120°F
Accu4 Spare Chamber with Plumbing, High Pressure
HACH