Instrumentación · Ingeniería · Paquete · PLC
0755-82221901 sales@ainstru.com

K1200-S00

K1200-S00
Orbisphere K1200 LDO Oxygen sensor, 0 - 2000 ppb, 28 mm Orbisphere fitting
HACH K1200-S00
The first maintenance-free optical oxygen sensor for nuclear power plants.

产品概述

The first maintenance-free optical oxygen sensor for nuclear power plants.

The Hach Orbisphere K1200 sensor is a high accuracy Luminescent Dissolved Oxygen (LDO) sensor that has been optimized for measurements in water processes in the power and steam industry. Designed for mechanical robustness to extend service life and optimize total cost of ownership, the sensor, combined with the Orbisphere Controllers, offers a new way to monitor dissolved oxygen in power plants. Based on good engineering practice the Orbisphere K1200 LDO sensor is suitable for nuclear plant application. Maintenance and operating costs are significantly reduced due to the absence of membrane and electrolyte and sensor accuracy is unaffected by process changes such as flow changes.

主要特点

  • One calibration per year:One zero point calibration per year is all that is needed with the K1200 sensor. Traditional electrochemical (EC) sensors display significant drift after only a few months, demanding regular re-calibration and substantial operator time. Due to its luminescent technology, the K1200 sensor is designed for minimal drift, resulting in it being the most stable sensor with the longest calibration interval in the nuclear industry.
  • No membranes = two minutes of maintenance:With no membranes to replace and no electrolyte solution to replenish, the K1200 is virtually maintenance-free requiring only two minutes of maintenance per year. Sensor accuracy is unaffected by process changes or low flow events with no polarization time, eliminating unnecessary operator interventions in radioactive zones. In addition, corrosive or hazardous chemicals are not required, making the annual task faster, easier and safer without reducing measurement precision.
  • Low cost retrofit:The complete system consists of a Controller, a flow chamber, the K1200 Luminescent Dissolved Oxygen Sensor and a 312xx Hydrogen Sensor for parallel measurement. The sensor is compatible with Hach Orbisphere 28 mm flow chambers, eliminating the need for engineering changes. Installation is fast and easy and does not require special preparation.
  • A new level of confidence:The K1200 optical sensor is the first to use luminescent measurement technology to measure low level oxygen in nuclear power plants. Since 1978, when the first patent on EC sensors was granted, Hach Orbisphere sensors have set the industry standard for oxygen measurement by delivering confidence to every water chemistry manager. The new K1200 maintains this tradition in the nuclear industry and offers significant operating and cost benefits.

技术参数

参数名称说明
Accuracy± 0.8 ppb or 2 % whichever is greater
Ambient Temperature-5 - 50 °C (23 - 122°F)
Analog Outputs3 Smart 0/4 to 20 mA (500 Ohms) programmable as linear or tri-linear, configurable to send diagnostics or alarm informations.
ApplicationNuclear
CalibrationSingle point zero calibration with standard 99.999% nitrogen (quality 5.0) or equivalent oxygen free gas
Certifications2004/108/EC - EN 61326-1
Communication Capabilities3 x 0/4-20 mA; RS485; Ethernet
Display Resolution0.1 ppb
Lower Limit of Detection (LOD)0.6 ppb
Mounting28 mm Orbisphere fitting
ParameterOxygen
Power Requirements (Voltage)100 - 230 VAC
Range0 - 2000 ppb dissolved O2 (DO)
RelaysMeasurement board: 3 measurement alarm relays (1A-30 VAC or 0.5A-50 VDC), configurable to send diagnostics information. Main board: 1 system alarm relay (1A-30 VAC or 0.5A-50 VDC).
Repeatability± 0.4 ppb or 1 % whichever is greater
Reproducibility± 0.8 ppb or 2 % whichever is greater
Response Time(90%) <10 s (gas phase); <30 s (liquid phase)
Sample Flow Rate50 - 300 mL/min
Sample Pressure1 - 20 bar absolute (14.5 - 290 psi)
Sample Temperature-5 - 50°C (23 - 122°F)
SensorsK1200 LDO
Storage Conditions-5 °C - 100 °C
Temperature RangeAccurate from -5 - 50 °C (23 - 122 °F) Resistant - temperature from -5 - 100 °C (23 - 212 °F)
Warranty12 months
Weight Sensor0.6
What's included?K1200 Sensor only. Controller 510, sensor cable, flow-chamber or process connection must be ordered separately.

应用领域

Boiler Water · Steam Cycle · Ultrapure Water

资料下载