产品概述
Allrounder probe for in pipe turbidity and suspended solids measurement in the industrial sector
The TSS sc TriClamp is a digital process probe for the extremely accurate in pipe measurement of turbidity and suspended solids in the industrial sector. It is installed by ball valve or weld-on fitting and can measure both on-line suspended solids and turbidity in one instrument. The two channel 90° scattered light turbidity method is in accordance with DIN EN ISO 7027. For the determination of suspended solids a scattered light measurement method at 120 ° is integrated. This eight channel measurement system, with an integrated bubble and temperature compensating software, enables the instrument to have a wide measuring range that effectively covers most applications, from the darkest pre-treated water to the freshest of spring waters. The TSS sc probes are compatible with the Hach SC200 and SC1000 controller platform. Furthermore, they are made of polished stainless steel with a scratch resistant and easy to clean sapphire window design to withstand harsh environments. Please contact Hach for further information. Request for a quote today!
主要特点
- Unique Multi-Beam Pulsed Infrared Light System:TSS sc probes have a double optical system with two pulsating infrared LEDs and four receivers. This combined multiple beam alternating light method with beam focusing facilitates accurate color independent measurement of turbidity and suspended solids.
- Automatic Compensation for Air Bubbles and Temperature Swings for Accuracy:Special software enables the system to recognize gas bubbles or temperature swings for a more accurate suspended solids or turbidity measurement.
- Standard Compliant and Precise - Ease of Calibration:The turbidity measurement complies with the standard DIN EN 27027 (ISO 7027). Calibration is not necessary. If the probe is used to measure the solids content, a one-point calibration suffices. For special applications, curves can be defined using several calibration points.
- Withstands Difficult Conditions and Harsh Environments:Probes are made of highly polished stainless steel with a scratch resistant sapphire window. They are design to withstand harsh environments and keep particles from sticking to the surface.
- Compatible with Hach SC controller platform:Measurement data is displayed and processed with the help of a digital Hach SC200 or SC1000 controller.
- Two parameters in one instrument--Suspended Solids and Turbidity:
技术参数
| 参数名称 | 说明 |
|---|---|
| Accuracy | Turbidity up to 1000 NTU: < 5% of the measured value ±0.01 NTU |
| Ambient Temperature | 0 - 60 °C (briefly 80 °C) |
| Application | Standard |
| Automatic Wiper | No |
| Cable Length | 10 m (optional extension cables available) |
| Calibration | Trubidity (TRB): Factory calibrated Solids (TS): To be calibrated by customer on site Zero point: Permanently calibrated in the factory |
| Calibration Method | Turbidity Formazin or Stablcal Standard (at 800 NTU). Requires a calibration kit. Suspended Solids Sample specific, based on gravimetric analysis with a correction factor procedure. |
| Controller Compatibility | SC200, SC1000, SC4500. All controllers sold separately |
| Diameter | 40 mm |
| Flow | Max. 3 m/s (the presence of air bubbles affects the measurement) |
| Includes | Turbidity & Suspended Solids sensor, user manual |
| Installation style | Weld On TriClamp Fitting |
| Length | 332 mm |
| Maintenance Interval | 1 h/month |
| Material | Body, Flange: Stainless Steel (1.4571, AISI 316Ti, UNS S31635), Optics: 1.4404, Gasket: FKM |
| Max Temperature | 60 |
| Measurement Method | Combined multiple beam alternating light method with infrared diode system and beam focusing Turbidity (TRB): 2-channel 90º scattered light measurement in accordance with DIN/EN 27027/ISO 7027, wavelength = 60 nm Solids (TS): 90º and 120º scattered light measurement, wavelength = 60 nm |
| Model | TSS sc |
| Mounting Configurations | TriClamp |
| Operating Temperature Range | 0 - 60 °C |
| Parameter | Turbidity, Suspended Solids |
| Pressure Range | < 10 bar or < 100 m |
| Range | Turbidity (TRB): 0.001 - 9999 NTU Solids (TSS): 0.001 - 500 g/L |
| Repeatability | TSS content: < 4 % Turbidity: < 3 % |
| Response Time | 1 s < T90 < 300 s (adjustable) |
| Response Time T90 | 1 - 300 s adjustable |
| Special Features | With Triclamp |
| Special Notes | Installation note: Distance sensor–wall > 50 cm (Turbidity) > 10 cm (TSS) |
| Warranty | 12 months |
| Weight | Approx. 1.6 kg |
| What's included? | Turbidity & Suspended Solids sensor, user manual |
应用领域
Pre-Treatment
同系列其他配置
| 型号 | |
|---|---|
| LXV323.99.30002 | TSS sc Turbidity and Suspended Solids Inline Probe, Stainless Steel |
| LXV324.99.30002 | TSS W sc Turbidity and Suspended Solids Inline Probe with Wiper, Stain |
| LXV323.99.10002 | TSS sc Turbidity and Suspended Solids Immersion Probe, Stainless Steel |
| LXV324.99.10002 | TSS W sc Turbidity and Suspended Solids Immersion Probe with Wiper, St |
| LXV324.99.20002 | TSS W sc Turbidity and Suspended Solids TriClamp Insertion Probe with |
资料下载
- Case Study: SOLITAX Sensors, LDO Probes; Accurate On-Line DO and TSS Monitoring Helps Optimize Petrochem Effluent Treatment
- Beer Recovery from Surplus Yeast Using the TSS Vari sc by Hach
- Measuring Turbidity and Suspended Solids in Power Plants
- Advanced On-Line Instrumentation Helps DAF Systems Lower Costs
- Introduction to Hach Services
- Complete Water Analysis Solutions : Upstream and Oil & Gas Industry
- Complete Water Analysis Solutions - Midstream and Liquefied Natural Gas Production
- Care Advantage Flyer : Complete service solution for all your instruments
Recomendados similares
HACHSonatax sc Sensor
HACHGelex Secondary Turbidity Standard, Stray Light
HACHTSS Portable Hand-held Turbidity, Suspended Solids, and Sludge Level System
HACHGasket for mounting flange, Solitax sc, TSS sc
HACHSolitax sc Sensor Online Course
HACHTurbidity Nephelometric Technology Online Course
HACHWelded branch DN125 for TSS sc TriClamp
HACH