Instrumentación · Ingeniería · Paquete · PLC
0755-82221901 sales@ainstru.com

LXV328.99.20002

LXV328.99.20002
Familia de producto: TSS EX1 sc Suspended Solids Sensors for Hazardous Locations
TSS EX1 sc Turbidity and Suspended Solids TriClamp Insertion Probe, Stainless Steel
HACH LXV328.99.20002
Turbidity and suspended solids in pipe probe specially designed for use in potentially explosive atmospheres (Class I Div 2)

产品概述

Turbidity and suspended solids in pipe probe specially designed for use in potentially explosive atmospheres (Class I Div 2)

The TSS EX1 sc TriClamp is a digital process probe for the extremely accurate in pipe measurement of turbidity and suspended solids specially designed for use in potentially explosive atmospheres (Class I Div 2). This in pipe probe is installed by ball valve fitting and can measure both on-line suspended solids and turbidity in one instrument. The 2 channel 90° scattered light turbidity method is in accordance with DIN EN ISO 7027. For the determination of suspended solids a scattered light measurement method at 120 ° is integrated. This eight channel measurement system, with an integrated bubble and temperature compensating software, enables the instrument to have a wide measuring range. The TSS EX1 sc probes are compatiblewith the Hach SC200 and SC1000 controller platform. Furthermore, they are made of polished stainless steel with a scratch resistant and easy to clean sapphire window design to withstand harsh environments. Please contact Hach for further information. Request for a quote today!

主要特点

  • Unique Multi-Beam Pulsed Infrared Light System:TSS EX1 sc probes have a double optical system with two pulsating infrared LEDs and four receivers. This combined multiple beam alternating light method with beam focusing facilitates accurate color independent measurement of turbidity and suspended solids.
  • Automatic Compensation for Air Bubbles and Temperature Swings for Accuracy:Special software enables the system to recognize gas bubbles or temperature swings for a more accurate suspended solids or turbidity measurement.
  • Standard Compliant and Precise - Ease of Calibration:The turbidity measurement complies with the standard DIN EN 27027 (ISO 7027). Calibration is not necessary. If the probe is used to measure the solids content, a one-point calibration suffices. For special applications, curves can be defined using several calibration points.
  • Withstands Difficult Conditions and Harsh Environments:TSS EX1 sc probes are made of highly polished stainless steel with a scratch resistant sapphire window. They are design to withstand harsh environments and keep particles from sticking to the surface.
  • Suitable for Use in Potentially Explosive Atmospheres:TSS EX1 sc probes have been specially designed for measurement in hazardous locations with Class I Div 2 requirements.

技术参数

参数名称说明
AccuracyTurbidity up to 1000 FNU/NTU: < 5 % of measurement value or ±0.01 NTU, whichever is greater
Ambient Temperature0 - 50 °C
ApplicationHazardous locations
Automatic WiperNo
Cable Length10 m
CalibrationTurbidity (TRB): Factory calibrated Solids (TS): To be calibrated by customer on site Zero point: Permanently calibrated in the factory
Calibration MethodTurbidity Formazin or Stablcal Standard (at 800 NTU). Requires a calibration kit. Suspended Solids Sample specific, based on gravimetric analysis with a correction factor procedure.
Controller CompatibilitySC200, SC1000, SC4500. All controllers sold separately
Diameter48.5 mm
FlowMax. 3 m/s (the presence of air bubbles affects the measurement)
IncludesTurbidity & Suspended Solids sensor, user manual
Installation styleBall Valve Insertion
Length448 mm
Maintenance Interval1 h/month
MaterialBody, Flange: Stainless Steel (1.4571, AISI 316Ti, UNS S31635), Optics: 1.4404, Gasket: FKM
Max Temperature50
Measurement MethodCombined multiple beam alternating light method with infrared diode system and beam focusing. Turbidity (TRB): 2-channel 90º scattered light measurement in accordance with DIN EN ISO 7027, wavelength = 860 nm Solids (TSS): 90º and 120º scattered light measurement, eight channel multi-angle measurement system, with an integrated bubble and temperature compensating, wavelength = 860 nm
ModelTSS EX1 sc
Mounting ConfigurationsTriClamp
Operating Temperature Range-10 - 50 °C
ParameterTurbidity, Suspended Solids
Pressure Range 10 bar or 100 m
RangeTurbidity (TRB): 0.001 - 9999 NTU Solids (TSS): 0.001 - 500 g/L
RepeatabilityTSS content: < 4 % Turbidity: < 3 %
Response Time1 s < T90 < 300 s (adjustable)
Response Time T901 - 300 s adjustable
Special FeaturesWith Triclamp
Warranty12 months
Weightapprox. 2.7 kg
What's included?Turbidity & Suspended Solids sensor, user manual

同系列其他配置

型号
LXV328.99.30002TSS EX1 sc Turbidity and Suspended Solids Inline Probe, Stainless Stee
LXV328.99.10002TSS EX1 sc Turbidity and Suspended Solids Immersion Probe, Stainless S

资料下载

3 modelos · clic para contraer