HI6321

Medidor de conductividad y resistividad de sobremesa

Medidor de conductividad de sobremesa de nueva generación

Measurement

  • Measurement of μS/cm, mS/cm (conductivity); Ω·cm, kΩ·cm and MΩ·cm (resistivity); ppm, ppt (TDS); ppt, PSU, % (salinity) as a function of temperature

  • Application-specific profiles allow fast and direct measurement without updating sensor and system settings

  • Activity log during measurement

  • Measurement stability indicator (set using stability criteria)

  • Reading modes: direct and direct/auto-hold

  • Temperature compensation can be set to automatic or manual

  • Audible and/or alarm messages when measurement exceeds predefined limits

  • Galvanic isolation for conductivity measurement


  • The HI6321 is a streamlined benchtop instrument with a large touchscreen display, consisting of the instrument housing and an integrated conductivity measurement module.

  • The HI6321 is compact and easy to operate, and includes Hanna's HI7631233 four-ring conductivity/resistivity/TDS/salinity probe.

  • The HI7631233 is recommended for water applications in a wide range of industrial processes. It provides stable measurement over a wide measuring range and does not require frequent calibration. The integrated temperature sensor measures the process temperature and adjusts the measured conductivity to the reference temperature by applying specialized compensation algorithms:

  • Linear: Applicable when the temperature coefficient of variation is assumed to be the same value for all measured temperatures.

  • Standard: Suitable for high-purity water measurement and documented in ASTM standard D5391-14. This setting should be used for resistivity measurements greater than 1 Mohm.cm.

  • Natural: Suitable for natural groundwater, well water or surface water (or water of similar composition) in accordance with ISO7888.

  • TDS is a calculated value based on the conductivity of the solution (TDS = coefficient x EC25). The TDS coefficient is the conversion factor used to change the EC measurement value to a ppm measurement value.

  • Sal psu: The practical salinity of seawater is related to the ratio of the conductivity of a normal seawater sample at 15°C and 1 atmosphere to that of a potassium chloride solution (KCl) with a mass of 32.4356 g/Kg at the same temperature and pressure. Under these conditions, this ratio is equal to 1 and S = 35. At temperatures from 0 to 35°C, the practical salinity scale is applicable to values from 0 to 42.00 psu.

  • Sal ppt: The measured value expressed in ppt is based on the natural seawater scale from 0.00 to 80.00 g/L, covering a temperature range of 10 to 31°C. It determines salinity based on the conductivity of the sample relative to standard seawater at 15°C and a salinity value of approximately 35 in seawater.

  • Sal%: In this range, 100% salinity corresponds to approximately 10% solids.

  • User Interface

  • 7-inch capacitive touchscreen with multi-touch support

  • Capacitive touchscreen, home and system menu keys

  • User-friendly icons and symbols allow users to easily navigate and interpret instrument functions.

  • Users can choose between five different views:

  • Simple GLP with calibration information

  • Full GLP with electrode status and calibration point details

  • Interactive graph with real-time updates

  • Tabular data including date, time and comments

  • Calibration

  • Standard single-point salinity calibration with 100% salinity standard, reading salinity scale set to %

  • Standard single or multiple conductivity calibration using standards

  • Non-volatile memory retains data and settings

  • Logging

  • Data log collection of at least 1000000 data points (with time and date stamp)

  • Logging types: manual, automatic, auto-hold

  • Sample ID for manual and auto-hold data

  • Connectivity and Services

  • Transfer logged data to USB drive

  • Includes log files of measurement and calibration data (as .csv files)

  • FTP and email for exporting logs via Ethernet and Wi-Fi connection

  • USB Type A for USB flash drive, keyboard and printer

  • USB Type C for USB flash drive and PC connection

  • Instrument Guided Help Section

  • Video-supported demonstrations of main functions


Conductivity EC Measuring range

0.000 to 9.999 μS/cm, 10.00 to 99.99 μS/cm, 100.0 to 999.9 μS/cm, 1.000 to 9.999 mS/cm, 10.00 to 99.99 mS/cm, 100.0 to 1000.0 mS/cm

Resolution

0.001 μS/cm, 0.01 μS/cm, 0.1 μS/cm, 0.001 mS/cm, 0.01 mS/cm, 0.1 mS/cm

Accuracy ±1% of reading (±0.01μS/cm)
Resistivity Measuring range

1.0 to 99.9 Ω•cm; 100 to 999 Ω•cm; 1.00 to 9.99 KΩ•cm; 10.0 to 99.9 KΩ•cm; 100 to 999 KΩ•cm; 1.00 to 9.99 MΩ•cm; 10.0 to 100.0 MΩ•cm

Resolution 0.1 Ω•cm; 1 Ω•cm; 0.01 KΩ•cm; 0.1 KΩ•cm; 1 KΩ•cm; 0.01 MΩ•cm; 0.1 MΩ•cm
Accuracy ±1% of reading (±1 Ω•cm)
TDS Measuring range 0.000 to 9.999 ppm, 10.00 to 99.99 ppm, 100.0 to 999.9 ppm, 1.000 to 9.999 ppt, 10.00 to 99.99 ppt, 100.0 to 400.0 ppt  (conversion factor 1.00 )
Resolution

0.001 ppm, 0.01 ppm, 0.1 ppm, 0.001 ppt, 0.01 ppt, 0.1 pp

Accuracy

±1% of reading (±0.01 ppm)

Salinity Measuring range Practical salinity: 0.00 to 42.00 psu; Seawater salinity: 0.00 to 80.00 ppt; Salinity percentage: 0.0 to 400.0%
Resolution

0.01 (practical salinity, seawater salinity); 0.1% salinity

Accuracy

±1% of reading

Temperature Measuring range

-20.0 to 120.0˚C

Resolution 0.1 °C
Accuracy

±0.2 °C

Conductivity calibration

Calibration points

Auto recommended / user-defined

Standard calibration solutions

84.00μS/cm,1.413mS/cm,5.000mS/cm,12.88mS/cm,80.00mS/cm,111.8mS/cm

Reminder

Off function

Daily: 0 minutes to 23 hours 59 minutes.

Periodic: 1 minute to 500 days, 23 hours 59 minutes

Resistivity calibration

 

Via conductivity calibration

Salinity calibration

 

1 calibration point (percentage)

Temperature compensation

 

Automatic or manual

View

Type

Direct/auto-hold

  Stability

Accurate; Medium; Fast

 

Sampling rate

1000 ms

Conductivity view

Primary

Measurement (EC, resistivity, TDS, salinity, temperature) stability status

Simple GLP

Basic view information

Last calibration data, electrode offset

Full GLP

Simple GLP information and detailed calibration points

Table

Measurement values displayed in a table at one-second intervals

Graph (plot)

pH (or mV) and temperature versus time graph can be panned or zoomed (zoom technology)

Logging

Type

Automatic, manual logging, auto-hold

Number of records

Maximum 50000 per file

Minimum 1000000 data points stored per user

Auto interval 1, 2, 5, 10, 30 seconds
   

1, 2, 5, 10, 15, 30, 60, 120, 150, 180 minutes

 

Sample ID

Increment mode

 

Export options

.csv file format

Users

 

Up to 9 users and administrator. Accounts (default)

Connectivity

USB-A

2 ports for keyboard input or USB-A drive

 

USB-C

1 port for PC connection and USB Type-C drive

 

Wi-Fi & Ethernet

FTP

Web server log transfer and download

Email

 

RS232

Connect peripherals

Power supply

 

DC adapter 100-240AC, 24 VDC 2.5A

Operating environment

 

0 - 50.0 °C

Dimensions/Weight

 

205 x 160 x 77mm / approx. 1.2kg

Ordering information

 

HI6321 meter, conductivity electrode HI7631233; electrode holder HI764060, capillary pipette; 24VDC power adapter; USB-C to USB-A cable; quick start guide with quality certificate

Tratamiento de Aguas Residuales Municipales

Tratamiento de Aguas Residuales Municipales — Hanna: caso de configuración desde stock

Resumen del proyecto

La instalación está en las etapas de entrada, biológica y salida de una planta municipal. El cliente quería configurar bien Hanna en la planta existente, con el objetivo de análisis portátil y de laboratorio con instrumentos Hanna.

Retos técnicos

Lo que realmente da problemas es alto contenido de sólidos y componentes corrosivos, con grandes variaciones diarias de caudal, y los parámetros del efluente oscilan y las sondas se ensucian con lodos y biopelícula. Todo ello implica que la configuración de Hanna deba incluir selección por parámetros, precisión y portabilidad.

Configuración

El suministro y la instalación se organizaron por la vía de stock: configurar desde stock y entregado con rapidez. En planta esto supuso selección por parámetros, precisión y portabilidad; se revisó línea a línea: cumplimiento estable del nivel 1A y reporte de datos en línea.

Comportamiento

Tras la puesta en marcha el resultado es resultados in situ, con menos verificaciones de laboratorio, y las lecturas se usan ya para ajustar el proceso.

Alimentación y Bebidas

Alimentación y Bebidas — Hanna: caso de configuración desde stock

Situación en planta

Se trata de un cliente del sector Alimentación y Bebidas cuya planta está en las etapas de dosificación, pasteurización, llenado y limpieza CIP. La tarea principal era análisis portátil y de laboratorio con instrumentos Hanna.

El problema

La dificultad está en jarabes, lácteos y bebidas proclives a incrustaciones, con estrictos requisitos higiénicos; a la vez, las limpiezas frecuentes exigen instrumentos resistentes a ciclos CIP/SIP con sosa caliente. Por eso la especificación implica selección por parámetros, precisión y portabilidad.

Qué hicimos

Adoptamos un enfoque de stock: configurar desde stock y entregado con rapidez. La entrega se construyó en torno a selección por parámetros, precisión y portabilidad; se verificó punto por punto: requisitos de seguridad alimentaria y diseño higiénico 3-A / EHEDG.

Resultado operativo

En servicio el sistema responde a lo previsto: resultados in situ, con menos verificaciones de laboratorio, con un mantenimiento rutinario notablemente menor.

Química y Farmacéutica

Química y Farmacéutica — Hanna: caso de configuración desde stock

Contexto de aplicación

reactores, vasijas de dosificación y tanques de producto terminado es un servicio típico de Química y Farmacéutica. El proyecto partió de Hanna, centrándose en análisis portátil y de laboratorio con instrumentos Hanna.

Limitaciones del proceso

la corrosión y la protección Ex acortan la vida útil y encarecen las paradas, y medios muy corrosivos con disolventes y cristalización, en parte en atmósferas explosivas no hace sino acentuarlo. Por eso la especificación de Hanna debe cubrir selección por parámetros, precisión y portabilidad.

Enfoque de suministro

La entrega se gestionó como stock: configurar desde stock y entregado con rapidez. La ejecución se centró en selección por parámetros, precisión y portabilidad, junto con certificación Ex y trazabilidad de datos según GMP.

Resultados en campo

La operación evidencia el beneficio: resultados in situ, con menos verificaciones de laboratorio, con una confianza claramente mayor en este punto de medición.

Energía y Electricidad

Energía y Electricidad — Hanna: caso de configuración desde stock

Perfil del cliente

El entorno típico en Energía y Electricidad es agua de alimentación, condensado de turbina y sistemas de desulfuración/DeNOx. El punto de medición es Hanna; el foco es análisis portátil y de laboratorio con instrumentos Hanna.

Dificultades clave

Las condiciones son alta temperatura y presión, con vapor y agua salina y, encima, deriva a alta temperatura y presión, con paradas no programadas muy costosas, algo que una configuración estándar no soporta mucho tiempo. El diseño parte de selección por parámetros, precisión y portabilidad.

Implementación

El alcance se canalizó por la vía de stock: configurar desde stock y entregado con rapidez. Técnicamente se apoya en selección por parámetros, precisión y portabilidad, y en estos requisitos: cumplimiento de normativas del sector y objetivos de disponibilidad.

Beneficio para el cliente

En la práctica esto significa resultados in situ, con menos verificaciones de laboratorio, lo que redujo la carga del equipo de mantenimiento.