Home    analyzer    conductivity    CLD134 Conductivity compact device
  • Field of application
    Smartec CLD134 is a measuring system for inductive conductivity in the food & beverage and life sciences industries. The combination of a transmitter and a sensor is interference-free and especially easy-to-use. It meets the highest hygienic and sterile demands thanks to its food-grade virgin PEEK body, its jointless and creviceless design and its hygienic certificates. Smartec CLD134 is therefore the ideal package to ensure the highest safety and quality of your products and processes.
    The hygienic Smartec CLD134 measures toroidal conductivity in food & beverage and life science applications:
    Phase separation of product/water and product/product mixtures in pipe systems
    Control of CIP processes in the return channel
    Concentration control in remaking of CIP cleaning agents
    Product monitoring in pipe systems, bottling plants, quality assurance
    Leakage monitoring
    It features the following protocols and interfaces:
    0/4...20mA
    HART
    PROFIBUS DP
    PROFIBUS PA

    Benefits
    Unique hygienic design avoids recontamination
    With all hygienic certificates required in hygienic and sterile applications
    Certified biocompatibility according to USP class VI
    Compliant with EG 2023/2006 and 1935/2004
    Suitable for cleaning-in-place (CIP) and sterilization-in-place (SIP)
    Durable thanks to encapsulated, jointless design

     

    Conductivity:

    Measuring Principle Inductive
    Application Food, beverage, dairies, breweries and pharmaceutical industries, biotechnology
    Characteristic Specifically intended for use in hygienic applications and especially suitable for following applications:
    Phase separation, control of CIP, concentration control, product monitoring, leakage monitoring
    Measuring range 100 µS/cm to 2000 mS/cm
    Cell constant k : 6.3 1/cm
    Measuring principle Inductive conductivity measurement
    Design Hygienic design with FDA certified virgin PEEK body including all hygienic certifications
    Material In contact with medium: Virgin PEEK
    Process temperature Max. 125 °C (max. 257 °F)
    For sterilization: up to 150 °C for max. 60 min
    (up to 302 °F for max. 60 min)
    Process pressure 12 bar abs up to 90 °C
    (174 psi abs up to 194 °F)

     
    Temperature sensor Pt 1000
    Connection Dairy pipe, aseptic-fitting, Clamp ISO 2852, SMS-2", Varivent N DN 40-125, Neumo Biocontrol D50
    Ingress protection IP67
    Output / communication Conductivity, concentration and temperature. 0/4 to 20 mA, HART, Profibus
    Additional certifications FDA, EHEDG, 3A
    Biological reactivity USP class VI part <87> and <88>
    EC Regulation No. 1935/2004

CLD134 Conductivity compact device

Toroidal conductivity measuring system for hygienic and sterile applications
CLD134 Conductivity compact device

everyone is watching: