Приборы · Инжиниринг · Комплектация · ПЛК
0755-82221901 sales@ainstru.com
LXV525.97KTO039
Семейство продуктов: TU5400sc Turbidimeter + SC4500 Controller
TU5400sc Ultra-High Precision Low Range Laser Turbidimeter, EPA, Auto Cleaning, System Check, RFID, Flow, SC4500, 2 Channels
HACH LXV525.97KTO039
The next standard in the evolution of turbidity

产品概述

The next standard in the evolution of turbidity

Compliant with EPA approved Hach Method 10258 and perfect for membrane filtration applications, the TU5400sc employs a patented optical design that sees more of your sample than any other turbidimeter, delivering the best low level precision and sensitivity while minimizing variability between measurements. For the first time you will be able to remove uncertainty of which measurement to trust between your lab and process instruments, thanks to identical 360° x 90° Detection® Technology in any TU5 Series® turbidimeter. The TU5400sc Ultra-High Precision Low Range Laser Turbidimeter reduces the time needed to get a turbidity measurement you can trust. With a 10mL sample size, response time is over 15 times faster than an FT660. A stable laser light source eliminates the need for annual lamp replacements. The turbidimeter’s vial surface area has been reduced by 98%, allowing you to save time cleaning. With a mounting pattern identical to the FT660, replacing your previous turb is fast and simple. All of this, along with the ability to measure to 0.0002 NTU, gives you the next standard in the evolution of turbidity. A flow meter is included with this turbidimeter that continuously measures flow through the instrument and displays the flow value on screen. Current flow can be output to SCADA via 4-20mA or digital communications. Ensure that your turbidimeter is always functioning properly with this TU5's System Check capability. With the ability to display measurement status and vial cleanliness, you can have full confidence in your turbidity measurement. RFID technology included with this turbidimeter delivers easy calibration and simple, paperless turbidity measurement comparisons between your online turbidimeter and the TU5200. Cleaning your turbidimeter takes valuable time that your staff could use to complete other tasks. The TU5 Series' automatic cleaning module that is included with this instrument completes cleaning for you, saving operator time and money.

主要特点

  • Groundbreaking 360° x 90° Detection™ Technology:The TU5 Series employs a patented optical design that sees more of your sample than any other turbidimeter, delivering the best low level precision and sensitivity while minimizing variability from test to test.
  • Everything about turbidity – faster:The TU5 Series dramatically reduces the time needed to get a turbidity measurement you can rely on, with 98% less online sample surface area to clean, sealed vials for calibration, and the elimination of the need for indexing and silicone oil in the lab. Not to mention, a smaller online sample volume means you will detect events almost immediately.
  • Easy Adoption:The familiar experience of a modern touchscreen, the ability to use your current Hach sensors, and the same footprint as the SC200, make installation and integration of the SC4500 Controller seamless.
  • No Time for Downtime:The SC4500's built-in predictive diagnostic software ensures measurement confidence and reduces the risk of unexpected equipment downtime by enabling proactive maintenance planning via MSM, including step-by-step instructions.

技术参数

参数名称说明
Accuracy±2% or 0.01 NTU from 0 - 40 NTU ±10% of reading from 40 - 1000 NTU based on Formazin primary standard
Cable Length1.6 m (5.25 ft), extendable up to 50 m (164 ft) for instrument without accessories
Calibration MethodFor Formazin and Stablcal®: 20 NTU from 0 to 40 NTU; at 20 FNU and 600 NTU for full range Custom calibration for up to 6-point calibrations
CertificationsCE compliant US FDA accession number: 1420493-000 EPA version, 1420492-000 ISO version Australian ACMA Marking
CommunicationSystem Check & RFID
Controller2 Channels, 100-240 VAC
Dimensions (H x W x D)249 mm x 268 mm x 190 mm
Enclosure RatingElectronic compartment IP55; all other functional units IP65 with process head/ACM attached to the TU5300sc/TU5400sc instrument
Fitting TypeSample quick connector: ¼-in. for ¼-in. tubing
InstrumentWith Flow Sensor and Automatic Cleaning Module
Light SourceClass 2 laser product, with embedded 650 nm (EPA 0.43 mW) or Class 1 laser product, with embedded 850 nm (ISO), max. 0.55 mW (complies with IEC/EN 60825-1 and to 21 CFR 1040.10 in accordance with Laser Notice No. 50)
MaterialASA Luran S 777K / RAL7000, TPE RESIN Elastocon® STK40, Thermoplastic Elastomer TPS-SEBS
ModelTU5400sc
Mounting ConfigurationsWall Mount
Operating HumidityRelative humidity: 5 - 95% at different temperatures, non-condensing
Operating Temperature Range0 - 50 °C (32 - 122 °F)
OptionsProcess Head, Flow Sensor, & Automatic Cleaning Module
RangeEPA: 0 - 700 NTU / FNU / TE/F / FTU 0 - 100 mg/L 0 - 175 EBC ISO: 0 - 1000 NTU / FNU / TE/F / FTU 0 - 100 mg/L 0 - 250 EBC
RegulatoryEPA
RepeatabilityBetter than 1% of reading or ±0.002 NTU (TU5300sc) or ±0.0006 NTU (TU5400sc) on Formazin at 25 °C (77 °F), whichever is greater
Resolution0.0001 NTU / FNU / TE/F / FTU / EBC
Response TimeTU5300sc: T90 <45 seconds at 100 mL/min TU5400sc: T90 <30 seconds at 100 mL/min
Sample Flow Rate100 - 1000 mL/min; optimal flow rate: 200 - 500 mL/min
Sample Pressure6 bar (87 psi) maximum, compared to air at sample temperature range from 2 - 40 °C (35.6 - 104 °F)
Sample Temperature2 - 60 °C (35 - 140 °F)
Signal Average TimeTU5300sc: 30 - 90 seconds TU5400sc: 1 - 90 seconds
Storage Conditions-40 - 60 °C (-40 - 140 °F)
Stray Light<10 mNTU
UnitsNTU, FNU, TE/F, FTU, EBC
VerificationLiquid: Stablcal®, Formazin (0.1 - 40 NTU) Dry: Glass Rod at <0.1 NTU
Weight5.95 lbs. (2.7 kg); 11 lbs. (5.0 kg) with all accessories
What's included?Controller, Non-Claros + Embedded Prognosys, without power cord; includes mounting hardware

应用领域

Filtration · Membrane Protection · Regulations · Water Reuse

同系列其他配置

型号
LXV525.97KTO038TU5400sc Ultra-High Precision Low Range Laser Turbidimeter, EPA Versio

资料下载

Моделей: 2 · свернуть