Product Overview

  • HI1414D/50 is a glass body, single reference junction pH electrode made with low temperature glass and a DIN connector. This preamplified electrode features a built-in temperature sensor for temperature compensated measurements in a single probe design.

  • The electrode has a unique open junction design in which there is a viscolene (hard gel) electrolyte layer that is free of silver chloride (AgCl) between the sample to be measured and the internal reference cell. The open junction design resists clogging and the probe’s flat tip makes it ideal for pH measurements on surfaces such as skin, leather, and paper. The HI1414D/50 is recommended for use with Hanna’s HI99181 pH meter.

  • Open junction

  • Flat tip

  • Built-in temperature sensor

  • Hanna Instruments offers a wide variety of pH electrodes that are designed for many different applications. The type of glass used for sensing pH, bulb shape, body material, type of junction, type of reference and electrolyte used are just some of the design considerations.

  • The HI1414D/50 uses low temperature (LT) glass, flat tip bulb, glass body, and open junction with viscolene gel electrolyte.

Image 1 Details:

  • Low Temperature Glass Formulation

  • The glass tip uses a special LT glass formulation with a resistance of about 50 megohms compared to general purpose (GP) with a resistance of about 100 megohms. This is beneficial since many food products are stored at low temperatures. As the temperature of the glass in the sample decreases, the resistance of the LT glass will increase approaching that of GP glass at ambient temperature. If GP glass were used, the resistance would increase to above the optimum resistance for the high impedance input of a pH meter. The HI1414D/50 is suitable to use with samples that measure from 0 to 50°C.

,

Image 2 Details:

  • Flat Glass Tip

  • A flat tip is recommended to optimize surface contact with skin, leather, paper, and samples that are not easily penetrated.

Image 3 Details:

  • Glass Body

  • The glass body is ideal for laboratory use. The glass is resistant to many harsh chemicals and is easily cleaned. The glass body also allows for a fast transfer of heat to the internal reference electrolyte. The mV generated by the reference cell is temperature dependent. The faster the electrode reaches equilibrium, the more stable the reference potential.

Image 4 Details:

  • Open Junction Reference

  • Suspended solids and proteins found in food products will clog a conventional ceramic reference junction. This clogging will impede the measurement circuit between the indicating electrode and the internal reference resulting in slower response time, erratic readings, and frequent electrode replacement. The open junction design consists of a solid gel interface (viscolene) between the sample and internal Ag/AgCl reference. This interface not only prevents silver from entering the sample, but also makes it impervious to clogging, thus providing fast response and stable readings.

Image 5 Details:

  • DIN Connector

  • The HI1414D/50 uses a DIN connector. This type of connector is generally proprietary to the meters they are supplied with and may not be interchangeable. The HI1414D/50 is recommended for use with Hanna’s HI99181 pH meter. Other types of connectors include BNC, screw-type, T-type, and 3.5mm to name a few.

SKU HI1414D/50
Product Name Flat Tip pH Electrode for Skin and Scalp with DIN Connector - HI1414D/50
Quote Required Yes
Measurement Range pH
Reference Cell Type single, Ag/AgCl
Junction Type/Flow Rate open
Type of Electrolyte viscolene
Body Material glass
Tip / Shape flat
Matching Pin no
Max. Pressure 0.1 bar
Temperature Operating Range 0 to 50°C (32 to 122°F)
Body Length/Overall Length 50 mm / 93.5 mm
Temperature Sensor yes
Outer Diameter 12 mm
Cable 7-pole; 1 m (3.3')
Connector Type DIN
Amplifier yes
Recommended Use scalp, skin