Liquiline System CA80AM
Technical Data
CompareAnalyser
| Measuring Principle | Colorimetric |
| Characteristic | Analyzer for ammonium in aqueous solutions |
| Measuring method | Comply with standard colorimetric measuring principle - indophenol blue method following ISO 7150-1, DIN 38406-5 and GB 7481-87 |
| Size | Housing (open version): 793 x 530 x 417 mm 31.22 x 20.87 x 16.42 in Housing (closed version): 793 x 530 x 463 mm 31.22 x 20.87 x 18.23 in Housing with base: 1723 x 530 x 463 mm 67.83 x 20.87 x 18.23 in |
| Design | Open design / cabinet: Plastic ASA-PC Stand housing: Plastic ASA-PC, steel coated |
| Process temperature | 4 to 40 °C (39 to 104 °F) |
| Ambient temperature | 5 to 40 °C (41 to 104 °F) Outdoor version: - 20 to 40 °C (-4 to 104 °F) |
| Process pressure | Unpressurized |
| Sample flow rate | Min. 5 ml/min (0.17 fl.oz/min) |
| Consistency of the sample | Low solids content Turbitity < 50 NTU, aqueous, homogenized |
| Specials | Easy upgrade to measuring station with up to four digital Memosens sensors Automatic calibration and cleaning User-configurable measuring, cleaning and calibration intervals Optional cooling module for longer reagent life time Automatic calibration and cleaning Optional 2 channel version User-definable measuring ranges Flexibale functionality and modular expandable Digital communication for remote access |
| Application | Monitoring and optimization of the cleaning capacity of municipal and industrial wastewater treatment plants Monitoring of the wastewater treatment plant outlet for documentation purposes Monitoring and optimization of activated sludge basins |
| Power supply | 100 to 120 VAC / 200 to 240 VAC ± 10% 24 VDC ± 10% 50 ± 1 or 60 ± 1.2 Hz |
| Output / communication | 2x 0/4 to 20 mA Optional: Webserver, Modbus, Ethernet/IP, Profibus DP |
| Input | 1 or 2 measuring channels Optinonal 1 to 4 digital sensor inputs for sensors with Memosens protocol |
| Measuring range | 0.05 to 20 mg/l NH4-N 0.5 to 50 mg/l NH4-N 1 to 100 mg/l NH4-N 0.5 to 50 mg/l with dilution function to maximum 10 to 1000 mg/l NH4-N |
| Consumables | Reagents and standard solutions CY80AM Cleaner CY800 Maintenance kit CAV800 |
The Liquiline System CA80AM performs high-precision ammonium measurement at all critical control points. As a member of the Liquiline platform, it is plug-and-play and can be easily upgraded to a measuring station - with minimal installation requirements. Automatic calibration and cleaning and low reagent consumption reduce operating and maintenance costs. Advanced diagnostics with remote access ensure process safety and provide process documentation for convenient process management.
Advantages
Convenient commissioning, plug and play thanks to Memosens technology and a user-friendly interface
Easy functional upgrade - even to a complete measuring station - by simply adding and connecting Memosens sensors
Advanced diagnostics with remote access increase process safety
Automatic calibration, automatic cleaning and an optional reagent cooling unit effectively reduce system maintenance and management costs
Standard colorimetric measuring principle - indophenol blue method - compliant with ISO 7150-1, DIN 38406-5 and GB 7481-87
Areas of application
The Liquiline System CA80AM monitors and controls:
Inlet of wastewater treatment plants
Outlet of wastewater treatment plants for documentation
Aeration basins
Water treatment
Purification capacity of municipal and industrial wastewater treatment plants
E+H Product Configurator
Please enter the product order code to configure
Product comparison
Please enter the order codes of the products to compare in the boxes below (up to three products)
Technical documentation
Technical data
Documents/Manuals/Software
Accessories/Spare parts
A
Measuring principle video
Source:Endress+Hauserwebsite public information
Case Studies
Municipal Wastewater Treatment — Wastewater Treatment Plant · EH-ONLINE Application Case
Project Background
Municipal Wastewater Treatment customers at the Wastewater Treatment Plant stage have long faced the challenge of insufficient accuracy in effluent quality monitoring and instruments prone to contamination causing data drift, affecting compliance discharge, placing higher demands on the stability and measurement accuracy of EH-ONLINE.
Solution
Based on the on-site conditions, the Ainstru technical team selected suitable EH-ONLINE for the customer and delivered one-stop service covering selection, installation, commissioning and tuning.
Results
- Measurement accuracy and response speed significantly improved, with stable and reliable data;
- Instruments adapt to complex Municipal Wastewater Treatment conditions, greatly reducing maintenance;
- Helping customers achieve continuous online monitoring and cost reduction, highly recognized by customers.
Food & Beverage — Food Production Line · EH-ONLINE Application Case
Project Background
Food & Beverage customers at the Food Production Line stage have long faced the challenge of high hygienic measurement requirements, with inline instruments needing to withstand CIP/SIP cleaning while maintaining stable accuracy, placing higher demands on the stability and measurement accuracy of EH-ONLINE.
Solution
Based on the on-site conditions, the Ainstru technical team selected suitable EH-ONLINE for the customer and delivered one-stop service covering selection, installation, commissioning and tuning.
Results
- Measurement accuracy and response speed significantly improved, with stable and reliable data;
- Instruments adapt to complex Food & Beverage conditions, greatly reducing maintenance;
- Helping customers achieve continuous online monitoring and cost reduction, highly recognized by customers.
Chemical & Pharmaceutical — Reactors and Storage Tanks · EH-ONLINE Application Case
Project Background
Chemical & Pharmaceutical customers at the Reactors and Storage Tanks stage have long faced the challenge of highly corrosive media and complex conditions, requiring corrosion-resistant and explosion-proof instruments, placing higher demands on the stability and measurement accuracy of EH-ONLINE.
Solution
Based on the on-site conditions, the Ainstru technical team selected suitable EH-ONLINE for the customer and delivered one-stop service covering selection, installation, commissioning and tuning.
Results
- Measurement accuracy and response speed significantly improved, with stable and reliable data;
- Instruments adapt to complex Chemical & Pharmaceutical conditions, greatly reducing maintenance;
- Helping customers achieve continuous online monitoring and cost reduction, highly recognized by customers.
Power & Energy — Power Plant Boiler and Turbine · EH-ONLINE Application Case
Project Background
Power & Energy customers at the Power Plant Boiler and Turbine stage have long faced the challenge of unstable measurement under high temperature and pressure, with high maintenance and downtime costs, placing higher demands on the stability and measurement accuracy of EH-ONLINE.
Solution
Based on the on-site conditions, the Ainstru technical team selected suitable EH-ONLINE for the customer and delivered one-stop service covering selection, installation, commissioning and tuning.
Results
- Measurement accuracy and response speed significantly improved, with stable and reliable data;
- Instruments adapt to complex Power & Energy conditions, greatly reducing maintenance;
- Helping customers achieve continuous online monitoring and cost reduction, highly recognized by customers.