Instrumentation · Engineering · Package · PLC
0755-82221901 sales@ainstru.com

TMT82-1LLR4/115

TMT82-1LLR4/115
Product family: iTEMP TMT82, Transmitter HART
Temperature transmitterIn Stock (8)
Endress+Hauser TMT82-1LLR4/115
iTEMP TMT82, Transmitter HART;Approval ATEX II1G Ex ia IIC T4/T5/T6;Communication/Output Signal HART; 1x 4-20mA;Electrical Connection Screw terminals, housing for DIN rail mounting 17.5mm;Field Housin

Order Code:TMT82-1LLR4/115

Corresponding ext. order code:TMT82-BAA3CB3A1A1A5+Z1

Name:iTEMP TMT82, Transmitter HART

Configuration

CodeNameOrderSpecification
010ApprovalBAATEX II1G Ex ia IIC T4/T5/T6
020Communication; Output SignalAHART; 1x 4-20mA
030Electrical Connection3Screw terminals, housing for DIN rail mounting 17.5mm
040Field HousingCW/o
050Configuration Universal InputB3Ch1: RTD 3-wire, Ch2: RTD 3-wire
060Sensor Type Input 1A1Pt100, IEC60751 (a=0.00385), -200...+850oC, min. span 10K
070Sensor Type Input 2A1Pt100, IEC60751 (a=0.00385)
080Input; InterconnectionA5Sensor backup; PV=CH1(or CH2)
895MarkingZ1Tagging (TAG), metal

Configuration Explained

This is a TMT82 dual-channel DIN rail temperature transmitter, HART protocol, ATEX intrinsically safe approval, can be installed in hazardous areas, 1 x 4-20mA HART output, dual Pt100 three-wire hot backup, automatic switchover to the other channel on probe failure, production is not interrupted, output can use channel 1 or channel 2, 17.5 mm narrow housing, snaps onto a 35 mm rail, compact cabinet layout, with metal tag plate, suitable for redundancy at critical measuring points in chemical plants.