Instrumentation · Engineering · Package · PLC
0755-82221901 sales@ainstru.com

31240.01

31240.01
Product family: Orbisphere 312xx Sensors
High pressure hydrogen EC sensor, stainless steel, guard ring: platinum, o-ring: EPDM (200 bar)
HACH 31240.01
技术参数参数名称说明AccuracyThe greater of ±1% of reading, or ±0.03 ppb, or ±1 Pa (Membrane 2956A)Accuracy 2The greater of ±1% of reading, or ±0.09 ppb, or ±6 Pa (Membrane 2952A)Accuracy 3The greater of ±1% of …

技术参数

参数名称说明
AccuracyThe greater of ±1% of reading, or ±0.03 ppb, or ±1 Pa (Membrane 2956A)
Accuracy 2The greater of ±1% of reading, or ±0.09 ppb, or ±6 Pa (Membrane 2952A)
Accuracy 3The greater of ±1% of reading, or ±1 ppb, or ±50 Pa (Membrane 2995A)
Accuracy 4The greater of ±1% of reading, or ±10 ppb, or ±1 kPa (Membrane 29015A)
Ambient Temperature-5 - 100 °C (23 - 212 °F)
ApplicationCorrosion rate
CertificationsIP68 / NEMA4
Flow RateTypical Flow Rate Membrane 2956A : 50 to 220 mL/min Membrane 2952A : 40 to 200 mL/min Membrane 2995A : 20 to 70 mL/min Membrane 29015A : 20 to 40 mL/min
MaterialO-Ring: EPDM
ParameterHydrogen
Protection ClassIP68 / NEMA4
Radiation Resistance LevelMembrane 2956A : 2x104 rad
Membrane 2952A : 108 rad
Membrane 2995A : 108 rad
Membrane 29015A : N/A
RangeRange at 25 °C Membrane 2956A : 0 - 75 ppb or 0 Pa - 5 kPa
Membrane 2952A : 0 - 300 ppb or 0 Pa - 20 kPa
Membrane 2995A : 0 - 3200 ppb or 0 Pa - 200 kPa
Membrane 29015A : 0 ppb - 32 ppm or 0 - 2000 kPa
Sample Pressure0 - 200 bar
Special FeaturesSilver guard ring
Temperature RangeAccurate from -5 - 50 °C (23 - 122 °F) Resistant - temperature from -5 - 100 °C (23 - 212 °F)
Warranty12 months
Weight140 to 700 grams, depending on the construction materia
What's included?Sensor and manual. Controller, sensor cable, flow-chamber or process connection must be ordered separately.

同系列其他配置

型号
31210.12EC Sensor, Hydrogen, PEEK, 50 bar, O-Ring FKM/FPM

资料下载

2 models · click to collapse