产品概述
Allows you to monitor hydrogen in the power production process.
Hydrogen EC sensor, PEEK, guard ring: platinum, O-ring: FKM/FPM (20 bar)
技术参数
| 参数名称 | 说明 |
|---|---|
| Accuracy | The greater of ±1% of reading, or ±0.03 ppb, or ±1 Pa (Membrane 2956A) |
| Accuracy 2 | The greater of ±1% of reading, or ±0.09 ppb, or ±6 Pa (Membrane 2952A) |
| Accuracy 3 | The greater of ±1% of reading, or ±1 ppb, or ±50 Pa (Membrane 2995A) |
| Accuracy 4 | The greater of ±1% of reading, or ±10 ppb, or ±1 kPa (Membrane 29015A) |
| Ambient Temperature | -5 - 100 °C (23 - 212 °F) |
| Application | Corrosion rate |
| Certifications | IP68 / NEMA4 |
| Flow Rate | Typical Flow Rate Membrane 2956A : 50 to 220 mL/min Membrane 2952A : 40 to 200 mL/min Membrane 2995A : 20 to 70 mL/min Membrane 29015A : 20 to 40 mL/min |
| Material | O-Ring: FKM/FPM |
| Parameter | Hydrogen |
| Protection Class | IP68 / NEMA4 |
| Radiation Resistance Level | Membrane 2956A : 2x10⁴ rad Membrane 2952A : 108 rad Membrane 2995A : 108 rad Membrane 29015A : N/A |
| Range | Range at 25 °C Membrane 2956A : 0 - 75 ppb or 0 Pa - 5 kPa Membrane 2952A : 0 - 300 ppb or 0 Pa - 20 kPa Membrane 2995A : 0 - 3200 ppb or 0 Pa - 200 kPa Membrane 29015A : 0 ppb - 32 ppm or 0 - 2000 kPa |
| Sample Pressure | 0 - 50 bar |
| Special Features | Silver guard ring |
| Temperature Range | Accurate from -5 - 50 °C (23 - 122 °F) Resistant - temperature from -5 - 100 °C (23 - 212 °F) |
| Warranty | 12 months |
| Weight | 140 to 700 grams, depending on the construction materia |
| What's included? | Sensor and manual. Controller, sensor cable, flow-chamber or process connection must be ordered separately. |
应用领域
Ultrapure Water
同系列其他配置
| 型号 | |
|---|---|
| 31240.01 | High pressure hydrogen EC sensor, stainless steel, guard ring: platinu |
资料下载
- Introduction to Hach Services
- Orbisphere Hydrogen Sensors
- User Manual : Orbisphere Model 31xxx Thermal Conductivity Sensors
- Installation & Maintenance Manual : Orbisphere Model 31xxx EC Sensors
- ORBISPHERE EC Sensor-German-Bedienerhandbuch
- ORBISPHERE EC Sensor Quick Guide
- User Manual : Orbisphere Model 31xxx Electrochemical Sensors
- Basic User Manual : Orbisphere Model 31xxx Electrochemical Sensors
Related Products
HACHSC1000 relay module for DIN-rail mounting, 4 Relays, max. 240V
View →
HACHNP6000sc Phosphate Analyzers
View →
A
HACHStainless Steel Chain Mount Kit
View →
A
HACHMercury Thermometer, Celsius, -10 to 200°C
View →
A
HACHMercury Thermometer, Fahrenheit, -30 to 120°F
View →
A
HACHAccu4 Spare Chamber with Plumbing, High Pressure
View →
HACHDETECTR,HI GN,10' CABLE
View →
A
HACHCOVER-SERVICE REPLACEMENT
View →