产品概述
Reliable Monitoring for Critical Water Quality
Managing Hardness in water systems is critical for protecting equipment, supporting operational efficiency, and meeting compliance requirements. The EZ1017sc Hardness analyzer delivers accurate, real-time measurement of Hardness (expressed as calcium carbonate) and enables precise monitoring of calcium levels to help address your operational and compliance needs. By connecting directly to the SC4500 Controller, it ensures seamless integration with your existing processes—delivering clear results you can trust. Key Features Flexible Customisation The EZ1017sc Hardness analyzer offers exceptional versatility, capable of monitoring both single and multiple water streams. With support for up to 8 streams and flexible 4x or 8x mA outputs, it delivers reliable and efficient performance tailored to your needs. It integrates with key industrial communication protocols including Modbus RTU, Modbus TCP/IP, Profinet, Profibus DP, and Ethernet/IP, adapting easily to your system requirements and reporting needs across a range of applications. Reliable and Easy to Use With durable construction and straightforward controls, this analyzer ensures dependable operation in demanding settings. The intuitive user interface makes installation, setup, and daily operation simple, reducing downtime and minimising maintenance. Explore Tailored Solutions Every facility is unique, and we're here to help you find the right fit. Connect with our experts to learn how the EZ1017sc Hardness analyzer can be configured for your water quality goals. We deliver solutions designed around your needs—reach out for guidance, details, and personalized support.
主要特点
- Performance in Challenging Environments:Designed to excel in tough conditions, the system features preconditioning solutions, including self-cleaning filtration panels that seamlessly integrate with EZ sc analyzers. These prefiltration panels ensure reliable, long-term performance while minimizing maintenance needs, making them ideal for demanding applications.
- Cost-Effective and Scalable Monitoring:With support for up to 8 streams, the EZ1017sc reduces per-sampling-point costs while delivering comprehensive insights. Its flexible, scalable design adapts to your evolving monitoring needs, from basic to complex analyses.
- Enhanced Operational Efficiency with Advanced Connectivity:Real-time data and advanced connectivity improve reaction times, enabling faster, more informed decisions. Integrate the EZ1017sc with your systems using configurable analogue or digital outputs for optimized performance.
技术参数
| 参数名称 | 说明 |
|---|---|
| Ambient Temperature | 50 - 86°F (10 - 30 °C) ± 4 °C deviation at 5 - 95% relative humidity (non-condensing) |
| Analog Outputs | Active 0 - 20 mA (or 4 - 20 mA) max. 500 Ohm load, standard: 4, optional: 8 |
| Automatic cleaning | Yes; Frequency freely programmable: 6 hours, 12 hours, daily, weekly |
| Calibration | Automatic; 2-point; offset or slope; Frequency freely programmable: 6 hours, 12 hours, daily, weekly. Note: manufacturer recommends that a calibration is done when the reagents are replaced |
| Certifications | CE, ETL certified to UL and CSA safety standards, UKCA |
| Cycle Time | No Dilution Default: 10 min. Continuous: 8 min. Internal Dilution: Default: 15 min. Continuous: 10 min. |
| Demineralized Water | Internal dilution option: Rinse water & Dilution water |
| Digital Outputs | Profibus DP or Modbus RTU |
| Dimensions (H x W x D) | 740 mm x 500 mm x 340 mm |
| Drain | Atmospheric pressure, vented, min. Ø 32 mm |
| Earth Connection | Dry and clean earth pole with low impedance (< 1 Ohm) using an earth cable of > 2.5 mm² |
| Instrument Air | Dry and oil free according to ISA-S7.0.01-1996 quality standard for instrument air. Used to flush the instrument in corrosive environement. Min. 0.2 bar - Max. 0.5 bar |
| Interferences | Some metal ions interfere by causing fading or indistinct end points or by stoichiometric consumption of EDTA. Suspended or colloidal organic matter can interfere with the end point. Large quantities of color and turbidity interfere. Fats, oil, proteins, surfactants and tar interfere. |
| Lower Limit of Detection (LOD) | No Dilution 0.005 - 0.25 mg/L CaCO₃: 0.005 mg/L 0.005 - 0.5 mg/L CaCO₃: 0.005 mg/L 0.025 - 1 mg/L CaCO₃: 0.025 mg/L Interal Dilution 0.125 - 5 mg/L CaCO₃: 0.125 mg/L 0.25 - 10 mg/L CaCO₃: 0.25 mg/L 0.625 - 25 mg/L CaCO₃: 0.625 mg/L 1.25 - 50 mg/L CaCO₃: 1.25 mg/L 1.875 - 75 mg/L CaCO₃: 1.875 mg/L 2.5 - 100 mg/L CaCO₃: 2.5 mg/L |
| Material | Hinged part: Thermoform ABS, door: PMMA Wall section: Galvanized steel, powder coated |
| Measurement method | Hydroxynaphthol blue colorimetric method |
| Measurement Method | Colorimetric |
| Number of sample streams | Standard: 1 Optional accessory for 2, 4, or 8 sample streams |
| Outputs | Optional: Claros Ethernet connection and Modbus TCP/IP Ethernet connector; LAN version; 10/100 Mbps or Profinet or Ethernet IP |
| Parameter | Hardness, Calcium |
| Power | 100 - 240 VAC, 50/60 Hz Max. power consumption: 120 VA |
| Precision | Better than 3% of full scale at: 0.005 - 0.25 mg/L CaCO₃ 0.005 - 0.5 mg/L CaCO₃; and better than 2% of full scale for standard test solutions. |
| Protection Class | IP44 |
| Range | No Dilution 0.005 - 0.25 mg/L CaCO₃ 0.005 - 0.5 mg/L CaCO₃ 0.025 - 1 mg/L CaCO₃ Interal Dilution 0.125 - 5 mg/L CaCO₃ 0.25 - 10 mg/L CaCO₃ 0.625 - 25 mg/L CaCO₃ 1.25 - 50 mg/L CaCO₃ 1.875 - 75 mg/L CaCO₃ 2.5 - 100 mg/L CaCO₃ |
| Reagent Requirements | Keep between 50 - 86 °F (10 - 30 °C) |
| Relays | 5 contacts, not user configurable: Malfuntion, maintenance, analysis ready, sample ready, sample ready (EZ9150) |
| Sample Flow Rate | 100 - 300 mL/min |
| Sample Pressure | By external overflow vessel (open to atmospheric pressure) |
| Sample Quality | Maximum particle size 100 µm, < 0.1 g/L; Turbidity < 50 NTU |
| Sample Temperature | 50 - 86 °F (10 - 30 °C) |
| Validation | Automatic; Frequency freely programmable: 6 hours, 12 hours, daily, weekly |
| Warranty | 12 months |
| Weight | Max. 77 lb (35 kg) |
| What's included? | EZ1017sc Calcium Hardness Analyzer, Instruction Manual, 1 x Double Bit Door Key, 1 x Mounting Brackets, 3 x empty 2.5L Reagent Containers with Fittings (for Color, Acid, and EDTA Solution) and 1 x empty 5L Reagent Container with Fittings (for Buffer) |
应用领域
Municipal Water
同系列其他配置
| 型号 | |
|---|---|
| EZ1012.9700140T | EZ1012sc Cyanide Colorimetric Analyzer, 1 stream, 4x mA Output |
| EZ1035.9700140T | EZ1035sc Silica Colorimetric Analyzer, 1 stream, 4x mA Output |
| EZ1036.9700140T | EZ1036sc Sulfate Colorimetric Analyzer, 1 stream, 4x mA Output |
| EZ1037.9700140T | EZ1037sc Sulfide Colorimetric Analyzer, 1 stream, 4x mA Output |
| EZ3005.9700140T | EZ3005sc Chloride Ion-Selective Electrode (ISE) Analyzer, 1 stream, 4x |
| EZ3008.9700140T | EZ3008sc Fluoride Ion-Selective Electrode (ISE) Analyzer, 1 stream, 4x |
| EZ3505.9700140T | EZ3505sc Chloride Ion-Selective Electrode (ISE) Standard Addition Anal |
| EZ3508.9700140T | EZ3508sc Fluoride Ion-Selective Electrode (ISE) Standard Addition Anal |
| EZ1016.9700140T | EZ1016sc Hardness, total Colorimetric Analyzer, 1 stream, 4x mA Output |
资料下载
- Maximum Insight with Minimal Effort : EZ sc verses EZ
- 최소한의 노력으로 최대의 효과 : EZ sc ~에 비해 EZ
- Máxima perceção com o mínimo de esforço : EZ sc em comparação com EZ
- Analisadores Online Série EZ
- Maximální přehled s minimálním úsilím : EZ sc vs. EZ
- Online analyzátory řady EZ
- Analyzer Seri EZ Online Solusi Lengkap
- Brochure EZ Series Online Analysers
Related Products
HACHSC1000 relay module for DIN-rail mounting, 4 Relays, max. 240V
HACHNP6000sc Phosphate Analyzers
Stainless Steel Chain Mount Kit
Mercury Thermometer, Celsius, -10 to 200°C
Mercury Thermometer, Fahrenheit, -30 to 120°F
Accu4 Spare Chamber with Plumbing, High Pressure
HACH