产品概述
The Challenge of Consistent Chloride Monitoring
Chloride monitoring in municipal and industrial water can be disrupted by fouling, drift and temperature swings -creating extra maintenance and compliance risk. The EZ3005sc uses discontinuous ISE measurement with temperature control to deliver stable, precise chloride results with minimal operator intervention. Automated operation helps reduce routine maintenance and lower operating cost while keeping reliable chloride data available for process control.
Designed for straightforward integration, the analyzer supports multiple analog and digital outputs - including Modbus, Profinet and Ethernet/IP - so you can bring results and diagnostics directly into your control system.
Benefits: - Discontinuous ISE measurement with temperature control for repeatable results - Automated operation reduces maintenance effort and operator time - Efficient reagent use helps lower total cost of ownership - Analog and digital connectivity: Modbus, Profinet and Ethernet/IP - Real-time results and diagnostics to support compliance and process optimization
主要特点
- Reliable, Stable Measurements:Avoid downtime from unreliable ISE data caused by temperature fluctuations. The EZ3005sc stabilizes both sample and electrode using a temperature controlled flow cell, eliminating measurement errors and delivering precise, repeatable chloride results operators can trust.
- Lower Maintenance Effort:Minimize downtime and service costs with the EZ3005sc. Automated 2 point calibration, reagent alerts, a rugged ISE electrode, and minimal moving parts extend maintenance intervals; so operators spend less time servicing instruments and more time on plant operations.
- Reduced Operating Cost:The EZ3005sc helps optimize operating costs by reducing chemical consumption and service visits. Multi stream capability monitors up to eight sample lines, lowering cost per sampling point and extending equipment life.
- Designed for easy integration into existing control systems:With a wide range of analogue and digital outputs, this analyzer aligns with plant specific requirements, ensuring compatibility with existing infrastructure and immediate access to real time measurement data.
技术参数
| 参数名称 | 说明 |
|---|---|
| Ambient Temperature | 50 - 86 °F (10 - 30 °C) ± 4 °C deviation at 5 - 95% relative humidity (non-condensing) |
| Analog Outputs | Active 0 - 20 mA (or 4 - 20 mA) max. 500 Ohm load, standard 4, optional: 8 |
| Automatic cleaning | Yes; Frequency freely programmable: 6 hours, 12 hours, daily, weekly |
| Calibration | Automatic; 2-point; Frequency freely programmable: 6 hours, 12 hours, daily, weekly. Note: manufacturer recommends that a calibration is done when the reagents are replaced |
| Certifications | CE, ETL certified to UL and CSA safety standards, UKCA |
| Cycle Time | Default: 5 minutes Continuous: 5 minutes |
| Demineralized Water | Does not apply |
| Digital Outputs | Relays: 5 contacts, not user configurable: Malfuntion, maintenance, analysis ready, sample ready, sample ready (EZ9150) Ethernet Connections: Optional: Claros Ethernet connection and Modbus TCP/IP Ethernet connector; LAN version; 10/100 Mbps or Profinet or Ethernet IP RS485 communication: Profibus DP or Modbus RTU |
| Dimensions (H x W x D) | 688 mm x 460 mm x 340 mm |
| Drain | Atmospheric pressure, vented, min. Ø 32 mm |
| Earth Connection | Earth connection Dry and clean earth pole with low impedance (< 1 Ohm) using an earth cable of > 2.5 mm² |
| Instrument Air | Dry and oil free according to ISA-S7.0.01-1996 quality standard for instrument air. Used to flush the instrument in corrosive environement. Min. 0.2 bar - Max. 0.5 bar |
| Interferences | Bromide [(Br)-], sulphide [(S)²-], iodide [(I)-], cyanide [(CN)-] ions may interfere. Mercury [(Hg)+] must be absent. Ammonia [NH₃] and thiosulphate [(S₂O₃)²-] may interfere. Fats, oil, proteins, surfactants and tar. |
| Lower Limit of Detection (LOD) | 2 - 100 mg/L Cl-: 2 mg/L 5 - 250 mg/L Cl-: 5 mg/L 10 - 500 mg/L Cl-: 10 mg/L 20 - 1000 mg/L Cl-: : 20 mg/L |
| Material | Hinged part: Thermoform ABS, Door: PMMA Wall section: Galvanized steel, powder coated |
| Measurement method | Discontinuous, direct measurement by combined Ion-Selective Electrode |
| Measurement Method | Ion-Selective Electrode |
| Number of sample streams | Standard: 1 Optional accessory for 2, 4, or 8 sample streams |
| Parameter | Chloride |
| Power | 100 - 240 VAC, 50/60 Hz Max. power consumption: 120 VA |
| Precision | Better than 2% full scale range for standard test solutions |
| Protection Class | IP44 |
| Range | 2 - 100 mg/L Cl- 5 - 250 mg/L Cl- 10 - 500 mg/L Cl- 20 - 1000 mg/L Cl- |
| Reagent Requirements | Keep between 50 - 86 °F (10 - 30 °C) |
| Sample Flow Rate | 100 - 300 mL/min |
| Sample Pressure | By external overflow vessel (open to atmospheric pressure) |
| Sample Quality | Maximum particle size 100 µm, < 0.1 g/L; Turbidity < 50 NTU |
| Sample Temperature | 50 - 86 °F (10 - 30 °C) |
| Validation | Automatic; Frequency freely programmable: 6 hours, 12 hours, daily, weekly |
| Warranty | 12 months |
| Weight | Max. 77 lb (35 kg) |
| What's included? | EZ3005sc Chloride Ion-Selective Electrode (ISE) Analyzer, Instruction Manual, 1 x Double Bit Door Key, 1 x Mounting Brackets, 1 x ISE Electrode and Chloride Electrode, combined, and 1 x empty 2.5L Reagent Container with Fittings (for Buffer Solution) |
应用领域
Municipal Water
同系列其他配置
| 型号 | |
|---|---|
| EZ1012.9700140T | EZ1012sc Cyanide Colorimetric Analyzer, 1 stream, 4x mA Output |
| EZ1035.9700140T | EZ1035sc Silica Colorimetric Analyzer, 1 stream, 4x mA Output |
| EZ1036.9700140T | EZ1036sc Sulfate Colorimetric Analyzer, 1 stream, 4x mA Output |
| EZ1037.9700140T | EZ1037sc Sulfide Colorimetric Analyzer, 1 stream, 4x mA Output |
| EZ3008.9700140T | EZ3008sc Fluoride Ion-Selective Electrode (ISE) Analyzer, 1 stream, 4x |
| EZ3505.9700140T | EZ3505sc Chloride Ion-Selective Electrode (ISE) Standard Addition Anal |
| EZ3508.9700140T | EZ3508sc Fluoride Ion-Selective Electrode (ISE) Standard Addition Anal |
| EZ1016.9700140T | EZ1016sc Hardness, total Colorimetric Analyzer, 1 stream, 4x mA Output |
| EZ1017.9700140T | EZ1017sc Hardness, calcium Colorimetric Analyzer, 1 stream, 4x mA Outp |
资料下载
- Brochure EZ Series Online Analysers
- EZ Series Online Analyzers
- Analizadores en continuo de la serie EZ
- Analisadores on-line da série EZ
- Analyseurs en ligne série EZ
- EZ 시리즈 온라인 분석기 완벽한 솔루션 제공
- EZ Series: Reliable Chloride Monitoring for Regulatory and Process Confidence
- EZ3000sc/EZ3500sc Series Analyzers Service Data Sheet
Related Products
HACHSC1000 relay module for DIN-rail mounting, 4 Relays, max. 240V
HACHNP6000sc Phosphate Analyzers
Stainless Steel Chain Mount Kit
Mercury Thermometer, Celsius, -10 to 200°C
Mercury Thermometer, Fahrenheit, -30 to 120°F
Accu4 Spare Chamber with Plumbing, High Pressure