Overview
The BL983317 and BL983327 are compact, panel mounted, process controllers for measuring conductivity of a process stream. BL983317: When in automatic mode, the dry contact relay is activated when a reading is below the set point. The relay can be used to activate a dosing pump to add chemical until the desired set point is reached. The device features a large LCD with protective cover. Users may choose from automatic or manual dosing modes.
Specifications
| Parameter | Value |
|---|---|
| SKU | BL983317-1 |
| Temperature Compensation | automatic, from 5 to 50°C (41 to 122°F) with β=2%/ºC |
| Calibration | manual, with calibration trimmer |
| Setpoint | adjustable setpoint: covers measure range |
| Dosing Contact | maximum 2A (internal fuse protection), 250 VAC or 30 VDC |
| Overtime | dosing relay is disabled if setpoint is not reached within the set time interval. Timer adjustable between aprox. 5 to 30 minutes, or disabled by jumper. |
| Power Supply | models "-1": 115/230 VAC; 50/60Hz |
| Input Power | input: 10 VA for 115/230 VAC, 50/60 Hz models; fuse protected; installation category II. |
| Dimensions | 83 x 53 x 99 mm (3.3 x 2.1 x 3.9'') |
| Weight | 115/230 VAC models 300 g (10.6 oz) |
Related Products
HANNA InstrumentsPortable multi-range conductivity meter
View →
HANNA InstrumentsLow range ammonia nitrogen portable photometer
View →
HANNA InstrumentsPortable multi-range turbidimeter
View →
HANNA InstrumentsPortable honey color analyzer
View →
HANNA InstrumentsSaltwater aquarium salinity meter
View →
HANNA InstrumentsPortable free chlorine photometer
View →
HANNA InstrumentsBluetooth smart pH meter
View →
HANNA InstrumentsPortable free and total chlorine photometer
View →