General purpose pH electrode
Special pH and ORP Electrodes Tips for the most accurate measurements Keep Electrode Hydrated It is advisable for pH electrodes to be kept in a storage solution when not in use. Placing the electrode

Overview

Special pH and ORP Electrodes Tips for the most accurate measurements Keep Electrode Hydrated It is advisable for pH electrodes to be kept in a storage solution when not in use. Placing the electrode in a small glass filled with storage solution is suitable. An option for pocket meters is to place a small piece of sponge in the meter’s cap and pour storage solution into the cap to wet the sponge.

Pouring off any excess solution beforehand, the cap can then be placed on the meter. If a storage solution is not available, the next best option is to use pH 4.01 buffer (pH 7.01 is also suitable to a lesser extent). Clean electrodes before use Clean the junction of your electrodes once a day or at least once a week to prevent junction clogging and to maintain accuracy.

Immerse the electrode in the proper cleaning solution for at least 15 to 20 minutes. Hanna offers a wide range of cleaning solutions for general purpose and specific applications. Replace electrodes once a year If your electrode takes too long to stabilise a reading, or readings fluctuate wildly, it is most likely time to replace the electrode.

The typical life span of any pH electrode is from 6 months to 1.5 years. Additional Tips Calibration and storage solutions should be changed regularly (i.e. monthly).

Calibrate the meter often if a high degree of accuracy is required. Remember that the calibration is as only as good the buffer being used (i.e. old or contaminated buffer solutions may not have the same value on the label).

Single-use calibration sachets, as opposed to bottles, ensure that your buffer solution is always fresh. If the meter takes an unusually long time to get a stable reading, the junction may be clogged. Rinse the probe with purified water after each use.

Specifications

ParameterValue
SKUHI1217-1
SKUHI-1217-1
DescriptionpH electrode
Referencesingle, Ag/AgCl
Junction / Flow Rateceramic, single
Electrolytegel
Max Pressure2 bar
RangepH: 0 to 13
Recommended Operating Temp.0 to 70°C (32 to 158°F)
Glass TypeGP (general purpose)
Tip /Shapespheric (dia: 5.0 mm)
Temperature Sensoryes
Amplifieryes
Body MaterialPEI
Cable typecoaxial; 1 m
Recommended Usegeneral purpose
ConnectionHI-1217-1 DIN**

Accessories

  • HI-70004P: HI-70004P pH 4.01 Buffer Sachets, 25 x 20 mL sachets
  • HI-70007P: HI-70007P pH 7.01 Buffer Sachets, 25 x 20 mL sachets
  • HI-70300L: HI-70300L pH Electrode Storage Solution, 500 mL
  • HI-7061L: HI-7061L Electrode Cleaning Solution - General Purpose, 500mL
  • HI-700601P: HI-700601P Cleaning Solution - General Purpose, 25 x 20ml sachets