Instrumentation · Engineering · Package · PLC
0755-82221901 sales@ainstru.com

HI2003-02

HI2003-02
HI-2003 Edge Conductivity Meter
HANNA Instruments HI2003-02
edge® EC is dedicated to testing Conductivity. Using Hanna's latest technology and design, edge® can be adapted to suit your working method. Select from hand held, wall mount or bench top - or switch

Overview

edge® EC is dedicated to testing Conductivity. Using Hanna's latest technology and design, edge® can be adapted to suit your working method. Select from hand held, wall mount or bench top - or switch between them all.

edge® is happy any which way. Simply plug in the electrode, then play. Measurement is simpler, quicker and more accurate than ever.

Results are fast and reliable. Combining high precision technology with the user-friendly dimensions and format of a small digital tablet, edge® is breaking new ground in electrochemistry. This is a must-have meter for your laboratory.

Cutting-edge design, capacitive touch keypad and an intuitive user interface are just a few of the great attributes that make edge® stand out as the world's first Hybrid meter on the market. Key features 4-ring potentiometric probe - covers all ranges from 0.00µS/cm to 500 mS/cm (absolute EC) Accuracy - ±1% of the reading (±0.05µS/cm or 1 digit, whichever is greater) Calibration - offset (0 µS/cm) and cell factor calibration, choice of 5 standards Auto-ranging or manual range selection EC, TDS and Salinity reading modes Temperature compensation - automatic, NoTC (absolute) GLP data - records date, time, offset, and cell constant value (K) Adjustable EC to TDS conversion factor Adjustable temperature correction factor Capacitive touch keys Large 140mm LCD display with clear full text readout Digital smart electrodes enable edge to automatically recognise and store sensor type, ID and calibration information - with 3.5mm phono connection Programmed for basic and advanced modes Log-on demand and interval logging PC interface via micro USB Data download onto USB stick via standard USB

Specifications

ParameterValue
SKUHI2003-02
EC Range0.00 to 29.99 µS/cm; 30.0 to 299.9 µS/cm; 300 to 2999 µS/cm; 3.00 to 29.99 mS/cm; 30.0 to 200.0 mS/cm; up to 500.0 mS/cm absolute EC
Resolution0.01 µS/cm; 0.1 µS/cm; 1 µS/cm; 0.01 mS/cm; 0.1 mS/cm
Accuracy (@25°C)±1% of reading (±0.5 µS/cm or 1 digit, whichever is greater)
Calibrationsingle cell factor calibration; six standards available: 84µS/cm, 1413 µS/cm, 5.00 mS/cm, 12.88 mS/cm, 80.0 mS/cm, 118.8 mS/cm, one point offset: 0.00 µS/cm
Temperature Coefficient0.00 to 6.00%/°C (for EC and TDS only), default value is 1.90%/°C
TDS Range0.00 to 14.99 mg/L (ppm); 15.0 to 149.9 mg/L (ppm); 150 to 1499 mg/L (ppm); 1.50 to 14.99 g/L; 15.0 to 100.0 g/L; up to 400.0 g/L absolute TDS using 0.80 conversion factor
TDS Factor0.40 to 0.80 (default value is 0.50)
Salinity Range0.0 to 400.0 % NaCl; 2.00 to 42.00 PSU; 0.0 to 80.0 g/L
Temperature Range-20.0 to 120.0 °C
Accuracy±0.5°C
Additional Specifications:
ProbeHI-763100 digital four-ring conductivity probe with 3.5 mm connector and 1m cable
Loggingup to 1000 (400 for basic mode) records organised in: manual log-on demand (max. 200 logs), manual log-on stability (max. 200 logs), interval logging (max. 600 samples; 100 lots)
Connectivity1 USB port for storage; 1 micro USB port for charging and PC connectivity
Environment0 to 50°C; RH max 95% non-condensing
Power Supply5 VDC adapter (included)
Dimensions2020 x 140 x 12 mm
Weight250 g
Body MaterialPEI
Type4-ring
Insulatorglass
ElectrodesPlatinum
Range0 to 200 mS/cm
Environmental Conditions-5 to 100°C
Body Length120mm
Temperature SensorYes
AmplifierNo
OtherUnique serial ID
Digitalyes
Cable1m
Connection3.5mm jack

Accessories

  • HI-7039L: HI-7039L 5000 uS/cm Conductivity Solution, 500mL bottle
  • HI-763100: HI-763100 Edge Digital EC Platinum 4 Ring Electrode
  • HI-7031L: HI-7031L 1413uS/cm Conductivity Calibration Solution, 500mL
  • HI-7030L: HI-7030L Conductivity Calibration Solution 12880uS/cm, 500mL

Manuals and documentation