Instrumentation · Engineering · Package · PLC
0755-82221901 sales@ainstru.com

HI2600-02

HI2600-02
HI-2600 pH EC and DO Benchtop meter
HANNA Instruments HI2600-02
HI-2600 pH EC and DO Benchtop meter HI-2600 benchtop meter enables fast, accurate measurements utilising the Hanna Instruments digital electrode / probes for pH (HI-11310), ORP (HI-36180), conductivit

Overview

HI-2600 pH EC and DO Benchtop meter HI-2600 benchtop meter enables fast, accurate measurements utilising the Hanna Instruments digital electrode / probes for pH (HI-11310), ORP (HI-36180), conductivity (HI-763100), or dissolved oxygen (HI-764080). Each digital electrode / probe has a unique serial number and the parameter that it measures is automatically identified by the meter. Once connected, the single sensor or sensors are ready to measure the parameter along with temperature.

The intuitive design simplifies configuration, calibration, measurement, data logging and transfer of data to a USB thumb drive or computer. The benchtop meter offers a Basic Mode that streamlines measurement configuration and is useful for routine applications. The compact meter (and probe) can be used as a portable device (using its rechargeable battery) or used as laboratory instrumentation.

The meter has a large display with capacitive keys which features a 5.5” (14 cm) LCD. The large display provides a 130 ° wide viewing angle. Main feature Automatic parameter recognition pH/ORP choice of measurement units: pH, mV; EC choice of measurement unit: Conductivity: µS/cm, mS/cm TDS - ppm (mg/L), g/L Salinity - NaCl, PSU, g/L DO choice of measurement unit: ppm (mg/L), % Adjustable resolution for pH and EC measurements CAL Check indicators - HI-11310, HI-763100, HI-764080 digital probes only Sensor Check indicators - HI-11311 and HI-12301 digital probes only Basic mode for simplified operation Dedicated GLP key GLP data included with logged data Simplified data transfer to a PC Up to 8 hour battery life when used as a portable device Internal clock and date Key Features Digital probe input - simply connect each probe via the 3.5 mm jack, Digital Smart Electrodes are automatically recognised.

Clear, full text readout - The HI-2600 presents clear, detailed text prompts at the bottom section of the screen. Two operating modes - HI-2600 can be used in Standard or Basic Operating Modes (pH and EC only). Standard Mode enables all features while Basic Mode displays a simplified screen and features —ideal for routine measurements.

Data logging - The HI-2600 provides ample data storage capacity, allowing you to save up to 1000 log records. Each data set comprehensively includes readings, GLP information, as well as date and time stamps. GLP - The sensor retains data from the most recent calibration, including date, time, and buffers utilised.

When connected to the HI-2600, this GLP information is automatically transferred. CAL Check - Hanna’s exclusive CAL Check feature analyses the pH electrode’s response in calibration buffers to detect potential issues such as contaminated solutions or dirty electrodes. Upon completing calibration, indicators reflecting the probe’s condition—based on its offset and slope characteristics—are displayed on the measurement screen.

Sensor Check (using HI-11311 and HI-12301 pH probes only) - When paired with compatible Hanna electrodes featuring a matching pin, HI-2600 continuously monitors the impedance of the pH measuring electrode, providing real-time alerts in case of glass breakage. During calibration, Sensor Check displays the junction condition, by evaluating the reference junction resistance. ORP measurement - Measure ORP (REDOX) in mV units by using a compatible ORP sensor.

Capacitive touch keypad - HI-2600 features sensitive capacitive touch buttons for accurate keystrokes for navigating menus and screens. Since they are part of the screen, the buttons can never get clogged with sample residue. Easy to read LCD - HI-2600 features a 5.5” (14 cm) LCD display that you can clearly view from over 5 m (16.4’).

The large display, with its wide 130° viewing angle, provides one of the easiest to read LCDs in the industry. Rechargeable battery - The integrated rechargeable battery offers maximum portability with a long 8-hour battery life. USB-C port for charging and data transfer - The HI-2600 features a streamlined USB-C port that facilitates charging the integrated rechargeable battery, transferring data to your computer, and exporting files directly to a flash drive.

What's included: HI-11310 pH digital electrode HI-764026 Electrode holder with the following accessories: base plate with integrated pivot pin; cable holder clip, attached pH calibration kit: pH 4 buffer solution (2 sachets); pH 7 buffer solution (4 sachets); pH 10 buffer solution sachets (2 sachets) Electrode cleaning solution (2 sachets) HI-920018 - 1.5m USB-C to USB-C cable USB-C power adapter Instrument quality certificate Probe / electrode quality certificate Battery safety insert Quick reference guide with QR code for instruction manual download Warranty - Meter has 2 years. 6 months from date of purchase for electrodes and probes More accessories to buy separately Electrodes HI-10530 Triple ceramic, double junction, glass body, refillable pH electrode with conical tip and temperature sensor HI-10430 Triple ceramic, double junction, glass body, refillable pH electrode with temperature sensor HI-11310 Glass body, double junction, refillable pH/temperature electrode HI-11311 Glass body, double junction, refillable pH/ temperature electrode with enhanced diagnostics HI-12300 Plastic body, double junction, gel filled, non-refillable pH / temperature electrode HI-12301 Plastic body, double junction, gel filled, non-refillable pH/temperature electrode with enhanced diagnostics HI-36180 Glass body, double junction, refillable ORP/temperature probe HI-763100 Digital EC / temperature probe HI-764080 Digital DO / temperature probe More solutions to buy separately pH solutions – 500mL (other sizes available) HI-7001L pH 1.68 buffer solution, 500 mL HI-7004L pH 4.01 buffer solution, 500 mL HI-7006L pH 6.86 buffer solution, 500 mL HI-7007L pH 7.01 buffer solution, 500 mL HI-7009L pH 9.18 buffer solution, 500 mL HI-7010L pH 10.01 buffer solution, 500 mL ORP solutions HI-7021L ORP test solution, 240 mV @ 25 °C, 500 mL HI-7022L ORP test solution, 470 mV @ 25 °C, 500 mL HI-7091L Reducing Pretreatment Solution, 500 mL bottle +14 g (set) HI-7092L Oxidizing Pretreatment Solution, 500 mL bottle pH electrode maintenance solutions HI-70300L Storage solution, 500 mL HI-70000P Electrode rinse sachets (25 sachets × 20 mL) HI-7061L General cleaning solution, 500 mL Conductivity solutions HI-7030L 12880 µS/cm calibration solution, 500 mL HI-7031L 1413 µS/cm calibration solution, 500 mL HI-7033L 84 µS/cm calibration solution, 500 mL HI-7034L 80000 µS/cm calibration solution, 500 mL HI-7035L 111800 µS/cm calibration solution, 500 mL HI-7039L 5000 µS/cm calibration solution, 500 mL HI-7037L 100% NaCl sea water standard solution, 500 mL Dissolved oxygen solutions HI-7040L Zero oxygen solution, 500 mL HI-7041S Polarographic dissolved oxygen electrolyte solution, 30 mL Other useful accessories HI-764026 Electrode holder for HI-2600 family HI-764080A/P Dissolved oxygen spare screw cap membranes for HI-764080 DO probe (5 pcs.) Models in the series HI-2600 pH, EC, DO Benchtop Meter with Data Logging HI-2620 pH and ORP Benchtop Meter HI-2621 pH and ORP Benchtop Meter with Data Logging HI-2630 EC & TDS Benchtop Meter HI-2631 EC, TDS and Salinity Benchtop Meter with Data Logging HI-2640 Dissolved Oxygen Benchtop Meter HI-2641 Dissolved Oxygen Benchtop Meter with Data Logging

Specifications

ParameterValue
SKUHI2600-02
pH - using HI-11310 pH electrode
pH Rangebasic mode: -2.00 to 16.00 pH; standard mode: -2.000 to 16.000 pH
Resolutionbasic mode: 0.01 pH; standard mode: 0.001 pH
Accuracy (@25°C)basic mode: ±0.01 pH; standard mode: ±0.002 pH
CalibrationBasic Mode: Automatic, up to three points, option to select 5 standard buffers. Standard mode: Automatic, up to five point calibration, 7 standard buffers available and two custom buffers.
Temperature CompensationAutomatic, -5.0 to 100.0°C (using integral temperature sensor)
LoggingUp to 1000* (in Standard Mode) records organised in: • Manual log-on demand, maximum 200 logs • Manual log-on-stability, maximum 200 logs • Interval logging in Standard Mode
mV - using HI-11310 pH electrode
mV pH Range±1000 mV
ORP – using HI-36180 ORP electrode
ORP Range±2000.0 mV
EC - using HI-763100 four-ring conductivity probe
EC Range0.00 to 29.99 µS/cm; 30.0 to 299.9 µS/cm; 300 to 2999 µS/cm; 3.00 to 29.99 mS/cm; 30.0 to 200.0 mS/cm; up to 500.0 mS/cm, and absolute conductivity (TDS) is the conductivity value without temperature compensation
Temperature Coefficient0.00 to 6.00%/°C (for EC and TDS only), default value is 1.90%/°C
TDS - using HI-763100 four-ring conductivity probe
TDS Range0.00 to 14.99 mg/L (ppm); 15.0 to 149.9 mg/L (ppm); 150 to 1499 mg/L (ppm); 1.50 to 14.99 g/L; 15.0 to 100.0 g/L; up to 400.0 g/L absolute TDS using 0.80 conversion factor
TDS Factor in Standard Mode0.40 to 0.80 (default value is 0.50)
Salinity - using HI-763100 four-ring conductivity probe
Salinity Range0.0 to 400.0 % NaCl; 2.00 to 42.00 PSU; 0.00 to 80.00 g/L
Calibration % NaCl onlyone point with HI-7037 standard
Dissolved Oxygen - using HI-764080 DO Probe
DO Range0.00 to 45.00 PPM (mg/L); 0.0 to 300.0% saturation
Accuracy±1.5% of reading (± 1 digit)
Pressure CompensationAutomatic 450.0 to 850.0 mmHg – (0.1 mmHg resolution)
Salinity Compensation0 to 40 g/L – (1g/L resolution)
Temperature
Temperature Range-20.0 to 120.0 °C
Additional Specifications
PC InterfaceUSB-C
Storage InterfaceUSB-C
Power supplyUSB Type C (5 VDC; 500mA)
Environment0 to 50˚C; Maximum 95% RH non-condensing
Dimensions205 x 160 x 77 mm
WeightApproximately 0.85 kg
Ordering InformationHI-2600 pH/EC/DO benchtop meter includes HI-11310 glass body refillable pH electrode, pH 4 buffer solution sachets (2), pH 7 buffer solution sachets (4), pH 10 buffer solution sachets (2), electrode cleaning solution sachets (2), electrode holder, 1.5 m USB-C to C USB-C cable. All HI-2600 series compatible pH, EC and DO digital probes are interchangeable with HI-2600 and can be ordered separately.

Accessories

  • HI-70004P: HI-70004P pH 4.01 Buffer Sachets, 25 x 20 mL sachets
  • HI-70007P: HI-70007P pH 7.01 Buffer Sachets, 25 x 20 mL sachets
  • HI-70010P: HI-70010P pH 10.01 Buffer Sachets, 25 x 20 mL sachets
  • HI-7006L: HI-7006L pH 6.86 Buffer Solution, 500 mL bottle
  • HI-7009L: HI-7009L pH 9.18 Buffer Solution, 500 mL bottle
  • HI-7021L: HI-7021L ORP test solution, 240mV, 500 mL bottle
  • HI-7022L: HI-7022L ORP test solution, 470mV, 500 mL bottle
  • HI-7061L: HI-7061L Electrode Cleaning Solution - General Purpose, 500mL
  • HI-7091L: HI-7091L ORP Reducing pretreatment solution
  • HI-7092L: HI-7092L ORP Oxidizing pretreatment solution