Instrumentation · Engineering · Package · PLC
0755-82221901 sales@ainstru.com

HI6000-3

A
Conductivity Module
The HI6000-3 Conductivity module is designed to be used with the HI6000 Multiparameter Modular System, featuring plug-and-play design and multiparameter measurement flexibility.

Overview

The HI6000-3 Conductivity module is designed to be used with the HI6000 Multiparameter Modular System, featuring plug-and-play design and multiparameter measurement flexibility. It measures conductivity, resistivity, TDS, and salinity, and supports the USP <645> reading mode for purified water analysis. The HI7631233 platinum four-ring probe is recommended for use with this module.

Specifications

ParameterValue
SKUHI6000-3
Conductivity Range0.000 to 9.999 μS/cm; 10.00 to 99.99 μS/cm; 100.0 to 999.9 μS/cm; 1.000 to 9.999 mS/cm; 10.00 to 99.99 mS/cm; 100.0 to 1000.0 mS/cm
Conductivity Resolution0.001 μS/cm; 0.01 μS/cm; 0.1 μS/cm; 0.001 mS/cm; 0.01 mS/cm; 0.1 mS/cm
Conductivity Accuracy±1 % of reading or ±0.010 μS/cm, whichever is greater
Cell Constant0.0500 to 200.0000 /cm
Resistivity Range1.0 to 99.9 Ω·cm; 100 to 999 Ω·cm; 1.00 to 9.99 KΩ·cm; 10.0 to 99.9 KΩ·cm; 100 to 999 KΩ·cm; 1.00 to 9.99 MΩ·cm; 10.0 to 100.0 MΩ·cm
Resistivity Resolution0.1 Ω·cm; 1 Ω·cm; 0.01 KΩ·cm; 0.1 KΩ·cm; 1 KΩ·cm; 0.01 MΩ·cm; 0.1 MΩ·cm
Resistivity Accuracy±1 % of reading or ±1 Ω•cm, whichever is greater
TDS Resolution0.001 ppm; 0.01 ppm; 0.1 ppm; 0.001 ppt; 0.01 ppt; 0.1 ppt
TDS Accuracy±1 % of reading or ±0.01 ppm, whichever is greater
Seawater Salinity Resolution0.01 PSU; 0.01 ppt; 0.1 %
Seawater Salinity Accuracy±1 % of reading
Temperature Resolution0.1 °C, 0.1 °F, 0.1 K
Temperature Accuracy±0.2 °C, ±0.4 °F, ±0.2 K