Instrumentation · Engineering · Package · PLC
0755-82221901 sales@ainstru.com

HI801-11

HI801-11
Holmium Oxide Glass Filter
The HI801-11 is a holmium oxide glass filter that is used to validate the wavelength accuracy of the HI801 and HI802 iris spectrophotometer.

Overview

The HI801-11 is a holmium oxide glass filter that is used to validate the wavelength accuracy of the HI801 and HI802 iris spectrophotometer. The filter is mounted in a 10 mm square anodized holder and comes with a plastic holder that protects the filter when not in use. Absorbance peaks are measured at 361, 446, 536, and 637 nm, and it is supplied with a certificate of analysis traceable to SRM 2034.

Specifications

ParameterValue
SKUHI801-11
Peak Wavelengths @ 25°C361, 446, 536, and 637 nm. Actual value traceable to SRM 2034 found on COA
Peak Wavelength Uncertainty @ 25°C<0.1 nm for all peaks
Dimensions10 x 10 mm
Recommended useWavelength check of HI801 and HI802 iris
Ordering InformationHI801-11 is supplied with quality certificate