HI88713-01

HI88713-01
Benchtop Turbidity Meter
HANNA Instruments HI88713-01
This is a benchtop turbidity meter using the ISO 7027 method, capable of measuring in FNU, FAU, NTU, and EBC units.

Overview

This is a benchtop turbidity meter using the ISO 7027 method, capable of measuring in FNU, FAU, NTU, and EBC units. It features normal, average, and continuous measuring modes, supports two- to five-point calibration, stores up to 200 records internally, and allows data transfer via USB. It has a backlit graphic LCD display with tutorial mode and contextual help.

Specifications

ParameterValue
SKUHI88713-01
FNU Mode Range0.00 to 1000 FNU
FNU Mode Resolution0.01 (0.00 to 9.99 FNU); 0.1 (10.0 to 99.9 FNU); 1 (100 to 1000 FNU)
FNU Mode Accuracy±2% of reading plus stray light
FAU Mode Range10.0 to 4000 FAU
FAU Mode Resolution0.1 (10.0 to 99.9 FAU); 1 (100 to 4000 FAU)
FAU Mode Accuracy @25°C/77°F± 10% of reading
NTU Ratio Mode Accuracy±2% of reading plus stray light; ±5% of reading above 1000 NTU
NTU Non-ratio Mode Range0.00 to 1000 NTU; 0.00 to 245 EBC
NTU Non-ratio Mode Accuracy @25°C/77°F±2% of reading plus stray light
Range Selectionautomatic
Repeatability±1% of reading or stray light, whichever is greater
Stray Light< 0.1 NTU (0.05 EBC)
Light Detectorsilicon photocell
Light SourceIR LED
MethodISO 7027 method
Measuring Modenormal, average, continuous.
Turbidity Standards<0.1, 15, 100, 750 FNU and 2000 NTU
Calibrationtwo, three, four or five-point calibration
Log Memory200 records
Serial InterfaceUSB
Environment0°C to 50°C (32 to 122°F); max 95% RH non-condensing
Power Supply12 Vdc