Instrumentation · Engineering · Package · PLC
0755-82221901 sales@ainstru.com

HI943500C

HI943500C
Conductivity Controller (Low Range)
HANNA Instruments HI943500C
Four different models are available with different measurement ranges to satisfy any application requirements. The output configuration for connecting a recorder or a PLC can be chosen between 0-20 or

Overview

Four different models are available with different measurement ranges to satisfy any application requirements. The output configuration for connecting a recorder or a PLC can be chosen between 0-20 or 4-20 mA. The LED on the front panel indicates the operating status of the controller.

The automatic temperature compensation (ATC) is performed directly by the HI-7638 probe with built-in temperature sensor. The front panel is protected behind a transparent splash-proof cover. Order Information: The HI-943500 series is supplied with mounting brackets and instructions.

Specifications

ParameterValue
SKUHI943500C
Range0.0 to 1999 µS/cm
Resolution1 µS/cm
Accuracy±2% F.S.
Temperature Compensationautomatic, 0 to 60°C (32 to 140°F), with ß=2%/°C
Recorder Output4-20 mA (isolated)
Setpoint Relay1, isolated, 2A, Max. 240 V, resistive load, 1,000,000 strokes
Alarm Relay1, isolated, 2A, Max. 240 V, resistive load, 1,000,000 strokes
Power Supply115/230 ±10% Vac; 50/60 Hz
EnclosureBlack anodised aluminium body; front and back with ABS; transparent splash-proof front cover
Environment-10 to 50°C (4 to 122°F); RH max 95% non-condensing
Panel Cutout141 x 69 mm (5.6 x 2.7")
Weight1 kg (2.2 lb.)

Accessories

  • HI-7638: HI-7638 Submersion Conductivity Sensor, 3 metre
  • HI-7638/10: HI-7638/10 Submersion Conductivity Sensor
  • HI-7638/20: HI-7638/20 4-ring EC probe with Pt rings, temperature NTC sensor for ATC