Overview
The HI98143 series is designed to accept signals directly from a pH electrode and a conductivity probe at the same time. Direct connection of the probes to the transmitter assures a positive electrical connection with no signal loss, making it ideal for remote process control applications. The output signals are proportional to the input signals but independent of changes in load or cable capacitance. Compensation for the effects of temperature for EC measurements is performed by the transmitter's Automatic Temperature Compensation circuitry.
Specifications
| Parameter | Value |
|---|---|
| SKU | HI98143-01 |
| Range | 0 to 14 pH; 0 to 10 mS/cm |
| Accuracy (@25°C/77°F) | ±0.5% f.s. pH; ±2% f.s. EC |
| Calibration | manual, 2 point, through trimmers: pH: offset and slope trimmers; EC: 0 and 5 mS/cm trimmers |
| EC Temp. Compensation | automatic, 0 to 60°C (32 to 132°F) with β=2%/°C |
| pH Electrode | HI1001 pH electrode (suggested, not included), HI1283 matching pin (not included) |
| EC Probe | HI3001 (not included) with cell constant 2.1 |
| Casing | IP54 |
| Power Supply | 12-24 VDC |
| Environment | 0 to 50°C (32 to 122°F); RH max 95% non-condensing |
| Dimensions | 160 x 105 x 31 mm (6.3 x 4.1 x 1.2'') |
| Weight | 280 g (9.9 oz.) |