Instrumentation · Engineering · Package · PLC
0755-8222 1901 sales@ainstru.com

CPS77D 非玻璃数字式pH电极

CPS77D 非玻璃数字式pH电极

非玻璃数字式PH电极

Memosens CPS77D is the digital specialist for hygienic applications. Its bacteria-proof reference with contamination-resistant gel guarantees stable measurements. The sensor is sterilizable with hot steam, autoclavable and features an unbreakable PEEK shaft assuring highest product safety. Memosens CPS77D is not the latest Memosens generation. To get information on the new Memosens CPS77E sensor with extended functionality, click here.
Compare

pH

Measuring Principle
Application Hygienic and sterile applications, food industry, pharmaceutical industry, biotechnology
Characteristic Sterilizable and autoclavable ISFET sensor for pH measurement
Digital with Memosens technology
Reference Double chamber reference system
Measuring range pH 0 to 14
Material Sensor shaft: PEEK
Seals: FFKM
Junction: ceramic
Dimension Diameter: 12 mm (0.47 inch)
Shaft lengths: 120, 225 and 360 mm
(4.72, 8.86 and 14.2 inch)
Process temperature -15 to 135°C (5 to 275°F)
Process pressure Max. 11 bar abs at 100 °C
(Max. 160 psi at 212 °F)
Temperature sensor Pt1000
Connection Inductive, digital connection head with Memosens technology
Ingress protection IP 68 (10 m (33 ft) water column, 25 °C (77 °F), 45 days
Certificates, declarations and approvals 3-A, FDA compatible
EHEDG Typ EL Class I aseptic
ASME BPE-2018
Regulation (EC) No. 1935/2004
USP , USP Class VI, 121 °C (250 °F)

The Memosens CPS77D digital electrode is the measurement specialist for hygienic applications. The bacteria-resistant reference lead and the contamination-resistant gel guarantee stable and reliable measurement results. The electrode withstands steam sterilization and autoclaving, and the PEEK electrode shaft cannot break, ensuring maximum product safety. In addition, Memosens digital technology offers maximum process and data integrity with simple operation. The electrode is moisture- and corrosion-resistant and can be calibrated in the laboratory for predictive maintenance.

Benefits
The sensor cannot break, ensuring maximum product safety
Biocompatibility certification
Suitable for steam sterilization and autoclaving
Long calibration intervals, low maintenance
Six times higher CIP stability compared with conventional ISFET pH electrodes
Optional pharmaceutical CoC certification
Non-contact inductive signal transmission for maximum process safety

Applications
Hygienic and sterile applications:
• Food industry
• Life sciences
• Biotechnology
Certified for all relevant hygienic applications, e.g. EHEDG, 3A, EU1935/2004, USP 87/88.

 

E+H Product Model Configuration Tool

Please enter the product model to configure

Product comparison

Please enter the product models to compare in the fields below; up to three products can be compared

 

   

Technical documentation

   

Technical data

 

 

Documents/Manuals/Software

 

 

Accessories/Spare parts

 

A

A

Source: Endress+Hauserpublic information from the website

Municipal Wastewater Treatment

Municipal Wastewater Treatment — Wastewater Treatment Plant · pH Application Case

Project Background

Municipal Wastewater Treatment customers at the Wastewater Treatment Plant stage have long faced the challenge of insufficient accuracy in effluent quality monitoring and instruments prone to contamination causing data drift, affecting compliance discharge, placing higher demands on the stability and measurement accuracy of pH.

Solution

Based on the on-site conditions, the Ainstru technical team selected suitable pH for the customer and delivered one-stop service covering selection, installation, commissioning and tuning.

Results

  • Measurement accuracy and response speed significantly improved, with stable and reliable data;
  • Instruments adapt to complex Municipal Wastewater Treatment conditions, greatly reducing maintenance;
  • Helping customers achieve continuous online monitoring and cost reduction, highly recognized by customers.
Food & Beverage

Food & Beverage — Food Production Line · pH Application Case

Project Background

Food & Beverage customers at the Food Production Line stage have long faced the challenge of high hygienic measurement requirements, with inline instruments needing to withstand CIP/SIP cleaning while maintaining stable accuracy, placing higher demands on the stability and measurement accuracy of pH.

Solution

Based on the on-site conditions, the Ainstru technical team selected suitable pH for the customer and delivered one-stop service covering selection, installation, commissioning and tuning.

Results

  • Measurement accuracy and response speed significantly improved, with stable and reliable data;
  • Instruments adapt to complex Food & Beverage conditions, greatly reducing maintenance;
  • Helping customers achieve continuous online monitoring and cost reduction, highly recognized by customers.
Chemical & Pharmaceutical

Chemical & Pharmaceutical — Reactors and Storage Tanks · pH Application Case

Project Background

Chemical & Pharmaceutical customers at the Reactors and Storage Tanks stage have long faced the challenge of highly corrosive media and complex conditions, requiring corrosion-resistant and explosion-proof instruments, placing higher demands on the stability and measurement accuracy of pH.

Solution

Based on the on-site conditions, the Ainstru technical team selected suitable pH for the customer and delivered one-stop service covering selection, installation, commissioning and tuning.

Results

  • Measurement accuracy and response speed significantly improved, with stable and reliable data;
  • Instruments adapt to complex Chemical & Pharmaceutical conditions, greatly reducing maintenance;
  • Helping customers achieve continuous online monitoring and cost reduction, highly recognized by customers.
Power & Energy

Power & Energy — Power Plant Boiler and Turbine · pH Application Case

Project Background

Power & Energy customers at the Power Plant Boiler and Turbine stage have long faced the challenge of unstable measurement under high temperature and pressure, with high maintenance and downtime costs, placing higher demands on the stability and measurement accuracy of pH.

Solution

Based on the on-site conditions, the Ainstru technical team selected suitable pH for the customer and delivered one-stop service covering selection, installation, commissioning and tuning.

Results

  • Measurement accuracy and response speed significantly improved, with stable and reliable data;
  • Instruments adapt to complex Power & Energy conditions, greatly reducing maintenance;
  • Helping customers achieve continuous online monitoring and cost reduction, highly recognized by customers.