Instrumentation · Engineering · Package · PLC
0755-82221901 sales@ainstru.com

PMP50 pressure transmitter

PMP50 pressure transmitter
Cerabar PMP50
Configurable Models 20 models
Pressure transmitter with metal film for high-precision measurement of liquids and gases
Compare

Continuous / Liquids

Measuring Principle Absolute and gauge pressure
Characteristic / Application Pressure transmitter with
metal membrane for accurate
measurement in liquids or gases.
Available with diaphragm seals.
Intuitive and clear wizard for
commissioning.
Specialities zero/span button
test clamp
RFID TAG for easy identification
Supply / Communication Ex d, Ex e, non-Ex:
10.5...35V DC
Ex i: 10.5...30V DC
Accuracy Standard:
up to 0.065 %
Platinum:
up to 0.055 %
Long term stability 0.1 % of URL/ year
0.2 % of URL/ 5 years
0.25 % of URL/ 10 years
Ambient temperature -40°C...+85°C
(-40°F...+185°F)
Process temperature Standard:
-40°C…+125°C
(-40°F...+257°F)
Diaphragm seal:
-70°C...+400°C
(-94°F...+752°F)
Process connection Thread:
G1/2, NPT1/2
Flange:
DN25...DN80,
NPS 1"...3"
Communication 4...20 mA HART
Certificates / Approvals ATEX, FM, IECEx, JPN Ex,
INMETRO, NEPSI
Safety approvals SIL
Design approvals EN10204-3.1,
NACE MR0103,
NACE MR0175

Pressure

Measuring Principle Absolute and gauge pressure
Characteristic Pressure transmitter with
metal membrane for accurate
measurement in liquids or gases.
Available with diaphragm seals.
Intuitive and clear wizard for
commissioning.
Supply voltage Ex d, Ex e, non-Ex:
10.5...35V DC
Ex i: 10.5...30V DC
Reference Accuracy Standard:
up to 0.065 %
Platinum:
up to 0.05 %
Long term stability 0.1 % of URL/ year
0.2 % of URL/ 5 years
0.25 % of URL/ 10 years
Process temperature Standard:
-40°C…+125°C
(-40°F...+257°F)
Diaphragm seal:
-70°C...+400°C
(-94°F...+752°F)
Ambient temperature -40°C...+85°C
(-40°F...+185°F)
Measuring cell 1 bar...400 bar
(14.5 psi...5800 psi)
Smallest calibratable span 20 mbar (0.29 psi)
Vacuum resistance Standard:10 mbar (0.145 psi)
Diaphragm seal: 50 mbar (0.725 psi)
Max. Turn down 100:1
Max. overpressure limit 600 bar (9,000 psi)
Process connection Thread:
G1/2, NPT1/2
Flange:
DN25...DN80,
NPS 1"...3"
Material process membrane 316L, AlloyC, Gold
Material gasket PTFE
Fill fluid Silicone oil, high temperature oil,
Inert oil
Material housing Aluminum, 316L
Communication 4...20 mA HART
Certificates / Approvals ATEX, FM, IECEx, JPN Ex,
INMETRO, NEPSI
Safety approvals SIL
Design approvals EN10204-3.1,
NACE MR0103,
NACE MR0175
Specialities Zero/Span buttons,
Testclamp,
RFID TAG for easy identification

Cerabar PMP50 – pressure transmitter
Pressure transmitter with metallic diaphragm for high-precision measurement of liquids and gases
Advantages
Software wizards enable fast and simple commissioning

Device developed in accordance with IEC 61508

Flexible write protection via hardware and/or software wizards

Pre-assembled manifold (pressure and leak tested) for faster installation
Field of application
The Cerabar PMP50 is a pressure transmitter with a metallic diaphragm for measuring liquids and gases. It is optimized for standard applications and offers intuitive and safe operating functions.

Pressure measuring range: up to 400 bar (6000 psi)

Process temperature up to 400°C (752°F) when using diaphragm seals

Accuracy: up to +/-0.055%

Benefits
Software wizards enable fast and simple commissioning

Device developed in accordance with IEC 61508

Flexible write protection via hardware and/or software wizards

Pre-assembled manifold (pressure and leak tested) for faster installation

Lean selection
Standard product

Reliable, robust, low maintenance

Spare Parts & Accessories 34

34 spare parts · click to collapse
Order codeDescription
71487198Blind plug, M20, 316L, O-ring EPDMConsists of: blind plug, O-ring PTFE coated 17.0x2.0. Usage: FHX50B, FTL51B, FTL62, FTL63, FTL64, FMG50, FMR60B, FMR62B, FMR63B, FMR66B, FMR67B, NMT81, PMC51B, PMC71B, PMD50, PMD55B, PMD75B, PMD78B, PMP50, PMP51B, PMP71B.Details
71679718Blind plug NPT1/2, Alu, Ex dConsists of: Plug IEC/ATEX Ex d marking, PTFE coated. Usage: FMG50, FMR60B, FMR62B, FMR63B, FMR66B, FMR67B, FTL41, FTL51B, FTL62, FTL64, PMC51B, PMC71B, PMD50, PMD55B, PMD75B, PMD78B, PMP50, PMP51B, PMP71B.Details
71487199Plug: M20, aluminum, EPDM sealing ringConsists of: blind plug, O-ring PTFE coated 17.0x2.0. Usage: FHX50B, FTL51B, FTL62, FTL63, FTL64, FMG50, FMR60B, FMR62B, FMR63B, FMR66B, FMR67B, NMT81, PMC51B, PMC71B, PMD50, PMD55B, PMD75B, PMD78B, PMP50, PMP51B, PMP71B.Details
71679715Blind plug NPT1/2, 316L, Ex dConsists of: Plug IEC/ATEX Ex d marking, PTFE coated. Usage: FMG50, FMR60B, FMR62B, FMR63B, FMR66B, FMR67B, FTL41, FTL51B, FTL62, FTL64, PMC71B, PMD50, PMD75B, PMD78B, PMP71B, PMP50.Details
71487201Blind plug, NPT1/2, aluminumConsists of: blind plug, PTFE coated. Usage: FHX50B, FTL51B, FTL62, FTL63, FTL64, FMG50, FMR60B, FMR62B, FMR63B, FMR66B, FMR67B, NMT81, PMC51B, PMC71B, PMD50, PMD55B, PMD75B, PMD78B, PMP50, PMP51B, PMP71B.Details
71679714Blind plug M20, Alu, Ex dConsists of: Plug IEC/ATEX Ex d marking, o-ring PTFE coated 17,0x2,0. Usage: FMG50, FMP51, FMP52, FMP54, FMP55, FMP56, FMP57, FMR51, FMR52, FMR53, FMR54, FMR57, FMR60B, FMR62B, FMR63B, FMR66B, FMR67B, FTL41, FTL51B, FTL62, FTL64, OFMP50, OFMP51, OFMP57, OFMR51, OFMR54, OFTL41, OFTL51B, PMC51B, PMC71B, PMD50, PMD55B, PMD75B, PMD78B, PMP50, PMP51B, PMP71B.Details
71487203Blind plug, NPT1/2, 316LConsists of: blind plug, PTFE coated. Usage: FHX50B, FTL51B, FTL62, FTL63, FTL64, FMG50, FMR60B, FMR62B, FMR63B, FMR66B, FMR67B, NMT81, PMC51B, PMC71B, PMD50, PMD55B, PMD75B, PMD78B, PMP50, PMP51B, PMP71B.Details
71679712Blind plug M20, 316L, Ex dConsists of: Plug IEC/ATEX Ex d marking o-ring PTFE coated 17,0x2,0. Usage: FMG50, FMP51, FMP52, FMP54, FMP55, FMP56, FMP57, FMR51, FMR52, FMR53, FMR54, FMR57, FMR60B, FMR62B, FMR63B, FMR66B, FMR67B, FTL62, FTL64, OFMP51, OFMP57, OFMR51,OFMR54, OFTL51B, PMC71B, PMD50, PMD75B, PMD78B, PMP50, PMP71B.Details
71435135Cable gland M20, plastic, suitable for models: FTL41, FTL51B, PMC51B, PConsists of: Cable gland, O-ring 17.0 x 2.0 PTFE coated. Usage: FHX50B, FMB70, FMG50, FMR60B, FMR62B, FMR63B, FMR66B, FMR67B, FTL41, FTL51B, FTL62, FTL63, FTL64, PMC51B, PMC71B, PMD50, PMD55B, PMD63B, PMD75B, PMD78B, PMP50, PMP51B, PMP63B, PMP71B.Details
71435136Cable gland M20, nickel-plated brass, suitable for: FMG50, FTL41, FTL51B, PConsists of: Cable gland, O-ring 17.0 x 2.0 PTFE coated. Usage: FHX50B, FMG50, FMR60B, FMR62B, FMR63B, FMR66B, FMR67B, FTL41, FTL51B, FTL62, FTL63, FTL64, PMC51B, PMC71B, PMD55B, PMD75B, PMD78B, PMP51B, PMP71B.Details
71435137Cable gland M20, 316L, suitable for models: FMG50, FTL51B, PMC51BConsists of: Cable gland, O-ring 17.0 x 2.0 PTFE coated. Usage: FHX50B, FMG50, FMR60B, FMR62B, FMR63B, FMR66B, FMR67B, FTL51B, FTL62, FTL63, FTL64, PMC51B, PMC71B, PMD55B, PMD75B, PMD78B, PMP51B, PMP71B.Details
71474251Enclosure grounding terminal L+PConsists of: spring washer, screw, earth handle, clamp. Usage Liquiphant: Only for digital communication. Usage: FHX50B, FMG50, PMD50, PMD55B, PMD63B, PMD75B, PMD78B, PMC51B, PMC71B, PMP50, PMP51B, PMP63B, PMP71B, FMR60B, FMR62B, FMR63B, FMR66B, FMR67B, FTL51B, FTL62, FTL63, FTL64.Details
71652584Cover Alu, closedSealing ring: HNBR. Usage: PMD50, PMP50.Details
71652584Cover Alu, closedSealing ring: HNBR. Usage: PMD50, PMP50.Details
71652585Cover, aluminium, with sight glassSight glass: Soda lime glass. Usage: PMD50, PMP50.Details
71652587Housing cover, aluminum, flameproofSealing ring: HNBR. Usage: PMD50, PMP50.Details
71652587Housing cover, aluminum, flameproofSealing ring: HNBR. Usage: PMD50, PMP50.Details
71652606Aluminium cover, sight glass, Ex dSealing ring: HNBR. Sight glass: Soda lime glass. Usage: PMD50, PMP50.Details
71652609Cover red Alu, sight glass, Ex dSealing ring: HNBR. Sight glass: Soda lime glass. Usage: PMD50, PMP50.Details
71652610Cover 316L, sight glass, Ex dSealing ring: HNBR. Sight glass: Soda lime glass. Usage: PMD50, PMP50.Details

Configurator / Comparison

ConfigureEnter a product model to configure
CompareEnter the product models to compare — up to three products

   

   

 

 

 

Source:Endress+Hauserpublic information from the website

Measuring principle video

Click to play
260 models · click to collapse
All documents for this model
4 documents · click to expand
BA

BA Cerabar PMP50 Chinese

Rev. 0326 64 pages
Preview
BA

BA Cerabar PMP50 English

Rev. 0326 72 pages
Preview
TI

TI Cerabar PMP50 Chinese

Rev. 0426 48 pages
Preview
TI

TI Cerabar PMP50 English

Rev. 0426 48 pages
Preview
Municipal Wastewater Treatment

Municipal Wastewater Treatment · Gauge and Absolute Pressure — Application Case

Site Conditions

This is a Municipal Wastewater Treatment customer whose plant is at the inlet, biological and outlet stages of a municipal plant. The main task was measuring gauge or absolute pressure to judge process state.

The Problem

The difficulty lies in high solids and corrosive fractions, with large diurnal flow swings; at the same time effluent parameters drift and probes foul with sludge and biofilm. Specification therefore means selected by turndown, overpressure protection and diaphragm material.

What We Did

We took a pressure range approach — pressure solution matched to process pressure and temperature. Delivery was built around selected by turndown, overpressure protection and diaphragm material, with point-by-point checks on: stable compliance with Grade 1A discharge limits and online data reporting.

Operating Outcome

In service the system performs as intended — spotting abnormal pressure early, before equipment is damaged, with routine maintenance falling noticeably.

Food & Beverage

Food & Beverage · Gauge and Absolute Pressure — Application Case

Application Context

dosing, pasteurising, filling and CIP stages is a typical Food & Beverage duty. The project started from Gauge and Absolute Pressure, focusing on measuring gauge or absolute pressure to judge process state.

Process Constraints

frequent cleaning demands instruments that withstand hot CIP/SIP caustic cycles, and syrups, dairy and beverages that scale easily, under strict hygienic requirements only amplify it. The Gauge and Absolute Pressure specification must therefore cover selected by turndown, overpressure protection and diaphragm material.

Delivery Approach

Delivery was handled as the pressure range — pressure solution matched to process pressure and temperature. Execution centred on selected by turndown, overpressure protection and diaphragm material, together with food-safety and hygienic design per 3-A / EHEDG.

Field Results

Operation shows the benefit clearly — spotting abnormal pressure early, before equipment is damaged, with confidence in this measuring point improving markedly.

Chemical & Pharmaceutical

Chemical & Pharmaceutical · Gauge and Absolute Pressure — Application Case

Customer Profile

The typical setting in Chemical & Pharmaceutical is reactors, metering vessels and finished-product tanks. The measuring point is Gauge and Absolute Pressure; the focus is measuring gauge or absolute pressure to judge process state.

Key Difficulties

Conditions are highly corrosive media with solvents and crystallisation, partly in explosive atmospheres and, on top of that, corrosion plus Ex protection shortens instrument life and makes downtime costly, which a standard configuration cannot handle for long. The design therefore starts with selected by turndown, overpressure protection and diaphragm material.

Implementation

The scope ran through the pressure range: pressure solution matched to process pressure and temperature. Technically it hinges on selected by turndown, overpressure protection and diaphragm material, and on these requirements: Ex certification and GMP traceability of records.

Customer Benefit

In practice this means spotting abnormal pressure early, before equipment is damaged, which cut the workload on the site maintenance team.

Power & Energy

Power & Energy · Gauge and Absolute Pressure — Application Case

Project Background

The project covers boiler feedwater, turbine condensate and FGD/DeNOx systems. The work centres on Gauge and Absolute Pressure and the core task is measuring gauge or absolute pressure to judge process state.

Operating Challenges

The medium and duty are characterised by high temperature and pressure with both steam and saline water, and drift under high temperature and pressure, with very costly unplanned outages. So Gauge and Absolute Pressure cannot be specified generically — it calls for selected by turndown, overpressure protection and diaphragm material.

Solution

The Ainstru engineering team worked through the pressure range: pressure solution matched to process pressure and temperature. The configuration involves selected by turndown, overpressure protection and diaphragm material, with these requirements in view: compliance with power-industry codes and unit availability targets.

Results

Since start-up the system has run steadily; the practical result is spotting abnormal pressure early, before equipment is damaged.