Instrumentation · Engineering · Package · PLC
0755-82221901 sales@ainstru.com

HI801-01

HI801-01
iris® Spectrophotometer
This is a spectrophotometer with a wavelength range of 340 to 900 nm, a split beam optical system and a tungsten halogen lamp.

Overview

This is a spectrophotometer with a wavelength range of 340 to 900 nm, a split beam optical system and a tungsten halogen lamp. It measures transmittance (%), absorbance and concentration, and supports 10 mm square, 50 mm rectangular, 16 mm round, 22 mm round and 13 mm round (vial) sample cells. Wavelength calibration is internal and automatic at power-on with visual feedback; up to 9999 measured values can be stored, with export capability in csv and pdf file formats.

Specifications

ParameterValue
SKUHI801-01
Wavelength Range340 to 900 nm
Wavelength Resolution1 nm
Wavelength Accuracy±1.5 nm
Photometric Range0.000 to 3.000 Abs
Photometric Accuracy5 mAbs at 0.000 to 0.500 Abs, 1 % at 0.500 to 3.000 Abs
Measurement Modetransmittance (%), absorbance and concentration
Sample Cell10 mm square, 50 mm rectangular, 16 mm round, 22 mm round, 13 mm round (vial)
Wavelength Selectionautomatic, based on the selected method (editable for user methods only)
Light Sourcetungsten halogen lamp
Optical Systemsplit beam
Wavelength Calibrationinternal, automatic at power-on with visual feedback
Stray Light<0.1 % T at 340 nm with NaNO2
Spectral Bandwidth5 nm
Number of Methodsup to 150 factory (96 pre-loaded), up to 100 user
Data Points Stored9999 measured values
Connectivity1x USB A (mass storage host), 1x USB B (mass storage device)
Battery Life3000 measurements or 8 hours
Power Supply15 Vdc power adapter, 10.8 Vdc Li-ion rechargeable battery
Environment0 to 50 °C (32 to 122 °F), 0 to 95% RH
Dimensions155 x 205 x 322 mm (6.1 x 8.0 x 12.6")
Weight3 kg (6.6 Ibs.)