Instrumentation · Engineering · Package · PLC
0755-82221901 sales@ainstru.com

CA76NA Sodium Ion Analyzer

CA76NA Sodium Ion Analyzer
Sodium Analyzer CA76NA
Potentiometric analysis system for boiler feed water, steam and condensate, and ion exchanger monitoring
Compare

Analyser

Measuring Principle Potentiometric / ISE
Characteristic Analyser for sodium in boiler feedwater, steam, condensate and ion exchangers
Size 850 x 450 x 250 mm
33.5 x 17.7 x 9.8 inch
Design Panel stainless steel 1.4301
Process temperature 10 to 40 °C
(50 to 104 °F)
Ambient temperature 5 to 40 °C
(41 to 104 °F)
Process pressure 1 to 5 bar (abs)
14.5 to 72.5 psi
Sample flow rate 10 to 15 l/h
2.64 to 3.96 gal/h
Application Online sodium monitoring in boiler feed water, water steam cycles, condensate and ion exchangers
Power supply 100 to 240 V AC, 50/60 Hz
Output / communication 0/4 to 20 mA
Measuring range 0.1 to 9999 µg/l (ppb) Na
Consumables CA76NA sodium electrode,
CA76NA pH combination electrode
CA76NA standard solution 5100 ppb Na

Protect your water-steam circuit and ion exchangers with the CA76NA sodium analyzer! If the sodium ion concentration in high-purity feed water or boiler feed water rises, harmful impurities have dissolved in the water. This can cause severe equipment damage and expensive repair costs. The CA76NA performs highly accurate water quality monitoring and quickly detects critical events. The sodium analyzer helps you develop a predictive maintenance strategy and ensure normal power plant operation.

Benefits

  • Protects expensive plant equipment used in the power industry: the CA76NA sodium analyzer monitors low sodium ion concentrations in the water-steam circuit and detects residues and impurities in the power plant.

  • Online measurement of sodium ion concentration in the pipe downstream of the vacuum pump provides early warning of condenser leaks.

  • Low dosing of DIPA reagent and standard solution reduces operating costs.

  • Precise measurement: independent pH reference electrode with high-precision temperature compensation optimizes pH adjustment.

  • Optimized investment costs: one analyzer with up to six sampling channels.

  • Perfectly adapted to process conditions: dedicated sampling channels with freely adjustable measuring intervals.

Applications

  • In the power industry and utilities, the CA76NA sodium analyzer monitors low sodium ion concentrations:

  • Indicates water and steam purity in boilers and turbines.

  • Provides early warning of condenser leaks or ion exchanger failure.

  • Controls the inlet water quality of power plants using fresh water as the water source.

Spare Parts & Accessories 25

25 spare parts · click to collapse
Order codeDescription
71358762CA76NA Accessories, Commissioning AccessoriesDetails
71358110CA76NA Accessory Sodium ElectrodeDetails
71358111CA76NA accessory pH combination electrodeDetails
71371663CA76NA electrodeDetails
71358761CA76NA Accessory, 5100 ppb Sodium Standard SolutionDetails
71358112CA76NA Accessory, Mixing UnitDetails
71583317CA76NA Mixing Unit V2.0Details
71358114CA76xx accessory: filter connection, PEF50, 50 µmDetails
71583288CA76NA analyzer sample inlet filterDetails
71358118CA76NA Accessory, 24 V DC Alkalization PumpDetails
71583343CA76NA Alkalization Pump V2.0Details
71358122CA76NA Accessory 24V DC Standard Solution PumpDetails
71583346CA76NA Standard Solution Pump V2.0Details
71358125CA76NA accessory: 24 V DC circulation pumpDetails
71583339CA76NA circulation pump V2.0Details
71358127CA76xx accessory, 24V 2-way 3-port solenoid valveDetails
71583311CA76NA valve (valve manifold unit)Details
71358129CA76NA Accessory 24V 2/2-Way Solenoid ValveDetails
71583305CA76NA valve (regulating unit)Details
71358130CA76NA Accessories Small Component KitDetails

Configurator / Comparison

ConfigureEnter a product model to configure
CompareEnter the product models to compare — up to three products

Source: Endress+Hauserpublic information from the website

All documents for this model
4 documents · click to expand
BA

BA CA76NA Chinese

Rev. 0525 64 pages
Preview
BA

BA CA76NA English

Rev. 0525 72 pages
Preview
TI

TI CA76NA Chinese

Rev. 0525 16 pages
Preview
TI

TI CA76NA English

Rev. 0525 16 pages
Preview
Municipal Wastewater Treatment

Municipal Wastewater Treatment · EH-ONLINE — Application Case

Project Overview

The site is the inlet, biological and outlet stages of a municipal plant. The customer wanted EH-ONLINE properly configured on the existing plant, with the aim of continuous water monitoring on the E+H online platform.

Technical Challenges

What actually causes trouble is high solids and corrosive fractions, with large diurnal flow swings, and effluent parameters drift and probes foul with sludge and biofilm. All of this means the EH-ONLINE configuration has to involve configured under one numbering scheme across platform protocols and sensors.

Configuration

Supply and installation were organised through the online water-analysis line: integrating around each online monitoring point. On site this meant configured under one numbering scheme across platform protocols and sensors, checked line by line against: stable compliance with Grade 1A discharge limits and online data reporting.

Performance

After commissioning the outcome is data flows through the platform, improving service and diagnostics, and the readings are now used for process tuning.

Food & Beverage

Food & Beverage · EH-ONLINE — Application Case

Site Conditions

This is a Food & Beverage customer whose plant is at dosing, pasteurising, filling and CIP stages. The main task was continuous water monitoring on the E+H online platform.

The Problem

The difficulty lies in syrups, dairy and beverages that scale easily, under strict hygienic requirements; at the same time frequent cleaning demands instruments that withstand hot CIP/SIP caustic cycles. Specification therefore means configured under one numbering scheme across platform protocols and sensors.

What We Did

We took a online water-analysis line approach — integrating around each online monitoring point. Delivery was built around configured under one numbering scheme across platform protocols and sensors, with point-by-point checks on: food-safety and hygienic design per 3-A / EHEDG.

Operating Outcome

In service the system performs as intended — data flows through the platform, improving service and diagnostics, with routine maintenance falling noticeably.

Chemical & Pharmaceutical

Chemical & Pharmaceutical · EH-ONLINE — Application Case

Application Context

reactors, metering vessels and finished-product tanks is a typical Chemical & Pharmaceutical duty. The project started from EH-ONLINE, focusing on continuous water monitoring on the E+H online platform.

Process Constraints

corrosion plus Ex protection shortens instrument life and makes downtime costly, and highly corrosive media with solvents and crystallisation, partly in explosive atmospheres only amplify it. The EH-ONLINE specification must therefore cover configured under one numbering scheme across platform protocols and sensors.

Delivery Approach

Delivery was handled as the online water-analysis line — integrating around each online monitoring point. Execution centred on configured under one numbering scheme across platform protocols and sensors, together with Ex certification and GMP traceability of records.

Field Results

Operation shows the benefit clearly — data flows through the platform, improving service and diagnostics, with confidence in this measuring point improving markedly.

Power & Energy

Power & Energy · EH-ONLINE — Application Case

Customer Profile

The typical setting in Power & Energy is boiler feedwater, turbine condensate and FGD/DeNOx systems. The measuring point is EH-ONLINE; the focus is continuous water monitoring on the E+H online platform.

Key Difficulties

Conditions are high temperature and pressure with both steam and saline water and, on top of that, drift under high temperature and pressure, with very costly unplanned outages, which a standard configuration cannot handle for long. The design therefore starts with configured under one numbering scheme across platform protocols and sensors.

Implementation

The scope ran through the online water-analysis line: integrating around each online monitoring point. Technically it hinges on configured under one numbering scheme across platform protocols and sensors, and on these requirements: compliance with power-industry codes and unit availability targets.

Customer Benefit

In practice this means data flows through the platform, improving service and diagnostics, which cut the workload on the site maintenance team.