Technical Data
CompareConcentration
| Measuring Principle | UV absorbance |
| Application | Measurement of protein concentrations, chromatography monitoring, filtration monitoring, concentration measurement of organic compounds, detection of aromates |
| Characteristic | Sensor for installation in the pipe with separate flow assembly OUA260 |
| Measuring range | 0 to 2.5 AU 0 to 50 OD (depending on optical pathlength) |
| Material | Sensor housing: stainless steel 316 |
| Process temperature | 0 to 90 °C (32 to 194 °F) continuous Max. 130 °C (266 °F) for 2 hours |
| Process pressure | Max. 100 bar (1450 psi) absolute, depending on the material, pipe size and process connection of the flow assembly |
| Ex certification | ATEX, FM |
| Ingress protection | IP65 (NEMA4) |
| Certificates, declarations and approvals | FDA |
The OUSAF44 UV absorbance sensor monitors product concentration in process media. Excellent filter performance ensures the highest linearity and constant agreement with laboratory results. It provides you with fast, reliable process information and increases product yield. The patented EasycalTM system enables NIST-traceable online dry calibration. The OUSAF44 is explosion-protection approved for use in hazardous areas (ATEX, FM) and is suitable for hygienic processes (CIP/SIP cleaning).
Benefits
Fast, reliable monitoring of product concentration for better process control and simplified quality control
Measured values in close agreement with laboratory results
The patented EasycalTM system enables NIST-traceable online dry calibration
Long sensor service life, stable operation and low maintenance
Excellent filter performance for the highest linearity
Resistant to CIP and SIP cleaning
FM and ATEX approval for hazardous area applications
Applications
The OUSAF44 sensor is used for UV absorbance measurement:
Protein concentration measurement
Chromatography control
Permeate and retentate monitoring
Organic compound concentration measurement
Aromatic hydrocarbon detection
Filter monitoring
International explosion-protection approvals, such as ATEX and FM, for use in hazardous areas
Spare Parts & Accessories 9
9 spare parts · click to collapse
| Order code | Description | |
|---|---|---|
| 71108535 | OUSAF44 lamp 280/295/302/313/365 nmFor OUSAF44, 46 ATEX and Non-Exp sensors | Details |
| 71108543 | OUSAF44 Reference Rod | Details |
| 71210149 | KIT OUSAF44 EasyCal Retrofit 254 nm | Details |
| 71210150 | KIT OUSAF44 EasyCal Retrofit 280 nm | Details |
| 71210151 | KIT OUSAF44 EasyCal Retrofit 313 nm | Details |
| 71210152 | KIT OUSAF44 EasyCal Retrofit 365 nm | Details |
| 71210153 | KIT OUSAF44 EasyCal Retrofit 302 nm | Details |
| 71210156 | KIT OUSAF44 EasyCal Retrofit 295 nm | Details |
| 71210161 | OUSxFxx Dust Cover Connector | Details |
Configurator / Comparison
Source: Endress+Hauserpublic information from the website
Measuring principle video
Click to playManuals & Documents 5
All documents for this modelProduct Videos
Case Studies
Municipal Wastewater Treatment · multiparameter — Application Case
Application Context
the inlet, biological and outlet stages of a municipal plant is a typical Municipal Wastewater Treatment duty. The project started from multiparameter, focusing on monitoring several water parameters at one measuring point.
Process Constraints
effluent parameters drift and probes foul with sludge and biofilm, and high solids and corrosive fractions, with large diurnal flow swings only amplify it. The multiparameter specification must therefore cover configured by parameter set, mounting and cleaning needs.
Delivery Approach
Delivery was handled as the analyser range — selecting as part of a complete water-analysis system. Execution centred on configured by parameter set, mounting and cleaning needs, together with stable compliance with Grade 1A discharge limits and online data reporting.
Field Results
Operation shows the benefit clearly — one point replacing several single-parameter units, saving capex and upkeep, with confidence in this measuring point improving markedly.
Food & Beverage · multiparameter — Application Case
Customer Profile
The typical setting in Food & Beverage is dosing, pasteurising, filling and CIP stages. The measuring point is multiparameter; the focus is monitoring several water parameters at one measuring point.
Key Difficulties
Conditions are syrups, dairy and beverages that scale easily, under strict hygienic requirements and, on top of that, frequent cleaning demands instruments that withstand hot CIP/SIP caustic cycles, which a standard configuration cannot handle for long. The design therefore starts with configured by parameter set, mounting and cleaning needs.
Implementation
The scope ran through the analyser range: selecting as part of a complete water-analysis system. Technically it hinges on configured by parameter set, mounting and cleaning needs, and on these requirements: food-safety and hygienic design per 3-A / EHEDG.
Customer Benefit
In practice this means one point replacing several single-parameter units, saving capex and upkeep, which cut the workload on the site maintenance team.
Chemical & Pharmaceutical · multiparameter — Application Case
Project Background
The project covers reactors, metering vessels and finished-product tanks. The work centres on multiparameter and the core task is monitoring several water parameters at one measuring point.
Operating Challenges
The medium and duty are characterised by highly corrosive media with solvents and crystallisation, partly in explosive atmospheres, and corrosion plus Ex protection shortens instrument life and makes downtime costly. So multiparameter cannot be specified generically — it calls for configured by parameter set, mounting and cleaning needs.
Solution
The Ainstru engineering team worked through the analyser range: selecting as part of a complete water-analysis system. The configuration involves configured by parameter set, mounting and cleaning needs, with these requirements in view: Ex certification and GMP traceability of records.
Results
Since start-up the system has run steadily; the practical result is one point replacing several single-parameter units, saving capex and upkeep.
Power & Energy · multiparameter — Application Case
Project Overview
The site is boiler feedwater, turbine condensate and FGD/DeNOx systems. The customer wanted multiparameter properly configured on the existing plant, with the aim of monitoring several water parameters at one measuring point.
Technical Challenges
What actually causes trouble is high temperature and pressure with both steam and saline water, and drift under high temperature and pressure, with very costly unplanned outages. All of this means the multiparameter configuration has to involve configured by parameter set, mounting and cleaning needs.
Configuration
Supply and installation were organised through the analyser range: selecting as part of a complete water-analysis system. On site this meant configured by parameter set, mounting and cleaning needs, checked line by line against: compliance with power-industry codes and unit availability targets.
Performance
After commissioning the outcome is one point replacing several single-parameter units, saving capex and upkeep, and the readings are now used for process tuning.