Technical Data
CompareThermometer
| Measuring Principle | |
| Characteristic / Application | metric style modular temperature assembly fast response time (reduced/tapered tip) threaded process connection without neck |
| Thermowell / protection tube | without (not intended to use with thermowell) |
| Insert / probe | mineral insulated (MI), flexible |
| Outer diameter protection tube / Insert | 6,0 mm (0,24'') 9,0 mm (0,35'') |
| Max. immersion length on request | up to 10.000,00 mm (393,70'') |
| Material protection tube/ thermowell | 1.4404 (316L) |
| Process connection | male thread: G1/2" G3/4" G1" NPT1/2" NPT3/4" NPT1" M20x1.5 |
| Tip shape | reduced |
| Surface roughness Ra | 0,8 μm (31,5 μin.) |
| Operating temperature range | PT 100: -50 °C ... 600 °C (-58 °F ...1.112 °F) |
| Max. process pressure (static) | at 20 °C: 20 bar (290 psi) |
| Accuracy | class A acc. to IEC 60751 class AA acc. to IEC 60751 |
| Response time | depending on configuration t50 = 3 s t90 = 6 s |
| Integration head transmitter | yes (4 … 20 mA; HART; PROFIBUS PA; FOUNDATION FIELDBUS) |
Product Overview
The thermometer is mainly used in the chemical industry and is also suitable for applications in other industries. Devices with a head-mounted transmitter are ready for immediate use and offer higher measuring accuracy and reliability than directly wired sensors. A wide range of process connections, versions and materials provides flexibility in use.
-
Benefits
-
Modular design with a standard terminal head conforming to DIN EN 50446 and customer-specific immersion length for high flexibility
-
Highly compatible insert, designed in accordance with DIN 43772
-
Short tapered/tapered tip for faster response time
-
Easy selection of the head transmitter: analog output 4 to 20 mA, HART®, PROFIBUS® PA or FF fieldbus
Spare Parts & Accessories 22
22 spare parts · click to collapse
| Order code | Description | |
|---|---|---|
| TMT82- | Temperature transmitter TMT822-wire temperature transmitter. Dual universal sensor input. Galvanic isolation 2kV (input/output). Application: RTD, TC, Ohm and mV. Presetting failure current high. Consider additional marine certificate requirements on application side. ::Highly reliable due to drift detection and sensor back up. ::Optional also for safety critical applications by means of SIL 2/3 acc. to IEC61508. | Enquire |
| TMT84- | Temperature transmitter TMT84Dual sensor input. Advanced diagnostics: sensor monitoring. Galvanic isolation 2kV. Application: RTC, TC, Ohm, mV. Current consumption: 11 mA. Mounting: head form B, DIN43729. Factory setup: CH1: Pt100, 3-wire, oC CH2: inactive Differing configuration optional selectable. | Enquire |
| TMT85- | Temperature transmitter TMT85Temperature transmitter with dual sensor input and advanced diagnostics for sensor monitoring. FOUNDATION Fieldbus protocol (ITK 5.0). Galvanic isolation 2 kV (sensor input to the output). Application: RTC, TC, Ohm, mV. Current consumption: 11 mA. Mounting: head form B, DIN43729. UL listed, CSA General Purpose. | Enquire |
| TXU10- | Instrument Commissioning Kit TXU10用于可以PC编程调试的设备 程序设置+用于带USB接口的PC的连接线 | Enquire |
| 71044061 | equipment package2x screws, 2x springs, 2x spring rings, 1x sealing CDI-connector | Details |
| TMT82- | Temperature transmitter TMT822-wire temperature transmitter. Dual universal sensor input. Galvanic isolation 2kV (input/output). Application: RTD, TC, Ohm and mV. Presetting failure current high. Consider additional marine certificate requirements on application side. ::Highly reliable due to drift detection and sensor back up. ::Optional also for safety critical applications by means of SIL 2/3 acc. to IEC61508. | Enquire |
| TMT84- | Temperature transmitter TMT84Dual sensor input. Advanced diagnostics: sensor monitoring. Galvanic isolation 2kV. Application: RTC, TC, Ohm, mV. Current consumption: 11 mA. Mounting: head form B, DIN43729. Factory setup: CH1: Pt100, 3-wire, oC CH2: inactive Differing configuration optional selectable. | Enquire |
| TMT85- | Temperature transmitter TMT85Temperature transmitter with dual sensor input and advanced diagnostics for sensor monitoring. FOUNDATION Fieldbus protocol (ITK 5.0). Galvanic isolation 2 kV (sensor input to the output). Application: RTC, TC, Ohm, mV. Current consumption: 11 mA. Mounting: head form B, DIN43729. UL listed, CSA General Purpose. | Enquire |
| TID10- | Temperature transmitter display TID10to plug on. Dot-Matrix measured value display. 12 DIP-switches buttom side. Settings: FF: Simulation, Write Lock. PA: Bus address, Write Lock. Compatible: TMT85, TMT84. | Enquire |
| 71044062 | equipment package2x screws, 1x sealing CDI-connector | Details |
| TMT182- | *** module temperature transmitter TMT182Temperature transmitter, Protocol HART. Application: RTD, TC, Ohm, mV. 2-wire 4-20mA, SIL2, galvanic isolation Fault reaction: NAMUR NE 43. Mounting: head form B, DIN43729. UL listed. GL (German Llyod) Marine. | Enquire |
| TXU10- | Instrument Commissioning Kit TXU10用于可以PC编程调试的设备 程序设置+用于带USB接口的PC的连接线 | Enquire |
| READWIN- | Readwin 2000 Commissioning Package READWINReadwin 2000 Operation + Read. software | Enquire |
| FXA195- | Communication interface FXA195Modem to connect HART field devices to a computer via USB. Commissioning of field devices with software tool. :: To connect to a intrinsically safe signal circuit. :: SIL2, IEC61508. | Enquire |
| TA30A- | Terminal box TA30ARetainer for terminal block or head transmitter + optional display. Cover: low, hinged, optional window. Material: aluminium. Coating: polyester powder-coating. Colour: body blue, cover grey. Ambient temp.: -50...150 oC (housing). Protection: IP66/68, NEMA 4X (housing). | Enquire |
| TA30D- | Junction box TA30DRetainer for 2 head transmitters or terminal blocks. Cover: high, hinged. Material: aluminium. Coating: polyester powder-coating. Colour: body blue, cover grey. Ambient temp.: -50...150 oC (housing). Protection: IP66/68, NEMA 4X (housing). | Enquire |
| TA20B- | Terminal box TA20BConnecting head to mount a measuring insert with head transmitter or a connection socket. To be mounted on an armature or protective sleeve. Material:Polyamid black Ambient temperature: max. +80 oC Protection: IP65 | Enquire |
| 71367414 | Copper sealing ringPacking unit: 10 pieces | Details |
| 71367420 | Copper sealing ringPacking unit: 10 pieces | Details |
| 71367425 | Copper sealing ringPacking unit: 10 pieces | Details |
Configurator / Comparison
Source:Endress+Hauserwebsite public information
Manuals & Documents 2
All documents for this modelCase Studies
Municipal Wastewater Treatment · thermistor — Application Case
Project Overview
The site is the inlet, biological and outlet stages of a municipal plant. The customer wanted thermistor properly configured on the existing plant, with the aim of high-accuracy temperature measurement with platinum resistance elements.
Technical Challenges
What actually causes trouble is high solids and corrosive fractions, with large diurnal flow swings, and effluent parameters drift and probes foul with sludge and biofilm. All of this means the thermistor configuration has to involve choosing Pt100 or Pt1000 by accuracy class, wiring system and insertion depth.
Configuration
Supply and installation were organised through the temperature range: temperature solution matched to measuring point and response needs. On site this meant choosing Pt100 or Pt1000 by accuracy class, wiring system and insertion depth, checked line by line against: stable compliance with Grade 1A discharge limits and online data reporting.
Performance
After commissioning the outcome is clear accuracy advantage at low temperatures, giving tighter process windows, and the readings are now used for process tuning.
Food & Beverage · thermistor — Application Case
Site Conditions
This is a Food & Beverage customer whose plant is at dosing, pasteurising, filling and CIP stages. The main task was high-accuracy temperature measurement with platinum resistance elements.
The Problem
The difficulty lies in syrups, dairy and beverages that scale easily, under strict hygienic requirements; at the same time frequent cleaning demands instruments that withstand hot CIP/SIP caustic cycles. Specification therefore means choosing Pt100 or Pt1000 by accuracy class, wiring system and insertion depth.
What We Did
We took a temperature range approach — temperature solution matched to measuring point and response needs. Delivery was built around choosing Pt100 or Pt1000 by accuracy class, wiring system and insertion depth, with point-by-point checks on: food-safety and hygienic design per 3-A / EHEDG.
Operating Outcome
In service the system performs as intended — clear accuracy advantage at low temperatures, giving tighter process windows, with routine maintenance falling noticeably.
Chemical & Pharmaceutical · thermistor — Application Case
Application Context
reactors, metering vessels and finished-product tanks is a typical Chemical & Pharmaceutical duty. The project started from thermistor, focusing on high-accuracy temperature measurement with platinum resistance elements.
Process Constraints
corrosion plus Ex protection shortens instrument life and makes downtime costly, and highly corrosive media with solvents and crystallisation, partly in explosive atmospheres only amplify it. The thermistor specification must therefore cover choosing Pt100 or Pt1000 by accuracy class, wiring system and insertion depth.
Delivery Approach
Delivery was handled as the temperature range — temperature solution matched to measuring point and response needs. Execution centred on choosing Pt100 or Pt1000 by accuracy class, wiring system and insertion depth, together with Ex certification and GMP traceability of records.
Field Results
Operation shows the benefit clearly — clear accuracy advantage at low temperatures, giving tighter process windows, with confidence in this measuring point improving markedly.
Power & Energy · thermistor — Application Case
Customer Profile
The typical setting in Power & Energy is boiler feedwater, turbine condensate and FGD/DeNOx systems. The measuring point is thermistor; the focus is high-accuracy temperature measurement with platinum resistance elements.
Key Difficulties
Conditions are high temperature and pressure with both steam and saline water and, on top of that, drift under high temperature and pressure, with very costly unplanned outages, which a standard configuration cannot handle for long. The design therefore starts with choosing Pt100 or Pt1000 by accuracy class, wiring system and insertion depth.
Implementation
The scope ran through the temperature range: temperature solution matched to measuring point and response needs. Technically it hinges on choosing Pt100 or Pt1000 by accuracy class, wiring system and insertion depth, and on these requirements: compliance with power-industry codes and unit availability targets.
Customer Benefit
In practice this means clear accuracy advantage at low temperatures, giving tighter process windows, which cut the workload on the site maintenance team.
Related Products
Endress+HauseriTHERM ModuLine TM101
Endress+HauseriTHERM ModuLine TM131
Endress+HauserRTD Thermometer TR10
Endress+HauseriTHERM ModuLine TM111
Endress+HauseriTHERM CompactLine TM311
Endress+HauserCompact Thermometer TMR31
Endress+HauserRTD Thermometer TR24
Endress+Hauser