Instrumentation · Engineering · Package · PLC
0755-8222 1901 sales@ainstru.com

TM311 热电阻温度计

TM311  热电阻温度计

iTHERM CompactLine TM311

Configurable Models 3 models
Our compact Pt100 temperature sensor iTHERM TM311 with IO-Link provides highly accurate temperature measurements. Learn more!
Compare

Thermometer

Measuring Principle Resistance Temperature Detector
Characteristic / Application metric and imperial style
compact temperature probe
fast response time
high accuracy
standard and hygienic design
hygienic process connection
threaded process connection
Thermowell / protection tube without
incl. thermowell
incl. Tee-/elbow thermowell
Insert / probe pipe version, isolated wires, not flexible
Outer diameter protection tube / Insert 3,0 mm (0,12'')
6,0 mm (0,24'')
Max. immersion length on request up to 600,00 mm (23,62'')
Material protection tube/ thermowell 1.4435 Delta-Ferrit < 1%
316L
Process connection compression fitting
Thread
weld in adapter
clamp connections acc. to ISO2852
screwed pipe joint acc. to DIN 11851
aseptic screwed pipe joint acc. to DIN 11864-1
metallic sealing system
thread acc. to ISO 228 for liquiphant adaptor
Varivent
SMS 1147
Tee-/elbow thermowell
Tip shape straight
reduced
Surface roughness Ra 0,76 μm (29,92 μin)
0,38 μm (14,96 μin)
0,38 μm (14,96 μin) electropolished
Operating temperature range PT 100:
-50 °C ...200 °C
(-58 °F ...392 °F)
Max. process pressure (static) at 20 °C: 50 bar (725 psi)
Accuracy class A acc. to IEC 60751
Response time t50 = 1 s
t90 = 1.5 s
Integration head transmitter no (4…20mA Signal, IO-Link)
Ex - approvals CSA C/US
Certification 3A, EHEDG, ASME BPE, FDA, TSE
1935/2004, 2023/2006, 10/2011
DNV, BV
RCM

Benefits
Fast installation and simple commissioning:
• Compact design, small dimensions, all stainless steel
• M12 connector, IP69 degree of protection, convenient wiring
• Universal output signals: IO-Link (self-check connection) and 4 … 20 mA (Pt100 four-wire connection)
• Models with preset measuring ranges can be ordered
• Recommended insertion lengths guarantee optimum results and the highest level of standardization
Innovative sensor technology, excellent measuring performance:
• Extremely short response time
• High accuracy even with small insertion depths
• Improved accuracy through sensor-transmitter matching
Safe measurement thanks to numerous approvals:
• Safety requirements in accordance with EN 61010-1 and CSA C/US
• Electromagnetic compatibility per NAMUR NE21
• Diagnostic information classified per NAMUR NE43
• Hygienic design per 3-A, EHEDG, ASME BPE, FDA, EC 1935/2004, EN
2023/2006, TSE/ADI, GB4806-2016 and GB9685-2016
• DNV (Det Norske Veritas) approval

Application
• Designed for hygienic and sterile applications in the food & beverage and pharmaceutical industries, meeting the highest standardization requirements of plant manufacturers and skid
builders
• Measuring range: –50 … +200 °C (–58 … +392 °F)
• Maximum pressure: 50 bar (725 psi)
• Degree of protection: IP69
• Output signals
• Without electronics: Pt100 (four-wire connection)
• With electronics: IO-Link, 4 … 20 mA current output, 1 x PNP switch output (depending on
wiring)

 

E+H product model configurator

Please enter the product model to configure

Product comparison

Please enter the product models to be compared in the input boxes below; up to three products can be compared

 

   

Technical information

   

Technical data

 

 

Documents/Manuals/Software

 

 

Accessories/Spare parts

 

   

Measuring principle video

Source: public information from the Endress+Hauser website

Model Name
TM311-18P9/0 热电阻温度计
TM311-AAB0BH2BAB4A1JA 热电阻温度计
TM311-AAB2BH1BBX1B1JA 热电阻温度计
Municipal Wastewater Treatment

Municipal Wastewater Treatment — Wastewater Treatment Plant · RTD Application Case

Project Background

Municipal Wastewater Treatment customers at the Wastewater Treatment Plant stage have long faced the challenge of insufficient accuracy in effluent quality monitoring and instruments prone to contamination causing data drift, affecting compliance discharge, placing higher demands on the stability and measurement accuracy of RTD.

Solution

Based on the on-site conditions, the Ainstru technical team selected suitable RTD for the customer and delivered one-stop service covering selection, installation, commissioning and tuning.

Results

  • Measurement accuracy and response speed significantly improved, with stable and reliable data;
  • Instruments adapt to complex Municipal Wastewater Treatment conditions, greatly reducing maintenance;
  • Helping customers achieve continuous online monitoring and cost reduction, highly recognized by customers.
Food & Beverage

Food & Beverage — Food Production Line · RTD Application Case

Project Background

Food & Beverage customers at the Food Production Line stage have long faced the challenge of high hygienic measurement requirements, with inline instruments needing to withstand CIP/SIP cleaning while maintaining stable accuracy, placing higher demands on the stability and measurement accuracy of RTD.

Solution

Based on the on-site conditions, the Ainstru technical team selected suitable RTD for the customer and delivered one-stop service covering selection, installation, commissioning and tuning.

Results

  • Measurement accuracy and response speed significantly improved, with stable and reliable data;
  • Instruments adapt to complex Food & Beverage conditions, greatly reducing maintenance;
  • Helping customers achieve continuous online monitoring and cost reduction, highly recognized by customers.
Chemical & Pharmaceutical

Chemical & Pharmaceutical — Reactors and Storage Tanks · RTD Application Case

Project Background

Chemical & Pharmaceutical customers at the Reactors and Storage Tanks stage have long faced the challenge of highly corrosive media and complex conditions, requiring corrosion-resistant and explosion-proof instruments, placing higher demands on the stability and measurement accuracy of RTD.

Solution

Based on the on-site conditions, the Ainstru technical team selected suitable RTD for the customer and delivered one-stop service covering selection, installation, commissioning and tuning.

Results

  • Measurement accuracy and response speed significantly improved, with stable and reliable data;
  • Instruments adapt to complex Chemical & Pharmaceutical conditions, greatly reducing maintenance;
  • Helping customers achieve continuous online monitoring and cost reduction, highly recognized by customers.
Power & Energy

Power & Energy — Power Plant Boiler and Turbine · RTD Application Case

Project Background

Power & Energy customers at the Power Plant Boiler and Turbine stage have long faced the challenge of unstable measurement under high temperature and pressure, with high maintenance and downtime costs, placing higher demands on the stability and measurement accuracy of RTD.

Solution

Based on the on-site conditions, the Ainstru technical team selected suitable RTD for the customer and delivered one-stop service covering selection, installation, commissioning and tuning.

Results

  • Measurement accuracy and response speed significantly improved, with stable and reliable data;
  • Instruments adapt to complex Power & Energy conditions, greatly reducing maintenance;
  • Helping customers achieve continuous online monitoring and cost reduction, highly recognized by customers.