Measuring principle:
-
This online COD analyzer uses the potassium dichromate method to measure COD: a mixture of water sample, potassium dichromate, silver sulfate solution (used as a catalyst to enable more complete oxidation of straight-chain aromatic hydrocarbon compounds) and concentrated sulfuric acid is heated to 175°C in the digestion cell. During this period, chromium ions as an oxidant are reduced from valence VI to valence III, causing a color change. The degree of color change corresponds to the content of organic compounds in the sample. The online COD analyzer directly displays the COD of the sample through colorimetric conversion.
-
The main interfering substance is chloride, which is removed by adding mercury sulfate to form a complex.
Instrument features:
-
The classic potassium dichromate oxidation is combined with highly efficient digestion, complying with the Chinese national standard dichromate method COD testing principle.
-
This online COD analyzer is equipped with an accurate optical metering system and has an automatic calibration function, making measurement more accurate.
-
It can be connected to HACH sc200 and sc1000 controllers, offering good expandability.
-
10~5000 mg/L three-range automatic switching, with no need to determine the range of the water sample COD.
-
The optimized internal structure of this online COD analyzer makes comparison between online data and laboratory data easier.
-
It has an automatic cleaning function. Combined with piston pump technology and a corrosion-resistant tubing design, maintenance is reduced.
Technical specifications
-
Measuring method Potassium dichromate high-temperature digestion, colorimetric determination
-
Measuring range 10~5,000 mg/L
-
Indication error (potassium hydrogen phthalate test) ±8%
-
Repeatability ≤3%
-
Zero drift (24 h) ±5 mg/L
-
Range drift (24 h) ±10 mg/L
-
Digestion time Automatic, selectable at approximately 3, 5, 10, 20, 30, 40, 60, 80, 100 or 120 minutes
-
Measurement interval Continuous measurement, 1~24 hour interval measurement, trigger-start measurement, user-defined interval
-
Calibration Selectable no calibration, automatic calibration or user-defined calibration (up to 5 points)
-
Output 2 current outputs: 0/4-20 mA, maximum load 500Ω
-
2 multifunctional output relays: 24V 1A
-
Read via SD card on sc200 or sc1000 controller
-
Service interface RS232
-
Digital communication MODBUS or Profibus, via sc200 or sc1000 controller
-
Display Via HACH sc200 or sc1000 display
-
Ambient temperature +5~+40℃
-
Power supply 220 VAC ±10%/50-60 Hz
-
Power consumption Approx. 100VA
-
Dimensions 535.80 mm× 711.5 mm× 376 mm
-
Weight Approx. 38 kg (excluding reagents)
Related reagents
| Order No. | Description | Parameters | Analysis method | Packaging | Range | Guaranteed shelf life (days) |
|---|---|---|---|---|---|---|
| LCW420 | Chemical oxygen demand (COD) maximum (max) reagent | - | - | - | - | 150 |
Spare parts
| Spare part Order No. | Instrument model | Chinese description | English description |
|---|---|---|---|
| COD05_07 | CODmax plus sc potassium dichromate method online COD analyzer | Piston pump connector cable | connector cable piston pump |
| COD07_00 | CODmax plus sc potassium dichromate method online COD analyzer | Bus cable | bus cable |
| COD08_11 | CODmax plus sc potassium dichromate method online COD analyzer | Side panel right | SIDE PANEL RIGHT? |
| COD08_21 | CODmax plus sc potassium dichromate method online COD analyzer | Side panel left | SIDE PANEL LEFT?? |
| COD10_11 | CODmax plus sc potassium dichromate method online COD analyzer | Screw | SCREWING / VERSCHRAUBUNG |
| COD43_00 | CODmax plus sc potassium dichromate method online COD analyzer | Safety lock | magnetic latching board |
| COD44_00 | CODmax plus sc potassium dichromate method online COD analyzer | Detector at dosing tube (2 pcs) | photoelement dosing unit complete |
| EHH021 | CODmax plus sc potassium dichromate method online COD analyzer | Piston holder | syring holder |
| EXP033 | CODmax plus sc potassium dichromate method online COD analyzer | Piston pump | piston pump |
| HHA_X | CODmax plus sc potassium dichromate method online COD analyzer | Connecting tube between air valve and digestion cell | Valve cuvette adapter assembly for maintenance |
| LCW420 | CODmax plus sc potassium dichromate method online COD analyzer | Reagent kit | CODMAX regent |
| LZV123 | CODmax plus sc potassium dichromate method online COD analyzer | Connecting cable between power board and main board | SET CONNECTOR CABLE |
| LZV208 | CODmax plus sc potassium dichromate method online COD analyzer | Temperature sensor | Temperature sensor complete |
| OBA027 | CODmax plus sc potassium dichromate method online COD analyzer | Photometer board | photometer board |
| YAA857 | CODmax plus sc potassium dichromate method online COD analyzer | RS485 board | INTERFACE BOARD RS 485 |
| YAA903 | CODmax plus sc potassium dichromate method online COD analyzer | Display card | LCD module card |
| YAA909 | CODmax plus sc potassium dichromate method online COD analyzer | Motor board | motor board |
| YAA952 | CODmax plus sc potassium dichromate method online COD analyzer | Power supply board | Power supply board(ZDA269) |
| YAB112 | CODmax plus sc potassium dichromate method online COD analyzer | Main board | Processor card |
| YY0000010 | CODmax plus sc potassium dichromate method online COD analyzer | Housing | HOUSING MOUNTED |
| YY0000040 | CODmax plus sc potassium dichromate method online COD analyzer | Safety panel | COD51_00,safety window complete |
| YY0000050 | CODmax plus sc potassium dichromate method online COD analyzer | Connecting tube between air valve and digestion cell | COD47_00,VALUE CUVETTE ADAPTOR COMPLETE |
| YY0000060 | CODmax plus sc potassium dichromate method online COD analyzer | Tubing assembly | COD50_00,TUBE SET COMPLETE |
| YY0000070 | CODmax plus sc potassium dichromate method online COD analyzer | Digestion cell protection plate | COD48_00,SAFETY PLATE ASSY |
| YY0000080 | CODmax plus sc potassium dichromate method online COD analyzer | Tubing and fittings | COD87_00,TUBE 10 COMPLETE |
| YY0000089 | CODmax plus sc potassium dichromate method online COD analyzer | O-ring (sealing rings at upper and lower ends of dosing tube) | EZD259,O-ring 6,07x1,78 Kalrez 6375 |
| YY0000090 | CODmax plus sc potassium dichromate method online COD analyzer | Annual maintenance kit | TOS_X,one year wearing spareparts |
| YY0000091 | CODmax plus sc potassium dichromate method online COD analyzer | Dosing tube solenoid valve (air horizontal valve) | EXV055,valve block top |
| YY0000092 | CODmax plus sc potassium dichromate method online COD analyzer | O-ring | LZV047,O-ring 2,2 x 1,6mm material: Viton |
| YY0000093 | CODmax plus sc potassium dichromate method online COD analyzer | Tubing connector | HRS196,fitting pharmed tube |
| YY0000104 | CODmax plus sc potassium dichromate method online COD analyzer | O-ring (sealing ring at upper end of digestion glass) | EZD260,O-ring 9.19 x 2.62mm Kalrez 6375 |
| YY0000110 | CODmax plus sc potassium dichromate method online COD analyzer | O-ring kit | EZD_X,O-ring kit for maintenance |
| YY0000113 | CODmax plus sc potassium dichromate method online COD analyzer | Dosing tube | HBK061,Dose cuvette |
| YY0000119 | CODmax plus sc potassium dichromate method online COD analyzer | Lower fixing module of digestion unit | COD10_01,cuvette holder bottom |
| YY0000120 | CODmax plus sc potassium dichromate method online COD analyzer | Piston pump assembly | COD46_00,PISTON PUMP ARRANGED |
| YY0000122 | CODmax plus sc potassium dichromate method online COD analyzer | Digestion cell | BBK036,cuvette |
| YY0000128 | CODmax plus sc potassium dichromate method online COD analyzer | Fan assembly | COD10_36, fan holder |
| YY0000139 | CODmax plus sc potassium dichromate method online COD analyzer | Sample inlet tube | ELS027,raw material, FEP 1,6x3,2 mm |
| YY0000140 | CODmax plus sc potassium dichromate method online COD analyzer | Upper fixing module of digestion unit | COD10_02,cuvette holder top |
| YY0000141 | CODmax plus sc potassium dichromate method online COD analyzer | Sample inlet/drain hose | ELS895,raw material, pharmed 65 2,4x5,6 mm |
| YY0000156 | CODmax plus sc potassium dichromate method online COD analyzer | Combination valve | EXV054,valve block bottom |
| YY0000195 | CODmax plus sc potassium dichromate method online COD analyzer | Safety cover | SAFETY COVER |
| YY0000219 | CODmax plus sc potassium dichromate method online COD analyzer | Air valve (digestion air valve) | EXV056,air valve ( with connector cable ) |
| YY0000287 | CODmax plus sc potassium dichromate method online COD analyzer | Upper connecting tube | ERA005,ADAPTOR END TOP |
| YY0000289 | CODmax plus sc potassium dichromate method online COD analyzer | Digestion air safety valve diaphragm | EVQ241,membrane FFKM |
| YY0000290 | CODmax plus sc potassium dichromate method online COD analyzer | Standard solution valve | EXV065,magn.valve type 6116,FFKM-membrane |
Ordering guide
-
9302600 CODmax plus sc Online automatic chemical oxygen demand analyzer
-
LCW420 Analyzer reagent
-
Note: The above CODmax plus sc order numbers do not include the sc controller. Please order the corresponding sc200 or sc1000 controller as required.
Source: Public information from the HACH water quality website
Casos de servicio
Tratamiento de Aguas Residuales Municipales — HACH-ONLINE: caso de aplicación
Antecedentes del proyecto
El proyecto abarca las etapas de entrada, biológica y salida de una planta municipal. El trabajo se centra en HACH-ONLINE y la tarea principal es monitorización continua de vertido con instrumentos Hach.
Condiciones de servicio
El medio y el servicio se caracterizan por alto contenido de sólidos y componentes corrosivos, con grandes variaciones diarias de caudal, y además los parámetros del efluente oscilan y las sondas se ensucian con lodos y biopelícula. Por eso HACH-ONLINE no puede especificarse de forma genérica: requiere configurar por rango, intervalo de reactivo y acondicionamiento.
Solución
El equipo técnico de Ainstru trabajó por la vía de análisis en línea: integrar en torno a cada punto de monitorización. La configuración incluye configurar por rango, intervalo de reactivo y acondicionamiento; se cuidaron estos requisitos: cumplimiento estable del nivel 1A y reporte de datos en línea.
Resultados
Desde la puesta en marcha el sistema funciona de forma estable; el resultado práctico es suministro estable de reactivos y consumibles, con mantenimiento previsible.
Alimentación y Bebidas — HACH-ONLINE: caso de aplicación
Resumen del proyecto
La instalación está en las etapas de dosificación, pasteurización, llenado y limpieza CIP. El cliente quería configurar bien HACH-ONLINE en la planta existente, con el objetivo de monitorización continua de vertido con instrumentos Hach.
Retos técnicos
Lo que realmente da problemas es jarabes, lácteos y bebidas proclives a incrustaciones, con estrictos requisitos higiénicos, y las limpiezas frecuentes exigen instrumentos resistentes a ciclos CIP/SIP con sosa caliente. Todo ello implica que la configuración de HACH-ONLINE deba incluir configurar por rango, intervalo de reactivo y acondicionamiento.
Configuración
El suministro y la instalación se organizaron por la vía de análisis en línea: integrar en torno a cada punto de monitorización. En planta esto supuso configurar por rango, intervalo de reactivo y acondicionamiento; se revisó línea a línea: requisitos de seguridad alimentaria y diseño higiénico 3-A / EHEDG.
Comportamiento
Tras la puesta en marcha el resultado es suministro estable de reactivos y consumibles, con mantenimiento previsible, y las lecturas se usan ya para ajustar el proceso.
Química y Farmacéutica — HACH-ONLINE: caso de aplicación
Situación en planta
Se trata de un cliente del sector Química y Farmacéutica cuya planta está en reactores, vasijas de dosificación y tanques de producto terminado. La tarea principal era monitorización continua de vertido con instrumentos Hach.
El problema
La dificultad está en medios muy corrosivos con disolventes y cristalización, en parte en atmósferas explosivas; a la vez, la corrosión y la protección Ex acortan la vida útil y encarecen las paradas. Por eso la especificación implica configurar por rango, intervalo de reactivo y acondicionamiento.
Qué hicimos
Adoptamos un enfoque de análisis en línea: integrar en torno a cada punto de monitorización. La entrega se construyó en torno a configurar por rango, intervalo de reactivo y acondicionamiento; se verificó punto por punto: certificación Ex y trazabilidad de datos según GMP.
Resultado operativo
En servicio el sistema responde a lo previsto: suministro estable de reactivos y consumibles, con mantenimiento previsible, con un mantenimiento rutinario notablemente menor.
Energía y Electricidad — HACH-ONLINE: caso de aplicación
Contexto de aplicación
agua de alimentación, condensado de turbina y sistemas de desulfuración/DeNOx es un servicio típico de Energía y Electricidad. El proyecto partió de HACH-ONLINE, centrándose en monitorización continua de vertido con instrumentos Hach.
Limitaciones del proceso
deriva a alta temperatura y presión, con paradas no programadas muy costosas, y alta temperatura y presión, con vapor y agua salina no hace sino acentuarlo. Por eso la especificación de HACH-ONLINE debe cubrir configurar por rango, intervalo de reactivo y acondicionamiento.
Enfoque de suministro
La entrega se gestionó como análisis en línea: integrar en torno a cada punto de monitorización. La ejecución se centró en configurar por rango, intervalo de reactivo y acondicionamiento, junto con cumplimiento de normativas del sector y objetivos de disponibilidad.
Resultados en campo
La operación evidencia el beneficio: suministro estable de reactivos y consumibles, con mantenimiento previsible, con una confianza claramente mayor en este punto de medición.
Recomendados similares
HACHAnalizador de fósforo total / ortofosfato Sigma
HACHAnalizador de dureza del agua en línea
HACHAnalizador automático multiparamétrico de agua
HACHAnalizador multiparamétrico de agua en línea
HACHAnalizador de sílice en línea
HACHAnalizador de nitrógeno amoniacal
HACHAnalizador de cloro libre en línea
HACH