Instrumentation · Engineering · Package · PLC
0755-8222 1901 sales@ainstru.com

CPS76E PH/ORP组合电极

 CPS76E  PH/ORP组合电极

Memosens CPS76E

Configurable Models 6 models
Memosens CPS76E is the heavy-duty specialist for simultaneous pH and ORP measurement. Its unique, contamination-resistant reference guarantees stable measurement in polluted, poisoning media and in media with low conductivity. Thanks to Memosens 2.0 digital technology, CPS76E combines maximum process integrity with simple operation. It resists moisture, enables lab calibration and offers extended storage of calibration and process data providing the perfect basis for predictive maintenance.
Compare

ORP / Redox

Measuring Principle Sensor ORP / Redox
Application Process technology and monitoring of processes with:
- Rapidly changing pH values
- High levels of electrode poisons, such as H2S
Characteristic Digital pH/ORP electrode for chemical process with an ion trap for poison-resistant reference
Measuring range ORP: –1 500 to 1 500 mV
Application B
• pH: 0 to 14
Application H
• pH: 0 to 12
Measuring principle Gel compact electrode with ceramic junction and ion trap for simultaneous measurement of pH, ORP and rH value (in rH mode)
Design All shaft lengths with temperature sensor
Advanced gel technology
Material Sensor shaft: Glass to suit process
pH membrane glass: Type B, Type N
Metal lead: Ag/AgCl
Open aperture: Ceramic junction,
zirconium dioxide
ORP measuring element: Platinum
O-ring: FKM
Process coupling: PPS fiber-glass reinforced
Nameplate: Ceramic metal oxide
Dimension Diameter: 12 mm (0.46 inch)
Shaft lengths: 120, 225, 360 and 425 mm
(4.68, 8.77, 14.04 and 16,57 inch)
Process temperature Application B and H:
0 to 140 ° C (32 to 284 ° F)
Version TB:
0 to 140 ° C (32 to 284 ° F)
Version TU, TP (pressurized reference):
0 to 140 ° C (32 to 284 ° F) (140 ° C (284 ° F) for a limited period (e.g. 20 min))
maximum 100 ° C (212 ° F) in continuous operation due to increasing
pressure drop at T> 100 ° C (212 ° F)
Process pressure Application B: 0.8 to 14 bar (11.6 to 203 psi) absolute
Application H: 0.8 to 7 bar (11.6 to 101,5 psi) absolute
Temperature sensor NTC 30k
Ex certification With ATEX, IECEx, CSA C/US, NEPSI, Japan Ex and INMETRO approvals for use in
hazardous areas Zone 0, Zone 1 and Zone 2
Connection Inductive, digital connection head with Memosens 2.0 technology
Ingress protection IP68

pH

Measuring Principle Potentiometric
Application Process technology and monitoring of processes with:
- Rapidly changing pH values
- High levels of electrode poisons, such as H2S
Characteristic Digital pH/ORP electrode for chemical process with an ion trap for poison-resistant reference
Measuring range ORP: –1 500 to 1 500 mV
Application B
• pH: 0 to 14
Application H
• pH: 0 to 12
Measuring principle Gel compact electrode with ceramic junction and ion trap for simultaneous measurement of pH, ORP and rH value (in rH mode)
Design All shaft lengths with temperature sensor
Advanced gel technology
Material Sensor shaft: Glass to suit process
pH membrane glass: Type B, Type N
Metal lead: Ag/AgCl
Open aperture: Ceramic junction,
zirconium dioxide
ORP measuring element: Platinum
O-ring: FKM
Process coupling: PPS fiber-glass reinforced
Nameplate: Ceramic metal oxide
Dimension Diameter: 12 mm (0.46 inch)
Shaft lengths: 120, 225, 360 and 425 mm
(4.68, 8.77, 14.04 and 16,57 inch)
Process temperature Application B and H:
0 to 140 ° C (32 to 284 ° F)
Version TB:
0 to 140 ° C (32 to 284 ° F)
Version TU, TP (pressurized reference):
0 to 140 ° C (32 to 284 ° F) (140 ° C (284 ° F) for a limited period (e.g. 20 min))
maximum 100 ° C (212 ° F) in continuous operation due to increasing
pressure drop at T> 100 ° C (212 ° F)
Process pressure Application B: 0.8 to 14 bar (11.6 to 203 psi) absolute
Application H: 0.8 to 7 bar (11.6 to 101,5 psi) absolute
Temperature sensor NTC 30k
Ex certification With ATEX, IECEx, CSA C/US, NEPSI, Japan Ex and INMETRO approvals for use in
hazardous areas Zone 0, Zone 1 and Zone 2
Connection Inductive, digital connection head with Memosens 2.0 technology
Ingress protection IP68

The Memosens CPS76E is the measurement specialist for demanding conditions and can measure pH and ORP simultaneously. The unique contamination-resistant reference system reliably measures heavily contaminated media and low-conductivity media. The CPS76E uses Memosens 2.0 digital technology with high process integrity and very simple operation. The waterproof, moisture-proof design allows laboratory sensor calibration and supports extended data storage for calibration and process parameters. It is the perfect basis for predictive maintenance.

Benefits
Memosens 2.0 supports an extended storage unit for calibration and process parameters, enabling better trend identification and providing the perfect basis for future predictive maintenance and improved IIoT services.

Simultaneous measurement of pH, ORP and rH values (in rH mode) provides rich process information and is a solid foundation for strict process control.

Optional pressurized reference system for reliable measurement of media prone to clogging, such as dispersions.

The electrode has a dedicated ion trap that effectively prevents toxic diffusion to the lead and reference system, ensuring a long service life.

The electrode can be installed upside down for flexible installation.

Non-contact inductive signal transmission for high process safety.

Minimizes production downtime, extends electrode service life and reduces operating costs.

Applications
Chemical process and process monitoring:

Rapidly fluctuating pH or ORP values

Highly toxic media, e.g. H2S

Approved by ATEX, IECEx, CSA C/US, NEPSI, JPN, INMETRO for use in explosion-protection zones 0, 1 and 2.

 

E+H Product Model Configuration Tool

Please enter the product model to configure

Product comparison

Please enter the product models to compare in the fields below; up to three products can be compared

 

   

Technical documentation

Technical data

 

 

Documents/Manuals/Software

 

 

Accessories/Spare parts

 

Source: Endress+Hauserpublic information from the website

Model Name
CPS76E-AA7BTB2 PH/ORP组合电极
CPS76E-NA7BTU2 PH/ORP组合电极
CPS76E-AA7HTP2 PH/ORP组合电极
CPS76E-AA7BTU2 PH/ORP组合电极
CPS76E-NA7BTB2 PH/ORP组合电极
CPS76E-NA7HTP2 PH/ORP组合电极
Municipal Wastewater Treatment

Municipal Wastewater Treatment — Wastewater Treatment Plant · pH Application Case

Project Background

Municipal Wastewater Treatment customers at the Wastewater Treatment Plant stage have long faced the challenge of insufficient accuracy in effluent quality monitoring and instruments prone to contamination causing data drift, affecting compliance discharge, placing higher demands on the stability and measurement accuracy of pH.

Solution

Based on the on-site conditions, the Ainstru technical team selected suitable pH for the customer and delivered one-stop service covering selection, installation, commissioning and tuning.

Results

  • Measurement accuracy and response speed significantly improved, with stable and reliable data;
  • Instruments adapt to complex Municipal Wastewater Treatment conditions, greatly reducing maintenance;
  • Helping customers achieve continuous online monitoring and cost reduction, highly recognized by customers.
Food & Beverage

Food & Beverage — Food Production Line · pH Application Case

Project Background

Food & Beverage customers at the Food Production Line stage have long faced the challenge of high hygienic measurement requirements, with inline instruments needing to withstand CIP/SIP cleaning while maintaining stable accuracy, placing higher demands on the stability and measurement accuracy of pH.

Solution

Based on the on-site conditions, the Ainstru technical team selected suitable pH for the customer and delivered one-stop service covering selection, installation, commissioning and tuning.

Results

  • Measurement accuracy and response speed significantly improved, with stable and reliable data;
  • Instruments adapt to complex Food & Beverage conditions, greatly reducing maintenance;
  • Helping customers achieve continuous online monitoring and cost reduction, highly recognized by customers.
Chemical & Pharmaceutical

Chemical & Pharmaceutical — Reactors and Storage Tanks · pH Application Case

Project Background

Chemical & Pharmaceutical customers at the Reactors and Storage Tanks stage have long faced the challenge of highly corrosive media and complex conditions, requiring corrosion-resistant and explosion-proof instruments, placing higher demands on the stability and measurement accuracy of pH.

Solution

Based on the on-site conditions, the Ainstru technical team selected suitable pH for the customer and delivered one-stop service covering selection, installation, commissioning and tuning.

Results

  • Measurement accuracy and response speed significantly improved, with stable and reliable data;
  • Instruments adapt to complex Chemical & Pharmaceutical conditions, greatly reducing maintenance;
  • Helping customers achieve continuous online monitoring and cost reduction, highly recognized by customers.
Power & Energy

Power & Energy — Power Plant Boiler and Turbine · pH Application Case

Project Background

Power & Energy customers at the Power Plant Boiler and Turbine stage have long faced the challenge of unstable measurement under high temperature and pressure, with high maintenance and downtime costs, placing higher demands on the stability and measurement accuracy of pH.

Solution

Based on the on-site conditions, the Ainstru technical team selected suitable pH for the customer and delivered one-stop service covering selection, installation, commissioning and tuning.

Results

  • Measurement accuracy and response speed significantly improved, with stable and reliable data;
  • Instruments adapt to complex Power & Energy conditions, greatly reducing maintenance;
  • Helping customers achieve continuous online monitoring and cost reduction, highly recognized by customers.