Instrumentation · Engineering · Package · PLC
0755-82221901 sales@ainstru.com

25636900Discontinued

25636900
Product family: TU5400sc Turbidimeter + SC200 Controller
TU5400sc Low Range Laser Turbidimeter, ISO Version with SC200 Controller, 24 VDC, 1 ChannelDiscontinued
HACH 25636900
The next standard in the evolution of turbidity

产品概述

The next standard in the evolution of turbidity

Compliant with ISO Method 7027-1 and perfect for membrane filtration applications, the TU5400sc employs a patented optical design that sees more of your sample than any other turbidimeter, delivering the best low level precision and sensitivity while minimizing variability between measurements. For the first time you will be able to remove uncertainty of which measurement to trust between your lab and process instruments, thanks to identical 360° x 90° Detection Technology in any TU5 Series turbidimeter.

The TU5400sc Ultra-High Precision Low Range Laser Turbidimeter dramatically reduces the time needed to get a turbidity measurement you can rely on. With a 10mL sample size, response time is over 15 times faster than an FT660. A stable laser light source eliminates the need for annual lamp replacements. The turbidimeter’s measurement surface area has been reduced by 98%, allowing you to save time cleaning.

All of this, along with the ability to measure to 0.0002 FNU, gives you the next standard in the evolution of turbidity.

主要特点

  • Groundbreaking 360° x 90° Detection™ Technology:The TU5 Series employs a patented optical design that sees more of your sample than any other turbidimeter, delivering the best low level precision and sensitivity while minimizing variability from test to test.
  • Matching lab and online results:For the first time you will be able to remove the uncertainty of which measurement to trust, thanks to identical 360° x 90° Detection™ Technology in both instruments.
  • Everything about turbidity – faster:The TU5 Series dramatically reduces the time needed to get a turbidity measurement you can rely on, with 98% less online sample surface area to clean, sealed vials for calibration, and the elimination of the need for indexing and silicone oil in the lab. Not to mention, a smaller online sample volume means you will detect events almost immediately.
  • No surprises:Prognosys™ monitors your TU5 Series online instrument, proactively alerting you to maintenance needs before your measurement becomes questionable. And a Hach Service Agreement protects your investment and helps ensure that you stay in compliance and on budget.

技术参数

参数名称说明
Accuracy±2% of reading plus 0.01 NTU from 0 - 40 NTU
Cable Length1.6 m (5.25 ft), extendable up to 50 m (164 ft) for instrument without accessories
Calibration MethodFor Formazin and Stablcal®: 20 NTU from 0 to 40 NTU; at 20 FNU and 600 NTU for full range Custom calibration for up to 6-point calibrations
CertificationsCE compliant US FDA accession number: 1420493-000 EPA version, 1420492-000 ISO version Australian ACMA Marking
Channels1
Controller1 Channel, 24 VDC
Controller OptionsYes
Dimensions (H x W x D)249 mm x 268 mm x 190 mm
Enclosure RatingElectronic compartment IP55; all other functional units IP65 with process head/ACM attached to the TU5300sc/TU5400sc instrument
Fitting TypeSample quick connector: ¼-in. for ¼-in. tubing
InstrumentStandard Process Head
Light SourceClass 2 laser product, with embedded 650 nm (EPA 0.43 mW) or Class 1 laser product, with embedded 850 nm (ISO), max. 0.55 mW (complies with IEC/EN 60825-1 and to 21 CFR 1040.10 in accordance with Laser Notice No. 50)
MaterialASA S 777K / RAL7000, TPE RESIN Elastocon STK40, Thermoplastic Elastomer TPS-SEBS
ModelTU5400sc
Mounting ConfigurationsWall Mount
Operating HumidityRelative humidity: 5 - 95% at different temperatures, non-condensing
Operating Temperature Range0 - 50 °C
OptionsProcess Head Only
Power Requirements (Voltage)24 VDC
RangeISO: 0 - 1000 NTU / FNU / TE/F / FTU 0 - 250 EBC
RegulatoryISO
RepeatabilityBetter than 1% of reading or ±0.0006 NTU on Formazin at 25 °C (77 °F), whichever is greater
Resolution0.0001 NTU / FNU / TE/F / FTU / EBC
Response TimeT90 <30 seconds at 100 mL/min
Sample Flow Rate100 - 1000 mL/min; optimal flow rate: 200 - 500 mL/min
Sample Pressure6 bar (87 psi) maximum, compared to air at sample temperature range from 2 - 40 °C (35.6 - 104 °F)
Sample Temperature2 to 60 °C (35 to 140 °F)
Signal Average Time1 to 90 seconds
Storage Conditions-40 - 60 °C (-40 - 140 °F)
Stray Light<10 mNTU
System CheckNo
UnitsNTU, FNU, TE/F, FTU, EBC
VerificationLiquid: Stablcal®, Formazin (0.1 - 40 NTU) Dry: Glass Rod at <0.1 NTU
Weight2.7 kg (5.0 kg with all accessories)
What's included?TU5400 sc Turbidimeter, User Manual, Wall Mount, Vial Replacement Tool, Screw Set, Drying Cartridge, Flow Regulator, Service Bracket

应用领域

Filtration · Water Reuse

同系列其他配置

型号
2979605TU5400sc Ultra-High Precision Low Range Laser Turbidimeter, EPA versio
2979805TU5400sc UltraHigh Precision Low Range Laser Turbidimeter, Flow Sensor
2979705TU5400sc Ultra-High Precision Low Range Laser Turbidimeter, EPA versio
2979855TU5400sc Ultra-High Precision Low Range Turbidimeter, Flow Sensor, Aut
2979905TU5400sc Low Range Laser Turbidimeter, Flow Sensor, Auto Cleaning, RFI
2979955TU5400sc Ultra-High Precision Low Range Turbidimeter, Flow Sensor, Aut
8627100TU5400sc Ultra-High Precision Low Range Laser Turbidimeter, EPA versio
8627300TU5400sc Ultra-High Precision Low Range Laser Turbidimeter, Flow Senso
8627500TU5400sc Ultra-High Precision Low Range Laser Turbidimeter, Flow Senso
25635700TU5400sc Ultra-High Precision Low Range Laser Turbidimeter, ISO Versio
25635800TU5400sc Ultra-High Precision Low Range Laser Turbidimeter, ISO Versio
25635900TU5400sc Low Range Laser Turbidimeter, Flow Sensor, RFID, System Check
25636000TU5400sc Low Range Laser Turbidimeter, Flow Sensor, RFID, System Check
25636100TU5400sc Low Range Laser Turbidimeter Flow Sensor, Auto Cleaning, RFID
25637000TU5400sc Low Range Laser Turbidimeter, Flow Sensor, RFID, System Check
25637100TU5400sc Low Range Laser Turbidimeter, Flow Sensor, RFID, System Check
25637200TU5400sc Low Range Laser Turbidimeter, Flow Sensor, Auto Cleaning, RFI
25637300TU5400sc Low Range Laser Turbidimeter, Flow Sensor, Auto Cleaning, RFI

资料下载

19 models · click to collapse