Instrumentation · Engineering · Package · PLC
0755-82221901 sales@ainstru.com

PAH500 PAH (polycyclic aromatic hydrocarbons) analyzer

PAH500 PAH (polycyclic aromatic hydrocarbons) analyzer
PAH (polycyclic aromatic hydrocarbons) analyzer
HACH PAH500
The measuring principle of the PAH500 sc polycyclic aromatic hydrocarbon analyzer is based on the fluorescence properties of PAHs. After excitation by a UV light source, PAHs emit light of longer wavelength after a short delay. The intensity of this light is measured, and it is proportional to the PAH concentration. This measuring principle is much more sensitive than absorption and scattered light measurement. PAH testing can even detect the slightest traces of PAH contamination in water. PAHs are a component of most mineral oil products and are a very specific indicator of oil contamination in water bodies and process water.

Application Industry:

The PAH500 sc PAH analyzer is suitable for PAH detection fields such as marine vessel wastewater monitoring.

Instrument Features:

  • The PAH500 sc PAH analyzer uses the ultraviolet fluorescence method for online measurement of polycyclic aromatic hydrocarbons (oil) content in water, with high sensitivity;

  • The PAH500 sc PAH analyzer is made of titanium alloy, capable of measuring seawater and high-salinity samples, with low maintenance;

  • The PAH500 sc PAH analyzer features turbidity correction, minimizing the influence of suspended particles in water on measurement values;

  • The PAH500 sc PAH analyzer can be connected to the sc controller platform, offering multiple calibration methods and data output modes.

Technical Specifications

  • Measuring parameter                                PAH (polycyclic aromatic hydrocarbons)

  • Measuring method                                PAH ultraviolet fluorescence measurement method

  • Light source                                       UV LED

  • Excitation wavelength                                255 nm

  • Detection wavelength                               370 nm

  • Measuring range                               0 to 900 μg/L PAH PHE (phenanthrene) equivalent

  • Display units                              ppb, ppm, μg/L, mg/L PAH

  • Measuring interval                       10 seconds

  • Accuracy                                     At constant temperature and flow rate, ± 5% or ± 5 μg/L (whichever is greater)

  • Detection limit                              3 μg/L PAH

  • Response time                             10 seconds

  • Turbidity compensation                             0 to 150 FNU (or 0 to 150 NTU); the SC1000 controller or other programmable controller calculates turbidity compensation based on PAH readings and turbidity readings

Ordering Guide
P/N EN description CN description
LXV541.99.1001H PAH500 Digital PAH Sensor, 1.0 m Cable, Bracket PAH500 Digital PAH Sensor, 1.0 m Cable, Bracket
LXV541.99.2001H PAH500 Digital PAH Sensor, 7.7 m Cable, Bracket PAH500 Digital PAH Sensor, 7.7 m Cable, Bracket
LXV541.99.3001H PAH500 Digital PAH Sensor, 15 m Cable, Bracket PAH500 Digital PAH Sensor, 15 m Cable, Bracket
LXV541.99.4001H PAH500 Digital PAH Sensor, 31 m Cable, Bracket PAH500 Digital PAH Sensor, 31 m Cable, Bracket
LZC541 PAH500 Sensor Calibration Kit PAH500 Sensor Calibration Kit
2563899 Phenanthrene Standard, 25 ppb PAH500 Phenanthrene Standard, 25 ppb, 40 mL
2563901 Phenanthrene Standard, 50 ppb PAH500 Phenanthrene Standard, 50 ppb, 40 mL
2088000 Alconox Detergent Detergent: Alconox
2088153 Liqui-Nox Detergent Detergent: Liqui-Nox
1227642 Isopropyl Alcohol Isopropyl Alcohol
244932 Sulfuric Acid, 5.25 N Sulfuric Acid, 5.25 N
9774000 Cleaning Swab, pk/50 Cleaning Swab, 50 pcs
2749320 Disposable Pipettes, pk/20 Disposable Pipettes, 5.0 mL, 20 pcs

Source: Public information from the HACH Water Quality website

Municipal Wastewater Treatment

Municipal Wastewater Treatment · Controllers and Sensors — Application Case

Site Conditions

This is a Municipal Wastewater Treatment customer whose plant is at the inlet, biological and outlet stages of a municipal plant. The main task was online water monitoring with a controller and sensor combination.

The Problem

The difficulty lies in high solids and corrosive fractions, with large diurnal flow swings; at the same time effluent parameters drift and probes foul with sludge and biofilm. Specification therefore means matching each controller channel to its sensor type.

What We Did

We took a Hach channel approach — supplying against the Hach original part numbers. Delivery was built around matching each controller channel to its sensor type, with point-by-point checks on: stable compliance with Grade 1A discharge limits and online data reporting.

Operating Outcome

In service the system performs as intended — factory-matched combinations that make faults easier to locate, with routine maintenance falling noticeably.

Food & Beverage

Food & Beverage · Controllers and Sensors — Application Case

Application Context

dosing, pasteurising, filling and CIP stages is a typical Food & Beverage duty. The project started from Controllers and Sensors, focusing on online water monitoring with a controller and sensor combination.

Process Constraints

frequent cleaning demands instruments that withstand hot CIP/SIP caustic cycles, and syrups, dairy and beverages that scale easily, under strict hygienic requirements only amplify it. The Controllers and Sensors specification must therefore cover matching each controller channel to its sensor type.

Delivery Approach

Delivery was handled as the Hach channel — supplying against the Hach original part numbers. Execution centred on matching each controller channel to its sensor type, together with food-safety and hygienic design per 3-A / EHEDG.

Field Results

Operation shows the benefit clearly — factory-matched combinations that make faults easier to locate, with confidence in this measuring point improving markedly.

Chemical & Pharmaceutical

Chemical & Pharmaceutical · Controllers and Sensors — Application Case

Customer Profile

The typical setting in Chemical & Pharmaceutical is reactors, metering vessels and finished-product tanks. The measuring point is Controllers and Sensors; the focus is online water monitoring with a controller and sensor combination.

Key Difficulties

Conditions are highly corrosive media with solvents and crystallisation, partly in explosive atmospheres and, on top of that, corrosion plus Ex protection shortens instrument life and makes downtime costly, which a standard configuration cannot handle for long. The design therefore starts with matching each controller channel to its sensor type.

Implementation

The scope ran through the Hach channel: supplying against the Hach original part numbers. Technically it hinges on matching each controller channel to its sensor type, and on these requirements: Ex certification and GMP traceability of records.

Customer Benefit

In practice this means factory-matched combinations that make faults easier to locate, which cut the workload on the site maintenance team.

Power & Energy

Power & Energy · Controllers and Sensors — Application Case

Project Background

The project covers boiler feedwater, turbine condensate and FGD/DeNOx systems. The work centres on Controllers and Sensors and the core task is online water monitoring with a controller and sensor combination.

Operating Challenges

The medium and duty are characterised by high temperature and pressure with both steam and saline water, and drift under high temperature and pressure, with very costly unplanned outages. So Controllers and Sensors cannot be specified generically — it calls for matching each controller channel to its sensor type.

Solution

The Ainstru engineering team worked through the Hach channel: supplying against the Hach original part numbers. The configuration involves matching each controller channel to its sensor type, with these requirements in view: compliance with power-industry codes and unit availability targets.

Results

Since start-up the system has run steadily; the practical result is factory-matched combinations that make faults easier to locate.