Instrumentation · Engineering · Package · PLC
0755-82221901 sales@ainstru.com

FP360 sc oil-in-water analyzer

FP360 sc oil-in-water analyzer
sc oil-in-water analyzer
HACH FP360
FP360 sc oil-in-water analyzer is an online oil-in-water analyzer using UV fluorescence analysis, with high sensitivity. This oil-in-water analyzer can be optionally equipped with a corrosion-resistant titanium alloy probe, is not affected by suspended particulate matter, offers multiple mounting options, and can be connected to the sc controller.

Typical Applications

The FP360 sc oil analyzer is an online oil-in-water analyzer designed specifically for monitoring trace mineral oil in water. It is used in municipal and industrial wastewater, water treatment and process water treatment (non-hazardous areas), industrial process monitoring, seawater desalination influent monitoring, membrane treatment, etc.

Principle of Operation

The FP360 sc oil analyzer is based on the fluorescence characteristics of PAHs and uses the UV fluorescence method to determine the concentration of polycyclic aromatic hydrocarbons in water. PAHs excited by UV light produce short delayed luminescence, and the light intensity is measured at a wavelength of 360 nm. By detecting the PAH concentration, this oil-in-water analyzer calculates the total oil content in the water body.

Instrument Features

  • The FP360 sc oil analyzer is an online oil-in-water analyzer using the UV fluorescence method, with high sensitivity.

  • The FP360 sc oil analyzer can be equipped with an optional titanium alloy probe for measuring seawater and high-salinity samples.

  • The FP360 sc oil analyzer is not affected by suspended particles in water and operates stably.

  • Multiple installation methods are available: immersion, flow-through, and insertion.

  • The FP360 sc oil analyzer can be connected to the sc controller platform.

Technical Specifications

  • Measuring range:                   0-15mg/L (oil in water), 0-500 μg/L PAH

  •                             0-150mg/L (oil in water), 0-5000 μg/L PAH

  • Accuracy:                   approx. ±5% of reading or ±1% of full scale

  • Resolution:               0.01ppb (PAH) in the lowest measuring range

  • Response time:           10s

  • Installation method:           Immersion, flow-through, insertion

  • Light source:                  Miniature xenon flash lamp with interference filter (peak wavelength 254nm, 25nm FWHM). Output energy is controlled by an internal reference UV diode.

  • Detector:               360nm photodiode with optical filter

  • Probe material:             Stainless steel, titanium alloy (seawater corrosion resistant)

Ordering Guide

  • FP360 sc sensor standalone Order No.

  • LXV441.99.11102    FP360 sc, measuring range 0-500ppb, stainless steel, 10 m cable, without cleaning unit

  • LXV441.99.11202    FP360 sc, measuring range 0-500ppb, stainless steel, 10 m cable, with cleaning unit

  • LXV441.99.11302    FP360 sc, measuring range 0-500ppb, stainless steel, 1.5 m cable, without cleaning unit

  • LXV441.99.21102    FP360 sc, measuring range 0-5000ppb, stainless steel, 10 m cable, without cleaning unit

  • LXV441.99.21202    FP360 sc, measuring range 0-5000ppb, stainless steel, 10 m cable, with cleaning unit

  • LXV441.99.21302    FP360 sc, measuring range 0-5000ppb, stainless steel, 1.5 m cable, without cleaning unit

  • (The above sensor Order Nos. do not include the sc200 controller. The above measuring ranges refer to the concentration of polycyclic aromatic hydrocarbons.)

Installation Accessories

  • LZX914.99.11110    Stainless steel chain mounting accessory

  • LZY699    Flow-through mounting accessory (with panel and hose)

  • LZY619    5 m compressed air hose, 6mm OD (connects to chain mounting)

  • LZY620    10 m compressed air hose, 6mm OD (connects to chain mounting)

  • LZY621    25 m compressed air hose, 6mm OD (connects to chain mounting)

Source: Public information from the HACH Water Quality website

Municipal Wastewater Treatment

Municipal Wastewater Treatment · Controllers and Sensors — Application Case

Site Conditions

This is a Municipal Wastewater Treatment customer whose plant is at the inlet, biological and outlet stages of a municipal plant. The main task was online water monitoring with a controller and sensor combination.

The Problem

The difficulty lies in high solids and corrosive fractions, with large diurnal flow swings; at the same time effluent parameters drift and probes foul with sludge and biofilm. Specification therefore means matching each controller channel to its sensor type.

What We Did

We took a Hach channel approach — supplying against the Hach original part numbers. Delivery was built around matching each controller channel to its sensor type, with point-by-point checks on: stable compliance with Grade 1A discharge limits and online data reporting.

Operating Outcome

In service the system performs as intended — factory-matched combinations that make faults easier to locate, with routine maintenance falling noticeably.

Food & Beverage

Food & Beverage · Controllers and Sensors — Application Case

Application Context

dosing, pasteurising, filling and CIP stages is a typical Food & Beverage duty. The project started from Controllers and Sensors, focusing on online water monitoring with a controller and sensor combination.

Process Constraints

frequent cleaning demands instruments that withstand hot CIP/SIP caustic cycles, and syrups, dairy and beverages that scale easily, under strict hygienic requirements only amplify it. The Controllers and Sensors specification must therefore cover matching each controller channel to its sensor type.

Delivery Approach

Delivery was handled as the Hach channel — supplying against the Hach original part numbers. Execution centred on matching each controller channel to its sensor type, together with food-safety and hygienic design per 3-A / EHEDG.

Field Results

Operation shows the benefit clearly — factory-matched combinations that make faults easier to locate, with confidence in this measuring point improving markedly.

Chemical & Pharmaceutical

Chemical & Pharmaceutical · Controllers and Sensors — Application Case

Customer Profile

The typical setting in Chemical & Pharmaceutical is reactors, metering vessels and finished-product tanks. The measuring point is Controllers and Sensors; the focus is online water monitoring with a controller and sensor combination.

Key Difficulties

Conditions are highly corrosive media with solvents and crystallisation, partly in explosive atmospheres and, on top of that, corrosion plus Ex protection shortens instrument life and makes downtime costly, which a standard configuration cannot handle for long. The design therefore starts with matching each controller channel to its sensor type.

Implementation

The scope ran through the Hach channel: supplying against the Hach original part numbers. Technically it hinges on matching each controller channel to its sensor type, and on these requirements: Ex certification and GMP traceability of records.

Customer Benefit

In practice this means factory-matched combinations that make faults easier to locate, which cut the workload on the site maintenance team.

Power & Energy

Power & Energy · Controllers and Sensors — Application Case

Project Background

The project covers boiler feedwater, turbine condensate and FGD/DeNOx systems. The work centres on Controllers and Sensors and the core task is online water monitoring with a controller and sensor combination.

Operating Challenges

The medium and duty are characterised by high temperature and pressure with both steam and saline water, and drift under high temperature and pressure, with very costly unplanned outages. So Controllers and Sensors cannot be specified generically — it calls for matching each controller channel to its sensor type.

Solution

The Ainstru engineering team worked through the Hach channel: supplying against the Hach original part numbers. The configuration involves matching each controller channel to its sensor type, with these requirements in view: compliance with power-industry codes and unit availability targets.

Results

Since start-up the system has run steadily; the practical result is factory-matched combinations that make faults easier to locate.